| Literature DB >> 1688333 |
J P Collier1, V A Surprenant, R E Jensen, M B Mayor.
Abstract
The combination of a cobalt-alloy head on a titanium-alloy femoral hip stem is widely accepted for press-fit and biologic fixation applications. Examination of 30 components retrieved at periods of 0.5 to 66.9 months for histologic examination of tissue ingrowth revealed that 56.6% of the tapered connections between head and stem showed evidence of crevice corrosion leading to concerns of metal ion release and the potential failure of head to stem fixation.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1688333
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176