A method to correct for temperature dependence and measure simultaneously dose and temperature using a plastic scintillation detector.

Francois Therriault-Proulx1, Landon Wootton, Sam Beddar.   

Abstract

Plastic scintillation detectors (PSDs) work well for radiation dosimetry. However, they show some temperature dependence, and a priori knowledge of the temperature surrounding the PSD is required to correct for this dependence. We present a novel approach to correct PSD response values for temperature changes instantaneously and without the need for prior knowledge of the temperature value. In addition to rendering the detector temperature-independent, this approach allows for actual temperature measurement using solely the PSD apparatus. With a temperature-controlled water tank, the temperature was varied from room temperature to more than 40 °C and the PSD was used to measure the dose delivered from a cobalt-60 photon beam unit to within an average of 0.72% from the expected value. The temperature was measured during each acquisition with the PSD and a thermocouple and values were within 1 °C of each other. The depth-dose curve of a 6 MV photon beam was also measured under warm non-stable conditions and this curve agreed to within an average of  -0.98% from the curve obtained at room temperature. The feasibility of rendering PSDs temperature-independent was demonstrated with our approach, which also enabled simultaneous measurement of both dose and temperature. This novel approach improves both the robustness and versatility of PSDs.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26407188      PMCID: PMC4613770          DOI: 10.1088/0031-9155/60/20/7927

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  35 in total

1.  Plastic scintillation dosimetry: optimal selection of scintillating fibers and scintillators.

Authors:  Louis Archambault; Jean Arsenault; Luc Gingras; A Sam Beddar; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

2.  Clinical trials of a urethral dose measurement system in brachytherapy using scintillation detectors.

Authors:  Natalka Suchowerska; Michael Jackson; Jamil Lambert; Yong Bai Yin; George Hruby; David R McKenzie
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-11-18       Impact factor: 7.038

3.  Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry.

Authors:  Louis Archambault; A Sam Beddar; Luc Gingras; René Roy; Luc Beaulieu
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

4.  Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide.

Authors:  J Lambert; Y Yin; D R McKenzie; S Law; N Suchowerska
Journal:  Phys Med Biol       Date:  2008-05-19       Impact factor: 3.609

5.  Technical note: removing the stem effect when performing Ir-192 HDR brachytherapy in vivo dosimetry using plastic scintillation detectors: a relevant and necessary step.

Authors:  Francois Therriault-Proulx; Sam Beddar; Tina M Briere; Louis Archambault; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

6.  Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations.

Authors:  Mathieu Guillot; Luc Gingras; Louis Archambault; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

7.  On possible temperature dependence of plastic scintillator response.

Authors:  Sam Beddar
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

8.  A mathematical formalism for hyperspectral, multipoint plastic scintillation detectors.

Authors:  Louis Archambault; François Therriault-Proulx; Sam Beddar; Luc Beaulieu
Journal:  Phys Med Biol       Date:  2012-10-12       Impact factor: 3.609

9.  Toward a real-time in vivo dosimetry system using plastic scintillation detectors.

Authors:  Louis Archambault; Tina M Briere; Falk Pönisch; Luc Beaulieu; Deborah A Kuban; Andrew Lee; Sam Beddar
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-16       Impact factor: 7.038

10.  Spectral discrimination of Cerenkov radiation in scintillating dosimeters.

Authors:  A M Frelin; J M Fontbonne; G Ban; J Colin; M Labalme; A Batalla; A Isambert; A Vela; T Leroux
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

View more
  1 in total

1.  Pre-clinical and clinical evaluation of the HYPERSCINT plastic scintillation dosimetry research platform for in vivo dosimetry during radiotherapy.

Authors:  Imke Schoepper; Sonja Dieterich; Earl Alonzo Trestrail; Michael Sean Kent
Journal:  J Appl Clin Med Phys       Date:  2022-02-21       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.