Image-based spectral distortion correction for photon-counting x-ray detectors.

Huanjun Ding1, Sabee Molloi.   

Abstract

PURPOSE: To investigate the feasibility of using an image-based method to correct for distortions induced by various artifacts in the x-ray spectrum recorded with photon-counting detectors for their application in breast computed tomography (CT).
METHODS: The polyenergetic incident spectrum was simulated with the tungsten anode spectral model using the interpolating polynomials (TASMIP) code and carefully calibrated to match the x-ray tube in this study. Experiments were performed on a Cadmium-Zinc-Telluride (CZT) photon-counting detector with five energy thresholds. Energy bins were adjusted to evenly distribute the recorded counts above the noise floor. BR12 phantoms of various thicknesses were used for calibration. A nonlinear function was selected to fit the count correlation between the simulated and the measured spectra in the calibration process. To evaluate the proposed spectral distortion correction method, an empirical fitting derived from the calibration process was applied on the raw images recorded for polymethyl methacrylate (PMMA) phantoms of 8.7, 48.8, and 100.0 mm. Both the corrected counts and the effective attenuation coefficient were compared to the simulated values for each of the five energy bins. The feasibility of applying the proposed method to quantitative material decomposition was tested using a dual-energy imaging technique with a three-material phantom that consisted of water, lipid, and protein. The performance of the spectral distortion correction method was quantified using the relative root-mean-square (RMS) error with respect to the expected values from simulations or areal analysis of the decomposition phantom.
RESULTS: The implementation of the proposed method reduced the relative RMS error of the output counts in the five energy bins with respect to the simulated incident counts from 23.0%, 33.0%, and 54.0% to 1.2%, 1.8%, and 7.7% for 8.7, 48.8, and 100.0 mm PMMA phantoms, respectively. The accuracy of the effective attenuation coefficient of PMMA estimate was also improved with the proposed spectral distortion correction. Finally, the relative RMS error of water, lipid, and protein decompositions in dual-energy imaging was significantly reduced from 53.4% to 6.8% after correction was applied.
CONCLUSIONS: The study demonstrated that dramatic distortions in the recorded raw image yielded from a photon-counting detector could be expected, which presents great challenges for applying the quantitative material decomposition method in spectral CT. The proposed semi-empirical correction method can effectively reduce these errors caused by various artifacts, including pulse pileup and charge sharing effects. Furthermore, rather than detector-specific simulation packages, the method requires a relatively simple calibration process and knowledge about the incident spectrum. Therefore, it may be used as a generalized procedure for the spectral distortion correction of different photon-counting detectors in clinical breast CT systems.

Entities:  

Mesh:

Year:  2012        PMID: 22482608      PMCID: PMC3316693          DOI: 10.1118/1.3693056

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  CT energy weighting in the presence of scatter and limited energy resolution.

Authors:  Taly Gilat Schmidt
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

Authors:  E Roessl; R Proksa
Journal:  Phys Med Biol       Date:  2007-07-17       Impact factor: 3.609

3.  Computed tomography with energy-resolved detection: a feasibility study.

Authors:  Polad M Shikhaliev
Journal:  Phys Med Biol       Date:  2008-02-19       Impact factor: 3.609

4.  An accurate method for direct dual-energy calibration and decomposition.

Authors:  H N Cardinal; A Fenster
Journal:  Med Phys       Date:  1990 May-Jun       Impact factor: 4.071

5.  Optimal "image-based" weighting for energy-resolved CT.

Authors:  Taly Gilat Schmidt
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

6.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

7.  Breast composition measurement with a cadmium-zinc-telluride based spectral computed tomography system.

Authors:  Huanjun Ding; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

8.  Comparison of two portable solid state detectors with an improved collimation and alignment device for mammographic x-ray spectroscopy.

Authors:  U Bottigli; B Golosio; G L Masala; P Oliva; S Stumbo; P Delogu; M E Fantacci; L Abbene; F Fauci; G Raso
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

9.  Compositional breast imaging using a dual-energy mammography protocol.

Authors:  Aurelie D Laidevant; Serghei Malkov; Chris I Flowers; Karla Kerlikowske; John A Shepherd
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

10.  CdZnTe detector in diagnostic x-ray spectroscopy.

Authors:  Satoshi Miyajima; Kotaro Imagawa; Masao Matsumoto
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

View more
  12 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications.

Authors:  Xuejin Liu; Mats Persson; Hans Bornefalk; Staffan Karlsson; Cheng Xu; Mats Danielsson; Ben Huber
Journal:  J Med Imaging (Bellingham)       Date:  2015-09-11

Review 3.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

4.  Quantitative material characterization from multi-energy photon counting CT.

Authors:  Adam M Alessio; Lawrence R MacDonald
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

5.  Tight-frame based iterative image reconstruction for spectral breast CT.

Authors:  Bo Zhao; Hao Gao; Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  Spectral CT using multiple balanced K-edge filters.

Authors:  Yothin Rakvongthai; William Worstell; Georges El Fakhri; Junguo Bian; Auranuch Lorsakul; Jinsong Ouyang
Journal:  IEEE Trans Med Imaging       Date:  2014-09-19       Impact factor: 10.048

7.  A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT.

Authors:  Jochen Cammin; Jennifer Xu; William C Barber; Jan S Iwanczyk; Neal E Hartsough; Katsuyuki Taguchi
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

8.  Breast tissue decomposition with spectral distortion correction: a postmortem study.

Authors:  Huanjun Ding; Bo Zhao; Pavlo Baturin; Farnaz Behroozi; Sabee Molloi
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

9.  Evaluation of a photon counting Medipix3RX cadmium zinc telluride spectral x-ray detector.

Authors:  Jeffrey F Marsh; Steven M Jorgensen; David S Rundle; Andrew J Vercnocke; Shuai Leng; Philip H Butler; Cynthia H McCollough; Erik L Ritman
Journal:  J Med Imaging (Bellingham)       Date:  2018-12-04

10.  Breast tissue characterization with photon-counting spectral CT imaging: a postmortem breast study.

Authors:  Huanjun Ding; Michael J Klopfer; Justin L Ducote; Fumitaro Masaki; Sabee Molloi
Journal:  Radiology       Date:  2014-05-07       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.