Francois Therriault-Proulx1, Landon Wootton, Sam Beddar.
Abstract
Plastic scintillation detectors (PSDs) work well for radiation dosimetry. However, they show some temperature dependence, and a priori knowledge of the temperature surrounding the PSD is required to correct for this dependence. We present a novel approach to correct PSD response values for temperature changes instantaneously and without the need for prior knowledge of the temperature value. In addition to rendering the detector temperature-independent, this approach allows for actual temperature measurement using solely the PSD apparatus. With a temperature-controlled water tank, the temperature was varied from room temperature to more than 40 °C and the PSD was used to measure the dose delivered from a cobalt-60 photon beam unit to within an average of 0.72% from the expected value. The temperature was measured during each acquisition with the PSD and a thermocouple and values were within 1 °C of each other. The depth-dose curve of a 6 MV photon beam was also measured under warm non-stable conditions and this curve agreed to within an average of -0.98% from the curve obtained at room temperature. The feasibility of rendering PSDs temperature-independent was demonstrated with our approach, which also enabled simultaneous measurement of both dose and temperature. This novel approach improves both the robustness and versatility of PSDs.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26407188 PMCID: PMC4613770 DOI: 10.1088/0031-9155/60/20/7927
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609