Alexandra D Jensen1, Marcus Winter, Sabine P Kuhn, Jürgen Debus, Olaf Nairz, Marc W Münter.
Abstract
PURPOSE: To investigate repositioning accuracy in particle radiotherapy in 6 degrees of freedom (DOF) and intensity-modulated radiotherapy (IMRT, 3 DOF) for two immobilization devices (Scotchcast masks vs thermoplastic head masks) currently in use at our institution for fractionated radiation therapy in head and neck cancer patients. METHODS AND MATERIALS: Position verifications in patients treated with carbon ion therapy and IMRT for head and neck malignancies were evaluated. Most patients received combined treatment regimen (IMRT plus carbon ion boost), immobilization was achieved with either Scotchcast or thermoplastic head masks. Position corrections in robotic-based carbon ion therapy allowing 6 DOF were compared to IMRT allowing corrections in 3 DOF for two standard immobilization devices. In total, 838 set-up controls of 38 patients were analyzed.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22458824 PMCID: PMC3337272 DOI: 10.1186/1748-717X-7-51
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Baseline characteristics
| radiation therapy | dose (Gy/GyRBE) | range (Gy/GyRBE) | |
|---|---|---|---|
| carbon ion therapy | 23.5 | 18 - 24.4 | |
| IMRT | 49 | 39 - 51 | |
| adenoid cystic carcinoma | 31 | ||
| malignant melanoma | 3 | ||
| mucoepidermoid carcinoma | 1 | ||
| nasopharyngeal carcinoma | 1 | ||
| osteosarcoma | 1 | ||
| adenocarcinoma | 1 | ||
| 20 | 18 | 38 | |
| 6 DOF | 162 | 146 | 308 |
| 3 DOF | 319 | 211 | 530 |
Figure 1Scotchcast mask.
Figure 2Thermoplastic head mask including shoulder fixation.
Figure 3Particle therapy position correction tool: 1: orthogonal x-rays acquired for position control at each session 2: digitally reconstructed radiographs (DRRs) from the planning CT-scan 3: slider to switch between the x-ray (position control) and DRR (planning scan). 4: manual correction tool for 6 DOF 5: resulting correction vector.
Figure 4a/b: Routine photon therapy position correction tool: 1: axial cone-beam CT/planning CT scan overlay; cone-beam CT: gray scale, planning CT: hot body 2: sagittal cone-beam CT/planning CT scan overlay 3: coronal cone-beam CT/planning CT scan overlay 4: slider to switch between the x-ray (position control) and DRR (planning scan). 5: resulting correction vector.
Corrections in 3 and 6 DOF; σc: centered standard deviation
| corrections in 6 DOF | ||||||
|---|---|---|---|---|---|---|
| 1.2 | 1.3 | 0.046 | 0.0 | 1.9 | 0.836 | |
| -0.3 | 1.0 | 0.153 | 0.5 | 1.6 | 0.127 | |
| 0.4 | 0.8 | 0.113 | 1.2 | 1.2 | 0.047 | |
| 0.3 | 0.6 | 0.142 | -0.1 | 0.9 | 0.396 | |
| 0.3 | 0.8 | 0.148 | 0.4 | 1.1 | 0.119 | |
| 0.2 | 0.6 | 0.175 | -0.2 | 1.2 | 0.258 | |
| 2.1 | 2.55 | |||||
| 0.8 | 1.3 | < 0.001 | ||||
| 0.6 | 0.9 | < 0.001 | ||||
| 0.6 | 0.7 | 0.062 | ||||
| 0.4 | 0.5 | 0.019 | ||||
| 0.5 | 0.7 | < 0.001 | ||||
| 0.4 | 0.6 | < 0.001 | ||||
| 0.4 | 2.3 | 0.133 | 0.6 | 1.9 | 0.107 | |
| -0.4 | 1.9 | 0.122 | 1.1 | 1.8 | 0.055 | |
| -0.4 | 2.3 | 0.145 | 0.00 | 1.8 | 1.000 | |
| 1.4 | 1.0 | < 0.001 | ||||
| 1.3 | 1.0 | < 0.001 | ||||
| 1.2 | 1.1 | 0.096 | ||||
| 3.5 | 3.0 | |||||
Action levels
| 6DOF | p = | 3DOF | p = | |||
|---|---|---|---|---|---|---|
| Action Levels | Scotchcast | Thermoplast | Scotchcast | Thermoplast | ||
| > 3 mm (component) | 3% | 23% | < 0.001 | 22% | 15% | 0.057 |
| > 3 mm (radial vector) | 7% | 28% | < 0.001 | 53% | 49% | 0.448 |