D A Chesler, S J Riederer, N J Pelc.
Abstract
A general expression is derived for the noise due to photon counting statistics in computed X-ray tomography. The variance is inversely proportional to the cube of the resolution distance. For scanners using a water box, the noise in the reconstructed image depends inversely on the number of detected primary photons, summed over all angles, that have passed through a resolution element. Predictions of this formula agree well with the results of computer simulations. It is shown how this formula can be used to determine such parameters as required X-ray flux, detector counting rate, and dose, with special emphasis on tradeoffs between these parameters and resolution. It is also shown that to determine the X-ray attenuation coefficient of a resolution element to a given precision, the number of photons required by computed X-ray tomography is close to a theoretical limit.Entities:
Mesh:
Year: 1977 PMID: 615896 DOI: 10.1097/00004728-197701000-00009
Source DB: PubMed Journal: J Comput Assist Tomogr ISSN: 0363-8715 Impact factor: 1.826