Literature DB >> 8288652

Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants.

J L Gilbert1, C A Buckley, J J Jacobs, K C Bertin, M R Zernich.   

Abstract

Two modular hip implants with a cobalt-alloy head and a cobalt-alloy stem were retrieved after a fracture had occurred in the neck region of the femoral component, eighty-five and seventy months after implantation. Both implants failed less than one millimeter distal to the taper junction between the head and the stem (outside of the taper). The fracture surfaces of the implant were investigated with the use of scanning electron microscopy, to determine the nature of the failure process. The fractures occurred at the grain boundaries of the microstructure and appeared to be the result of three factors: porosity at the grain boundaries; intergranular corrosive attack, initiated both at the head-neck taper and at the free surface; and cyclic fatigue-loading of the stem. The corrosive attack of the free surface was initiated, in part, by the egression of surface grains and by the ingression of fluid into the intergranular regions. Sectioned surfaces showed extensive intergranular corrosive attack in the prosthetic neck localized in the region of the head-neck taper junction and penetrating deeply into the microstructure.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8288652     DOI: 10.2106/00004623-199401000-00014

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  30 in total

1.  Fracture of a modular femoral neck after total hip arthroplasty: a case report.

Authors:  Geoffrey Wright; Scott Sporer; Robert Urban; Joshua Jacobs
Journal:  J Bone Joint Surg Am       Date:  2010-06       Impact factor: 5.284

Review 2.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

3.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

Review 4.  Corrosion of Metallic Biomaterials: A Review.

Authors:  Noam Eliaz
Journal:  Materials (Basel)       Date:  2019-01-28       Impact factor: 3.623

5.  Modern trunnions are more flexible: a mechanical analysis of THA taper designs.

Authors:  David A Porter; Robert M Urban; Joshua J Jacobs; Jeremy L Gilbert; José A Rodriguez; H John Cooper
Journal:  Clin Orthop Relat Res       Date:  2014-09-30       Impact factor: 4.176

Review 6.  The biological response to orthopaedic implants for joint replacement: Part I: Metals.

Authors:  Emmanuel Gibon; Derek F Amanatullah; Florence Loi; Jukka Pajarinen; Akira Nabeshima; Zhenyu Yao; Moussa Hamadouche; Stuart B Goodman
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-06-21       Impact factor: 3.368

7.  Severe metal-induced osteolysis many years after unipolar hip endoprosthesis.

Authors:  Matthew A Mann; Dylan Tanzer; Michael Tanzer
Journal:  Clin Orthop Relat Res       Date:  2013-01-05       Impact factor: 4.176

8.  Case reports: fractures of threaded cups: rare complications of a well-established implant.

Authors:  Arnd T Hoburg; Manav Mehta; Stephan Tohtz; Carsten Perka
Journal:  Clin Orthop Relat Res       Date:  2008-09-27       Impact factor: 4.176

9.  Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty.

Authors:  H John Cooper; Craig J Della Valle; Richard A Berger; Matthew Tetreault; Wayne G Paprosky; Scott M Sporer; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2012-09-19       Impact factor: 5.284

10.  Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?

Authors:  Robin Pourzal; Deborah J Hall; Nguyen Q Ha; Robert M Urban; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

View more