Literature DB >> 24732073

Preliminary evaluation of the dosimetric accuracy of the in vivo plastic scintillation detector OARtrac system for prostate cancer treatments.

Slade J Klawikowski1, Clint Zeringue, Landon S Wootton, Geoffrey S Ibbott, Sam Beddar.   

Abstract

A promising, new, in vivo prostate dosimetry system has been developed for clinical radiation therapy. This work outlines the preliminary end-to-end testing of the accuracy and precision of the new OARtrac scintillation dosimetry system. We tested 94 calibrated plastic scintillation detector (PSD) probes before their final integration into endorectal balloon assemblies. These probes had been calibrated at The University of Texas MD Anderson Cancer Center Dosimetry Laboratory. We used a complete clinical OARtrac system including the PSD probes, charge coupled device camera monitoring system, and the manufacturer's integrated software package. The PSD probes were irradiated at 6 MV in a Solid Water® phantom. Irradiations were performed with a 6 MV linear accelerator using anterior-posterior/posterior-anterior matched fields to a maximum dose of 200 cGy in a 100 cm source-axis distance geometry. As a whole, the OARtrac system has good accuracy with a mean error of 0.01% and an error spread of ±5.4% at the 95% confidence interval. These results reflect the PSD probes' accuracy before their final insertion into endorectal balloons. Future work will test the dosimetric effects of mounting the PSD probes within the endorectal balloon assemblies.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24732073      PMCID: PMC4083835          DOI: 10.1088/0031-9155/59/9/N27

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  13 in total

1.  Plastic scintillation dosimetry: optimal selection of scintillating fibers and scintillators.

Authors:  Louis Archambault; Jean Arsenault; Luc Gingras; A Sam Beddar; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

2.  In vivo dosimeters for HDR brachytherapy: a comparison of a diamond detector, MOSFET, TLD, and scintillation detector.

Authors:  Jamil Lambert; Tatsuya Nakano; Sue Law; Justin Elsey; David R McKenzie; Natalka Suchowerska
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

3.  Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry.

Authors:  Louis Archambault; A Sam Beddar; Luc Gingras; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

4.  Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications.

Authors:  Fréderic Lacroix; Louis Archambault; Luc Gingras; Mathieu Guillot; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

5.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

Review 6.  Water equivalent plastic scintillation detectors in radiation therapy.

Authors:  A S Beddar
Journal:  Radiat Prot Dosimetry       Date:  2006-08-01       Impact factor: 0.972

7.  Increased risk of biochemical and local failure in patients with distended rectum on the planning CT for prostate cancer radiotherapy.

Authors:  Renaud de Crevoisier; Susan L Tucker; Lei Dong; Radhe Mohan; Rex Cheung; James D Cox; Deborah A Kuban
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

8.  Real-time in vivo rectal wall dosimetry using plastic scintillation detectors for patients with prostate cancer.

Authors:  Landon Wootton; Rajat Kudchadker; Andrew Lee; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-01-17       Impact factor: 3.609

9.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

10.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

View more
  2 in total

Review 1.  A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy.

Authors:  S O'Keeffe; D McCarthy; P Woulfe; M W D Grattan; A R Hounsell; D Sporea; L Mihai; I Vata; G Leen; E Lewis
Journal:  Br J Radiol       Date:  2015-03-11       Impact factor: 3.039

2.  Real-time in vivo dosimetry for SBRT prostate treatment using plastic scintillating dosimetry embedded in a rectal balloon: a case study.

Authors:  Justin L Cantley; Chee-Wai Cheng; Fredrick B Jesseph; Tarun K Podder; Valdir C Colussi; Bryan J Traughber; Lee E Ponsky; Rodney J Ellis
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.