Literature DB >> 23289127

Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty.

H John Cooper1, Craig J Della Valle, Richard A Berger, Matthew Tetreault, Wayne G Paprosky, Scott M Sporer, Joshua J Jacobs.   

Abstract

BACKGROUND: Corrosion at the modular head-neck junction of the femoral component in total hip arthroplasty has been identified as a potential concern, although symptomatic adverse local tissue reactions secondary to corrosion have rarely been described.
METHODS: We retrospectively reviewed the records of ten patients with a metal-on-polyethylene total hip prosthesis, from three different manufacturers, who underwent revision surgery for corrosion at the modular head-neck junction.
RESULTS: All patients presented with pain or swelling around the hip, and two patients presented with recurrent instability. Serum cobalt levels were elevated prior to the revision arthroplasty and were typically more elevated than were serum chromium levels. Surgical findings included large soft-tissue masses and surrounding tissue damage with visible corrosion at the femoral head-neck junction; the two patients who presented with instability had severe damage to the hip abductor musculature. Pathology specimens consistently demonstrated areas of tissue necrosis. The patients were treated with debridement and a femoral head and liner exchange, with use of a ceramic femoral head with a titanium sleeve in eight cases. The mean Harris hip score improved from 58.1 points preoperatively to 89.7 points at a mean of 13.0 months after the revision surgery (p=0.01). Repeat serum cobalt levels, measured in six patients at a mean of 8.0 months following revision, decreased to a mean of 1.61 ng/mL, and chromium levels were similar to prerevision levels. One patient with moderate hip abductor muscle necrosis developed recurrent instability after revision and required a second revision arthroplasty.
CONCLUSIONS: Adverse local tissue reactions can occur in patients with a metal-on-polyethylene bearing secondary to corrosion at the modular femoral head-neck taper, and their presentation is similar to the adverse local tissue reactions seen in patients with a metal-on-metal bearing. Elevated serum metal levels, particularly a differential elevation of serum cobalt levels with respect to chromium levels, can be helpful in establishing this diagnosis.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23289127      PMCID: PMC3444948          DOI: 10.2106/jbjs.k.01352

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  39 in total

1.  Dislocation after polyethylene liner exchange in total hip arthroplasty.

Authors:  Henry R Boucher; Chris Lynch; Anthony M Young; C Anderson Engh; Charles Engh
Journal:  J Arthroplasty       Date:  2003-08       Impact factor: 4.757

2.  Corrosion of modular hip prostheses.

Authors:  E B Mathiesen; J U Lindgren; G G Blomgren; F P Reinholt
Journal:  J Bone Joint Surg Br       Date:  1991-07

3.  Corrosion between the components of modular femoral hip prostheses.

Authors:  J P Collier; V A Surprenant; R E Jensen; M B Mayor; H P Surprenant
Journal:  J Bone Joint Surg Br       Date:  1992-07

4.  Corrosion at the interface of cobalt-alloy heads on titanium-alloy stems.

Authors:  J P Collier; V A Surprenant; R E Jensen; M B Mayor
Journal:  Clin Orthop Relat Res       Date:  1991-10       Impact factor: 4.176

5.  Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty. A case report.

Authors:  O Svensson; E B Mathiesen; F P Reinholt; G Blomgren
Journal:  J Bone Joint Surg Am       Date:  1988-09       Impact factor: 5.284

6.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation.

Authors:  W H Harris
Journal:  J Bone Joint Surg Am       Date:  1969-06       Impact factor: 5.284

7.  Corrosion and wear at the modular interface of uncemented femoral stems.

Authors:  S D Cook; R L Barrack; A J Clemow
Journal:  J Bone Joint Surg Br       Date:  1994-01

8.  Early morbidity of modular exchange for polyethylene wear and osteolysis.

Authors:  William L Griffin; Thomas K Fehring; J Bohannon Mason; Thomas H McCoy; Susan Odum; Christi Sychterz Terefenko
Journal:  J Arthroplasty       Date:  2004-10       Impact factor: 4.757

9.  Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants.

Authors:  J L Gilbert; C A Buckley; J J Jacobs; K C Bertin; M R Zernich
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

10.  Investigations on the galvanic corrosion of multialloy total hip prostheses.

Authors:  L C Lucas; R A Buchanan; J E Lemons
Journal:  J Biomed Mater Res       Date:  1981-09
View more
  137 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  New technology: safety, efficacy, and learning curves.

Authors:  Paul J Duwelius; Javad Parvizi; Laura Matsen Ko
Journal:  Clin Orthop Relat Res       Date:  2013-12-07       Impact factor: 4.176

3.  CORR Insights®: Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem.

Authors:  Michael G Zywiel
Journal:  Clin Orthop Relat Res       Date:  2013-11-27       Impact factor: 4.176

4.  Taper design affects failure of large-head metal-on-metal total hip replacements.

Authors:  Nader A Nassif; Danyal H Nawabi; Kirsten Stoner; Marcella Elpers; Timothy Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

Review 5.  The surgical options and clinical evidence for treatment of wear or corrosion occurring with THA or TKA.

Authors:  Charles A Engh; Henry Ho; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

6.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

Review 7.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 8.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

9.  CORR Insights®: do serologic and synovial tests help diagnose infection in revision hip arthroplasty with metal-on-metal bearings or corrosion?

Authors:  Carl Deirmengian
Journal:  Clin Orthop Relat Res       Date:  2015-01-09       Impact factor: 4.176

10.  Diagnosis and Management of Adverse Reactions to Metal Debris.

Authors:  Richard A Wawrose; Kenneth L Urish
Journal:  Oper Tech Orthop       Date:  2019-10-01
View more