Literature DB >> 2071637

Corrosion of modular hip prostheses.

E B Mathiesen1, J U Lindgren, G G Blomgren, F P Reinholt.   

Abstract

Nine uncemented hip prostheses of modular design were revised because of late infection (2 cases), femoral stem loosening or fracture (2), loosening of threaded cups (3) and protrusion of bipolar cups (2). At surgery we found tissue discoloration and macroscopic corrosion in four of the nine prostheses, located at the head-neck junction. Histological examination in most cases showed extensive necrosis associated with metal particles, indicating metal toxicity. Metallographic examination of the prosthetic heads revealed structural imperfections and inhomogeneity of the metal. We suggest that the crevice between the head and neck is a potential site of corrosion in modular designs.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 2071637     DOI: 10.1302/0301-620X.73B4.2071637

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  14 in total

1.  Factors to consider in joint prosthesis systems.

Authors:  Larry M Wolford
Journal:  Proc (Bayl Univ Med Cent)       Date:  2006-07

Review 2.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 3.  Metal-on-metal total hip arthroplasty: risk factors for pseudotumours and clinical systematic evaluation.

Authors:  Ming Han Lincoln Liow; Young-Min Kwon
Journal:  Int Orthop       Date:  2016-10-20       Impact factor: 3.075

4.  Otto Aufranc Award: Large Heads Do Not Increase Damage at the Head-neck Taper of Metal-on-polyethylene Total Hip Arthroplasties.

Authors:  Georgios K Triantafyllopoulos; Marcella E Elpers; Jayme C Burket; Christina I Esposito; Douglas E Padgett; Timothy M Wright
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

5.  Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck.

Authors:  H John Cooper; Robert M Urban; Richard L Wixson; R Michael Meneghini; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2013-05-15       Impact factor: 5.284

6.  Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty.

Authors:  H John Cooper; Craig J Della Valle; Richard A Berger; Matthew Tetreault; Wayne G Paprosky; Scott M Sporer; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2012-09-19       Impact factor: 5.284

7.  Titanium alloy stem as a cause for adverse reaction to metal debris after bipolar hemiarthroplasty.

Authors:  Masaaki Sakamoto; Hitoshi Watanabe; Hidetaka Higashi; Hitoshi Kubosawa
Journal:  Case Rep Orthop       Date:  2014-03-04

8.  The innovation trap.

Authors:  Nils P Hailer
Journal:  Acta Orthop       Date:  2016-02-06       Impact factor: 3.717

9.  Fretting and Corrosion in Modular Shoulder Arthroplasty: A Retrieval Analysis.

Authors:  Johannes A Eckert; Ulrike Mueller; Sebastian Jaeger; Benjamin Panzram; J Philippe Kretzer
Journal:  Biomed Res Int       Date:  2016-06-28       Impact factor: 3.411

10.  A case of a C-stem fracture at the head-neck junction and a review of the literature.

Authors:  David Morley; Ian Starks; Justin Lim
Journal:  Case Rep Orthop       Date:  2012-12-09