Computed tomography with energy-resolved detection: a feasibility study.

Polad M Shikhaliev1.   

Abstract

The feasibility of computed tomography (CT) with energy-resolved x-ray detection has been investigated. A breast CT design with multi slit multi slice (MSMS) data acquisition was used for this study. The MSMS CT includes linear arrays of photon counting detectors separated by gaps. This CT configuration allows for efficient scatter rejection and 3D data acquisition. The energy-resolved CT images were simulated using a digital breast phantom and the design parameters of the proposed MSMS CT. The phantom had 14 cm diameter and 50/50 adipose/glandular composition, and included carcinoma, adipose, blood, iodine and CaCO3 as contrast elements. The x-ray technique was 90 kVp tube voltage with 660 mR skin exposure. Photon counting, charge (energy) integrating and photon energy weighting CT images were generated. The contrast-to-noise (CNR) improvement with photon energy weighting was quantified. The dual energy subtracted images of CaCO3 and iodine were generated using a single CT scan at a fixed x-ray tube voltage. The x-ray spectrum was electronically split into low- and high-energy parts by a photon counting detector. The CNR of the energy weighting CT images of carcinoma, blood, adipose, iodine, and CaCO3 was higher by a factor of 1.16, 1.20, 1.21, 1.36 and 1.35, respectively, as compared to CT with a conventional charge (energy) integrating detector. Photon energy weighting was applied to CT projections prior to dual energy subtraction and reconstruction. Photon energy weighting improved the CNR in dual energy subtracted CT images of CaCO3 and iodine by a factor of 1.35 and 1.33, respectively. The combination of CNR improvements due to scatter rejection and energy weighting was in the range of 1.71-2 depending on the type of the contrast element. The tilted angle CZT detector was considered as the detector of choice. Experiments were performed to test the effect of the tilting angle on the energy spectrum. Using the CZT detector with 20 degrees tilting angle decreased the tailing of the measured x-ray spectrum as compared to a conventional CZT detector. It was concluded that the energy-resolved MSMS CT with tilted angle CZT detector is potentially feasible and could provide a unique combination of photon counting, energy weighting, scatter rejection and single kVp dual energy subtraction CT imaging.

Entities:  

Mesh:

Year:  2008        PMID: 18296774     DOI: 10.1088/0031-9155/53/5/020

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  39 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

3.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

4.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

5.  Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition.

Authors:  Xin Liu; Lifeng Yu; Andrew N Primak; Cynthia H McCollough
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

6.  The effects of extending the spectral information acquired by a photon-counting detector for spectral CT.

Authors:  Taly Gilat Schmidt; Kevin C Zimmerman; Emil Y Sidky
Journal:  Phys Med Biol       Date:  2015-01-23       Impact factor: 3.609

Review 7.  Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology.

Authors:  Shuai Leng; Michael Bruesewitz; Shengzhen Tao; Kishore Rajendran; Ahmed F Halaweish; Norbert G Campeau; Joel G Fletcher; Cynthia H McCollough
Journal:  Radiographics       Date:  2019 May-Jun       Impact factor: 5.333

8.  Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system.

Authors:  Shuai Leng; Zhicong Yu; Ahmed Halaweish; Steffen Kappler; Katharina Hahn; Andre Henning; Zhoubo Li; John Lane; David L Levin; Steven Jorgensen; Erik Ritman; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-12-22

9.  150-μm Spatial Resolution Using Photon-Counting Detector Computed Tomography Technology: Technical Performance and First Patient Images.

Authors:  Shuai Leng; Kishore Rajendran; Hao Gong; Wei Zhou; Ahmed F Halaweish; Andre Henning; Steffen Kappler; Matthias Baer; Joel G Fletcher; Cynthia H McCollough
Journal:  Invest Radiol       Date:  2018-11       Impact factor: 6.016

10.  Spectral performance of a whole-body research photon counting detector CT: quantitative accuracy in derived image sets.

Authors:  Shuai Leng; Wei Zhou; Zhicong Yu; Ahmed Halaweish; Bernhard Krauss; Bernhard Schmidt; Lifeng Yu; Steffen Kappler; Cynthia McCollough
Journal:  Phys Med Biol       Date:  2017-08-21       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.