Literature DB >> 24694136

A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT.

Jochen Cammin1, Jennifer Xu2, William C Barber3, Jan S Iwanczyk3, Neal E Hartsough3, Katsuyuki Taguchi1.   

Abstract

PURPOSE: Energy discriminating, photon-counting detectors (PCDs) are an emerging technology for computed tomography (CT) with various potential benefits for clinical CT. The photon energies measured by PCDs can be distorted due to the interactions of a photon with the detector and the interaction of multiple coincident photons. These effects result in distorted recorded x-ray spectra which may lead to artifacts in reconstructed CT images and inaccuracies in tissue identification. Model-based compensation techniques have the potential to account for the distortion effects. This approach requires only a small number of parameters and is applicable to a wide range of spectra and count rates, but it needs an accurate model of the spectral distortions occurring in PCDs. The purpose of this study was to develop a model of those spectral distortions and to evaluate the model using a PCD (model DXMCT-1; DxRay, Inc., Northridge, CA) and various x-ray spectra in a wide range of count rates.
METHODS: The authors hypothesize that the complex phenomena of spectral distortions can be modeled by: (1) separating them into count-rate independent factors that we call the spectral response effects (SRE), and count-rate dependent factors that we call the pulse pileup effects (PPE), (2) developing separate models for SRE and PPE, and (3) cascading the SRE and PPE models into a combined SRE+PPE model that describes PCD distortions at both low and high count rates. The SRE model describes the probability distribution of the recorded spectrum, with a photo peak and a continuum tail, given the incident photon energy. Model parameters were obtained from calibration measurements with three radioisotopes and then interpolated linearly for other energies. The PPE model used was developed in the authors' previous work [K. Taguchi et al., "Modeling the performance of a photon counting x-ray detector for CT: Energy response and pulse pileup effects," Med. Phys. 38(2), 1089-1102 (2011)]. The agreement between the x-ray spectra calculated by the cascaded SRE+PPE model and the measured spectra was evaluated for various levels of deadtime loss ratios (DLR) and incident spectral shapes, realized using different attenuators, in terms of the weighted coefficient of variation (COVW), i.e., the root mean square difference weighted by the statistical errors of the data and divided by the mean.
RESULTS: At low count rates, when DLR < 10%, the distorted spectra measured by the DXMCT-1 were in agreement with those calculated by SRE only, with COVW's less than 4%. At higher count rates, the measured spectra were also in agreement with the ones calculated by the cascaded SRE+PPE model; with PMMA as attenuator, COVW was 5.6% at a DLR of 22% and as small as 6.7% for a DLR as high as 55%.
CONCLUSIONS: The x-ray spectra calculated by the proposed model agreed with the measured spectra over a wide range of count rates and spectral shapes. The SRE model predicted the distorted, recorded spectra with low count rates over various types and thicknesses of attenuators. The study also validated the hypothesis that the complex spectral distortions in a PCD can be adequately modeled by cascading the count-rate independent SRE and the count-rate dependent PPE.
© 2014 American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2014        PMID: 24694136      PMCID: PMC3979165          DOI: 10.1118/1.4866890

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors.

Authors:  Katsuyuki Taguchi; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; William C Barber
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Computed tomography in color: NanoK-enhanced spectral CT molecular imaging.

Authors:  Dipanjan Pan; Ewald Roessl; Jens-Peter Schlomka; Shelton D Caruthers; Angana Senpan; Mike J Scott; John S Allen; Huiying Zhang; Grace Hu; Patrick J Gaffney; Eric T Choi; Volker Rasche; Samuel A Wickline; Roland Proksa; Gregory M Lanza
Journal:  Angew Chem Int Ed Engl       Date:  2010-12-10       Impact factor: 15.336

3.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography.

Authors:  J P Schlomka; E Roessl; R Dorscheid; S Dill; G Martens; T Istel; C Bäumer; C Herrmann; R Steadman; G Zeitler; A Livne; R Proksa
Journal:  Phys Med Biol       Date:  2008-07-08       Impact factor: 3.609

4.  Signal formation and decay in CdTe x-ray detectors under intense irradiation.

Authors:  A Jahnke; R Matz
Journal:  Med Phys       Date:  1999-01       Impact factor: 4.071

5.  Energy-selective reconstructions in X-ray computerized tomography.

Authors:  R E Alvarez; A Macovski
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

6.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

7.  Modeling the performance of a photon counting x-ray detector for CT: energy response and pulse pileup effects.

Authors:  Katsuyuki Taguchi; Mengxi Zhang; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; Einar Nygard; Neal E Hartsough; Benjamin M W Tsui; William C Barber
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

8.  Image-based spectral distortion correction for photon-counting x-ray detectors.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

9.  Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging.

Authors:  Jan S Iwanczyk; Einar Nygård; Oded Meirav; Jerry Arenson; William C Barber; Neal E Hartsough; Nail Malakhov; Jan C Wessel
Journal:  IEEE Trans Nucl Sci       Date:  2009       Impact factor: 1.679

10.  Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical Applications.

Authors:  Stefano Del Sordo; Leonardo Abbene; Ezio Caroli; Anna Maria Mancini; Andrea Zappettini; Pietro Ubertini
Journal:  Sensors (Basel)       Date:  2009-05-12       Impact factor: 3.576

View more
  16 in total

1.  Emerging Breast Imaging Technologies on the Horizon.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  Semin Ultrasound CT MR       Date:  2017-09-13       Impact factor: 1.875

Review 2.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

3.  The effects of extending the spectral information acquired by a photon-counting detector for spectral CT.

Authors:  Taly Gilat Schmidt; Kevin C Zimmerman; Emil Y Sidky
Journal:  Phys Med Biol       Date:  2015-01-23       Impact factor: 3.609

4.  Estimation of signal and noise for a whole-body research photon-counting CT system.

Authors:  Zhoubo Li; Shuai Leng; Zhicong Yu; Steffen Kappler; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-22

5.  Experimental comparison of empirical material decomposition methods for spectral CT.

Authors:  Kevin C Zimmerman; Taly Gilat Schmidt
Journal:  Phys Med Biol       Date:  2015-03-27       Impact factor: 3.609

6.  Evaluation of models of spectral distortions in photon-counting detectors for computed tomography.

Authors:  Jochen Cammin; Steffen Kappler; Thomas Weidinger; Katsuyuki Taguchi
Journal:  J Med Imaging (Bellingham)       Date:  2016-05-06

Review 7.  Energy-sensitive photon counting detector-based X-ray computed tomography.

Authors:  Katsuyuki Taguchi
Journal:  Radiol Phys Technol       Date:  2017-01-30

8.  Pulse pileup analysis for a double-sided silicon strip detector using variable pulse shapes.

Authors:  Jinghui Wang; Linchuan Chen; Mats Persson; Paurakh L Rajbhandary; Praneeth Kandlakunta; Gabriella Carini; Rebecca Fahrig
Journal:  IEEE Trans Nucl Sci       Date:  2019-05-15       Impact factor: 1.679

9.  A Spectral CT Method to Directly Estimate Basis Material Maps From Experimental Photon-Counting Data.

Authors:  Taly Gilat Schmidt; Rina Foygel Barber; Emil Y Sidky
Journal:  IEEE Trans Med Imaging       Date:  2017-04-24       Impact factor: 10.048

10.  A neural network-based method for spectral distortion correction in photon counting x-ray CT.

Authors:  Mengheng Touch; Darin P Clark; William Barber; Cristian T Badea
Journal:  Phys Med Biol       Date:  2016-07-29       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.