Dosimetry of a cone beam CT device for oral and maxillofacial radiology using Monte Carlo techniques and ICRP adult reference computational phantoms.

J J Morant1, M Salvadó, I Hernández-Girón, R Casanovas, R Ortega, A Calzado.   

Abstract

OBJECTIVES: The aim of this study was to calculate organ and effective doses for a range of available protocols in a particular cone beam CT (CBCT) scanner dedicated to dentistry and to derive effective dose conversion factors.
METHODS: Monte Carlo simulations were used to calculate organ and effective doses using the International Commission on Radiological Protection voxel adult male and female reference phantoms (AM and AF) in an i-CAT CBCT. Nine different fields of view (FOVs) were simulated considering full- and half-rotation modes, and also a high-resolution acquisition for a particular protocol. Dose-area product (DAP) was measured.
RESULTS: Dose to organs varied for the different FOVs, usually being higher in the AF phantom. For 360°, effective doses were in the range of 25-66 μSv, and 46 μSv for full head. Higher contributions to the effective dose corresponded to the remainder (31%; 27-36 range), salivary glands (23%; 20-29%), thyroid (13%; 8-17%), red bone marrow (10%; 9-11%) and oesophagus (7%; 4-10%). The high-resolution protocol doubled the standard resolution doses. DAP values were between 181 mGy cm(2) and 556 mGy cm(2) for 360°. For 180° protocols, dose to organs, effective dose and DAP were approximately 40% lower. A conversion factor (DAP to effective dose) of 0.130 ± 0.006 μSv mGy(-1) cm(-2) was derived for all the protocols, excluding full head. A wide variation in dose to eye lens and thyroid was found when shifting the FOV in the AF phantom.
CONCLUSIONS: Organ and effective doses varied according to field size, acquisition angle and positioning of the beam relative to radiosensitive organs. Good positive correlation between calculated effective dose and measured DAP was found.

Entities:  

Mesh:

Year:  2012        PMID: 22933532      PMCID: PMC3667542          DOI: 10.1259/dmfr/92555893

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  19 in total

1.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

2.  An assessment of bone marrow and bone endosteum dosimetry methods for photon sources.

Authors:  Choonik Lee; Choonsik Lee; Amish P Shah; Wesley E Bolch
Journal:  Phys Med Biol       Date:  2006-10-06       Impact factor: 3.609

3.  Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD.

Authors:  S Lofthag-Hansen; A Thilander-Klang; A Ekestubbe; E Helmrot; K Gröndahl
Journal:  Dentomaxillofac Radiol       Date:  2008-02       Impact factor: 2.419

4.  Effective dose from cone beam CT examinations in dentistry.

Authors:  J A Roberts; N A Drage; J Davies; D W Thomas
Journal:  Br J Radiol       Date:  2008-10-13       Impact factor: 3.039

5.  Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

Authors:  John B Ludlow; Marija Ivanovic
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-05-27

6.  Validation of a Monte Carlo simulation for dose assessment in dental cone beam CT examinations.

Authors:  J J Morant; M Salvadó; R Casanovas; I Hernández-Girón; E Velasco; A Calzado
Journal:  Phys Med       Date:  2011-07-31       Impact factor: 2.685

7.  Photoelectron enhancement of the absorbed dose from X rays to human bone marrow: experimental and theoretical studies.

Authors:  S D King; F W Spiers
Journal:  Br J Radiol       Date:  1985-04       Impact factor: 3.039

8.  Cone-beam computed tomography for routine orthodontic treatment planning: a radiation dose evaluation.

Authors:  Maria Alves Garcia Silva; Ulrich Wolf; Frank Heinicke; Axel Bumann; Heiko Visser; Edgar Hirsch
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-05       Impact factor: 2.650

9.  Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation.

Authors:  John B Ludlow; Laura E Davies-Ludlow; Stuart C White
Journal:  J Am Dent Assoc       Date:  2008-09       Impact factor: 3.634

10.  Effective dose: how should it be applied to medical exposures?

Authors:  C J Martin
Journal:  Br J Radiol       Date:  2007-07-23       Impact factor: 3.039

View more
  15 in total

1.  Dosimetric assessment of the exposure of radiotherapy patients due to cone-beam CT procedures.

Authors:  Mariana Baptista; Salvatore Di Maria; Sandra Vieira; Joana Santos; Joana Pereira; Miguel Pereira; Pedro Vaz
Journal:  Radiat Environ Biophys       Date:  2018-11-03       Impact factor: 1.925

2.  Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields.

Authors:  Anna Kelaranta; Marja Ekholm; Paula Toroi; Mika Kortesniemi
Journal:  Dentomaxillofac Radiol       Date:  2015-08-27       Impact factor: 2.419

3.  TLD measurements and Monte Carlo calculations of head and neck organ and effective doses for cone beam computed tomography using 3D Accuitomo 170.

Authors:  Marina Ernst; Peter Manser; Karl Dula; Werner Volken; Marco Fm Stampanoni; Michael K Fix
Journal:  Dentomaxillofac Radiol       Date:  2017-07-27       Impact factor: 2.419

4.  Evaluation of the diagnostic efficacy of two cone beam computed tomography protocols in reliably detecting the location of the inferior alveolar nerve canal.

Authors:  Aditya Tadinada; Sydney Schneider; Sumit Yadav
Journal:  Dentomaxillofac Radiol       Date:  2017-03-13       Impact factor: 2.419

5.  Head and neck effective dose and quantitative assessment of image quality: a study to compare cone beam CT and multislice spiral CT.

Authors:  Cosimo Nardi; Cinzia Talamonti; Stefania Pallotta; Paola Saletti; Linda Calistri; Cesare Cordopatri; Stefano Colagrande
Journal:  Dentomaxillofac Radiol       Date:  2017-06-20       Impact factor: 2.419

6.  Effective dose span of ten different cone beam CT devices.

Authors:  D Rottke; S Patzelt; P Poxleitner; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-04-12       Impact factor: 2.419

7.  Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation.

Authors:  R Pauwels; G Zhang; C Theodorakou; A Walker; H Bosmans; R Jacobs; R Bogaerts; K Horner
Journal:  Br J Radiol       Date:  2014-10       Impact factor: 3.039

Review 8.  Technical aspects of dental CBCT: state of the art.

Authors:  R Pauwels; K Araki; J H Siewerdsen; S S Thongvigitmanee
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

9.  Effective dose reduction using collimation function in digital panoramic radiography and possible clinical implications in dentistry.

Authors:  Daniel Benchimol; Juha Koivisto; Nils Kadesjö; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2018-05-31       Impact factor: 2.419

Review 10.  Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units.

Authors:  J B Ludlow; R Timothy; C Walker; R Hunter; E Benavides; D B Samuelson; M J Scheske
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.