Literature DB >> 16154217

Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging.

Mandy Humphreys1, M Teressa Guerrero Urbano, Cefas Mubata, Elizabeth Miles, Kevin J Harrington, Margaret Bidmead, Christopher M Nutting.   

Abstract

PURPOSE: To evaluate set-up reproducibility of a cabulite shell and determine CTV-PTV margins for head and neck intensity-modulated-radiotherapy.
MATERIALS AND METHODS: Twenty patients were entered into the study. A total of 354 anterior and lateral isocentric electronic portal images (EPIs) were compared to simulator reference images.
RESULTS: About 94% of all translational displacements were < or =3 mm, and 99% < or =5 mm. The overall systematic error was 0.9 mm (+/-1.0SD) in the Right-Left, 0.7 mm (+/-0.9SD) in the Superior-Inferior and -0.02 mm (+/-1.1SD) in the Anterior-Posterior directions. The corresponding SDs of the random errors were +/-0.4, +/-0.6 and +/-0.7 mm. The estimated margins required from CTV-PTV were calculated according to the Van Herk formula was 2.9, 2.6 and 3.3 mm, respectively.
CONCLUSIONS: This head and neck immobilisation system is of sufficient accuracy for its use with IMRT treatments and a 3 mm CTV-PTV margin has been adopted.

Entities:  

Mesh:

Year:  2005        PMID: 16154217     DOI: 10.1016/j.radonc.2005.06.039

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  9 in total

1.  Correction of systematic set-up error in breast and head and neck irradiation through a no-action level (NAL) protocol.

Authors:  Eva M Lozano; Luis A Pérez; Javier Torres; Carmen Carrascosa; Miguel Sanz; Fermín Mendicote; Antonio Gil
Journal:  Clin Transl Oncol       Date:  2011-01       Impact factor: 3.405

2.  Anisotropic margin expansions in 6 anatomic directions for oropharyngeal image guided radiation therapy.

Authors:  Adam D Yock; Adam S Garden; Laurence E Court; Beth M Beadle; Lifei Zhang; Lei Dong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-07-29       Impact factor: 7.038

3.  Inter- and intrafractional positional uncertainties in pediatric radiotherapy patients with brain and head and neck tumors.

Authors:  Chris Beltran; Matthew J Krasin; Thomas E Merchant
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-06-03       Impact factor: 7.038

4.  Dosimetric impact of intrafractional patient motion in pediatric brain tumor patients.

Authors:  Chris Beltran; John Trussell; Thomas E Merchant
Journal:  Med Dosim       Date:  2009-02-07       Impact factor: 1.482

5.  Impact of residual setup error on parotid gland dose in intensity-modulated radiation therapy with or without planning organ-at-risk margin.

Authors:  Anna Delana; Loris Menegotti; Andrea Bolner; Luigi Tomio; Aldo Valentini; Frank Lohr; Valentina Vanoni
Journal:  Strahlenther Onkol       Date:  2009-08-28       Impact factor: 3.621

6.  Robotic-based carbon ion therapy and patient positioning in 6 degrees of freedom: setup accuracy of two standard immobilization devices used in carbon ion therapy and IMRT.

Authors:  Alexandra D Jensen; Marcus Winter; Sabine P Kuhn; Jürgen Debus; Olaf Nairz; Marc W Münter
Journal:  Radiat Oncol       Date:  2012-03-29       Impact factor: 3.481

7.  Evaluation of radiotherapy techniques for radical treatment of lateralised oropharyngeal cancers : Dosimetry and NTCP.

Authors:  D McQuaid; A Dunlop; S Nill; C Franzese; C M Nutting; K J Harrington; K L Newbold; S A Bhide
Journal:  Strahlenther Onkol       Date:  2016-06-13       Impact factor: 3.621

8.  Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device.

Authors:  Tejpal Gupta; Supriya Chopra; Avinash Kadam; Jai Prakash Agarwal; P Reena Devi; Sarbani Ghosh-Laskar; Ketayun Ardeshir Dinshaw
Journal:  Radiat Oncol       Date:  2007-12-14       Impact factor: 3.481

9.  Assessment of set-up errors in the radiotherapy of patients with head and neck cancer: standard vs. individual head support.

Authors:  Sabina Androjna; Valerija Zager Marcius; Primoz Peterlin; Primoz Strojan
Journal:  Radiol Oncol       Date:  2020-06-06       Impact factor: 2.991

  9 in total