Effective dose estimation in cone-beam computed tomography for dental use by Monte-Carlo simulation optimizing calculation numbers using a step-and-shoot method.

Yoshihiro Ozaki1, Hiroshi Watanabe1, Tohru Kurabayashi1.   

Abstract

OBJECTIVE: The objective of this study was to perform effective dose estimation in cone-beam CT for dental use (CBCT) using a Monte-Carlo simulation employing a step-and-shoot method as well as to determine the optimal number of steps.
METHODS: We simulated 3DX Accuitomo FPD8 as a CBCT model and estimated the effective doses of a large and a small field of view (FOV) examination against the virtual Rando phantom using a particle and heavy ion transport code system. We confirmed the results compared to those from a thermo-luminescence dosemeter (TLD) system in a real phantom and investigated how the reduced angle calculations could be accepted.
RESULTS: The effective doses of both FOVs estimated with each one degree were almost the same as those estimated from the TLD measurements. Considering the effective doses and the itemized organ doses, simulation with 5° and 10° is acceptable for the large and small FOV, respectively. We tried to compare an effective dose with a large FOV as well as with multiple small FOVs covering the corresponding area and found that the effective dose from six small FOVs was approximately 1.2 times higher than that of the large FOVs.
CONCLUSION: We successfully performed a Monte-Carlo simulation using a step-and-shoot method and estimated the effective dose in CBCT. Our findings indicate that simulation with 5° or 10° is acceptable based on the FOV size, while a small multiple FOV scan is recommended from a radiation protection viewpoint.

Entities:  

Keywords:  Monte Carlo Method; Radiation Dosimeters; Radiation, Dental; cone-beam computed tomography; radiation protection

Mesh:

Year:  2021        PMID: 33929892      PMCID: PMC8474132          DOI: 10.1259/dmfr.20210084

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   3.525


  18 in total

1.  Uncertainty analysis of absorbed dose calculations from thermoluminescence dosimeters.

Authors:  T H Kirby; W F Hanson; D A Johnston
Journal:  Med Phys       Date:  1992 Nov-Dec       Impact factor: 4.071

2.  Effective dose range for dental cone beam computed tomography scanners.

Authors:  Ruben Pauwels; Jilke Beinsberger; Bruno Collaert; Chrysoula Theodorakou; Jessica Rogers; Anne Walker; Lesley Cockmartin; Hilde Bosmans; Reinhilde Jacobs; Ria Bogaerts; Keith Horner
Journal:  Eur J Radiol       Date:  2010-12-31       Impact factor: 3.528

3.  Basic training requirements for the use of dental CBCT by dentists: a position paper prepared by the European Academy of DentoMaxilloFacial Radiology.

Authors:  J Brown; R Jacobs; E Levring Jäghagen; C Lindh; G Baksi; D Schulze; R Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-10-16       Impact factor: 2.419

4.  Radiochromic films for dental CT dosimetry: a feasibility study.

Authors:  O Rampado; S D Bianchi; A Peruzzo Cornetto; V Rossetti; R Ropolo
Journal:  Phys Med       Date:  2012-06-30       Impact factor: 2.685

5.  Validation of a Monte Carlo simulation for dose assessment in dental cone beam CT examinations.

Authors:  J J Morant; M Salvadó; R Casanovas; I Hernández-Girón; E Velasco; A Calzado
Journal:  Phys Med       Date:  2011-07-31       Impact factor: 2.685

6.  TLD measurements and Monte Carlo calculations of head and neck organ and effective doses for cone beam computed tomography using 3D Accuitomo 170.

Authors:  Marina Ernst; Peter Manser; Karl Dula; Werner Volken; Marco Fm Stampanoni; Michael K Fix
Journal:  Dentomaxillofac Radiol       Date:  2017-07-27       Impact factor: 2.419

7.  EFFECTIVE DOSE IN TWO DIFFERENT DENTAL CBCT SYSTEMS: NEWTOM VGi AND PLANMECA 3D MID.

Authors:  Mohammad Ghaedizirgar; Reza Faghihi; Reza Paydar; Sedigheh Sina
Journal:  Radiat Prot Dosimetry       Date:  2017-11-01       Impact factor: 0.972

Review 8.  An analysis of effective dose optimization and its impact on image quality and diagnostic efficacy relating to dental cone beam computed tomography (CBCT)

Authors:  Margarete B McGuigan; Henry F Duncan; Keith Horner
Journal:  Swiss Dent J       Date:  2018-04-16

Review 9.  Cone-beam computed tomography and its applications in dental and maxillofacial radiology.

Authors:  R K W Schulze; N A Drage
Journal:  Clin Radiol       Date:  2020-05-23       Impact factor: 2.350

10.  The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice.

Authors:  Andreas Stratis; Guozhi Zhang; Reinhilde Jacobs; Ria Bogaerts; Hilde Bosmans
Journal:  Eur Radiol       Date:  2019-07-01       Impact factor: 5.315

View more
  1 in total

1.  Distrontium Cerate as a Radiopaque Component of Hydraulic Endodontic Cement.

Authors:  Kunlanun Dumrongvute; Sherif Adel; Takahiro Wada; Nobuyuki Kawashima; Chinalai Piyachon; Hiroshi Watanabe; Tohru Kurabayashi; Takashi Okiji; Motohiro Uo
Journal:  Materials (Basel)       Date:  2021-12-31       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.