| Literature DB >> 27929503 |
Justin L Cantley1, Chee-Wai Cheng, Fredrick B Jesseph, Tarun K Podder, Valdir C Colussi, Bryan J Traughber, Lee E Ponsky, Rodney J Ellis.
Abstract
A novel FDA approved in vivo dosimetry device system using plastic scintillating detectors placed in an endorectal balloon to provide real-time in vivo dosimetry for prostatic rectal interface was tested for use with stereotactic body radiotherapy (SBRT). The system was used for the first time ever to measure dose during linear accelerator based SBRT. A single patient was treated with a total dose of 36.25 Gy given in 5 fractions. Delivered dose was measured for each treatment with the detectors placed against the anterior rectal wall near the prostate rectal interface. Measured doses showed varying degrees of agreement with computed/ planned doses, with average combined dose found to be within 6% of the expected dose. The variance between measurements is most likely due to uncertainty of the detector location, as well as variation in the placement of a new balloon prior to each fraction. Distance to agreement for the detectors was generally found to be within a few millimeters, which also suggested that the differences in measured and calculated doses were due to positional uncertainty of the detectors during the SBRT, which had sharp dose falloff near the penumbra along the rectal wall. Overall, the use of a real time in vivo dosimeter provided a level of safety and improved confidence in treatment delivery. We are evaluating the device further in an IRB-approved prospective partial prostate SBRT trial, and believe further clinical investigations are warranted.Entities:
Mesh:
Year: 2016 PMID: 27929503 PMCID: PMC5690496 DOI: 10.1120/jacmp.v17i6.6508
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1OARtrac system: (top left) schematic of the PSD sensors and the fiducial (marker bead); (top right) overhead schematic of the endorectal balloon with the PSD sensor connected. Note that the sensor is attached to the top side of the balloon. Bottom left: PSD sensor outside of the endorectal balloon. Bottom right: Overhead view of the PSD sensor inside of the endorectal balloon and the fiber optic cables that will connect to the CCD camera system. The sensor is attached to the top side of the balloon, which would be placed against the anterior rectal wall.
Measured and expected doses for the proximal detector. All doses are in cGy
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|
| Measured Dose | 417.11 | 603.90 | 425.91 | 291.71 | 420.66 | 2159.29 |
| Pinnacle Dose | 458 | 458 | 458 | 458 | 458 | 2290 |
| % Difference |
|
|
|
|
|
|
| MIM Dose | 531 | 399 | 497 | 395 | 474 | 2296 |
| % Difference |
|
|
|
|
|
|
Measured and expected doses for the distal detector. All doses are in cGy
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|
| Measured Dose | 433.25 | 323.17 | 593.22 | 692.72 | 521.98 | 2564.34 |
| Pinnacle Dose | 456.7 | 456.7 | 456.7 | 456.7 | 456.7 | 2283.5 |
| % Difference |
|
|
|
|
|
|
| MIM Dose | 429 | 407 | 435 | 549 | 457 | 2277 |
| % Difference |
|
|
|
|
|
|
Figure 2Axial view of the prostate with isodose lines for one fraction of the SBRT treatment. Each isodose line represents a change of 50 cGy. Slices moving superior to inferior from top left to bottom right.
Figure 3Sagittal view of the prostate with isodose lines for one fraction of the SBRT treatment. Estimated positions of the proximal and distal detectors can be seen in the central image. Each isodose line represents a change of 50 cGy. Slices moving patient left to patient right from top left to bottom right.
Distance to agreement for the two PSD detectors for each treatment fraction
|
|
|
|
|
| |
|---|---|---|---|---|---|
| DTA – Proximal Detector (mm) | 4.5 | 5.0 | 2.5 | 3.5 | 2.0 |
| DTA – Distal Detector (mm) | 0.6 | 9.0 | 4.5 | 4.0 | 2.5 |