Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors.

L Gilbeau1, M Octave-Prignot, T Loncol, L Renard, P Scalliet, V Grégoire.   

Abstract

PURPOSE: Setup accuracy is an important factor influencing the definition of the planning target volume (PTV). The purpose of this study was to compare the setup accuracy of three different thermoplastic masks used for immobilization of patients with brain or head and neck tumors.
MATERIALS AND METHODS: Thirty patients with brain or head and neck tumors were consecutively assigned to one of three different thermoplastic masks (Posifix): head mask with three fixation points (3 FP, ten patients), head and shoulder mask with four fixation points (4 FP, ten patients), head and shoulder mask with five fixation points (5 FP, four fixations plus an additional one on the top of head, ten patients). Once a week, during the session with a 6 MV linac (Elekta), orthogonal (antero-posterior and lateral) portal images were acquired for three fictitious isocenters placed during the simulation at the level of the head, the neck and the shoulders. Portal images and digitized simulator films were compared using the PIPS pro software, and displacements in antero-posterior (A-P), cranio-caudal (C-C) and medio-lateral (M-L) directions were calculated. From these displacements, 2D or 3D errors were also calculated.
RESULTS: A total of 915 portal images were obtained, of which 98% could be analyzed. For the whole population, total displacements reached a standard deviation (SD) of 2.2 mm at the level of the head and the neck. Systematic and random displacements were in the same order of magnitude and reached a SD of 1.8 mm. Patient setup was slightly worse at the shoulder level with a total displacement of 2.8 mm (1 SD) for both the C-C and the M-L directions. There again, the systematic and the random components were in the same order of magnitude below 2.4 mm (+/-SD). For isocenters in the head and in the neck, there was no substantial difference in the setup deviation between the three masks. The setup reproducibility was found to be significantly worse (P=0.01) at the level of the shoulders with the 3 FP mask. For the 2D random error, 1 SD of 2.3 mm was observed compared to 0.8 and 1.2 mm for the 4 and 5 FP masks, respectively. Lastly, 90% of the 3D total deviations were below 4.5 mm for the head and the neck. In the shoulder region, 90% of the 2D total deviations were below 5.5 mm.
CONCLUSION: Thermoplastic masks provide an accurate patient immobilization. At the shoulder level, setup variations are reduced when 4 or 5 FP masks are used. These data could be used for the assessment of margins for the PTV.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11166866     DOI: 10.1016/s0167-8140(00)00280-2

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  39 in total

1.  A gradient-based method for segmenting FDG-PET images: methodology and validation.

Authors:  Xavier Geets; John A Lee; Anne Bol; Max Lonneux; Vincent Grégoire
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-03-13       Impact factor: 9.236

2.  Respiratory gated [18F]FDG PET/CT for target volume delineation in stereotactic radiation treatment of liver metastases.

Authors:  R A Bundschuh; N Andratschke; J Dinges; M N Duma; S T Astner; M Brügel; S I Ziegler; M Molls; M Schwaiger; M Essler
Journal:  Strahlenther Onkol       Date:  2012-03-24       Impact factor: 3.621

3.  Comparison of the effectiveness of different immobilization systems in different body regions using daily megavoltage CT in helical tomotherapy.

Authors:  K-F Cheng; V W C Wu
Journal:  Br J Radiol       Date:  2014-01-07       Impact factor: 3.039

4.  Quality of patient positioning during cerebral tomotherapy irradiation using different mask systems.

Authors:  C Leitzen; T Wilhelm-Buchstab; S Garbe; C Lütter; T Müdder; B Simon; H H Schild; H Schüller
Journal:  Strahlenther Onkol       Date:  2013-12-11       Impact factor: 3.621

5.  Accurate positioning for head and neck cancer patients using 2D and 3D image guidance.

Authors:  Hyejoo Kang; Dale M Lovelock; Ellen D Yorke; Sergey Kriminski; Nancy Lee; Howard I Amols
Journal:  J Appl Clin Med Phys       Date:  2010-10-27       Impact factor: 2.102

6.  A prospective analysis of inter- and intrafractional errors to calculate CTV to PTV margins in head and neck patients.

Authors:  J Cacicedo; J F Perez; R Ortiz de Zarate; O del Hoyo; F Casquero; A Gómez-Iturriaga; A Lasso; E Boveda; P Bilbao
Journal:  Clin Transl Oncol       Date:  2014-07-19       Impact factor: 3.405

7.  Radial displacement of clinical target volume in node negative head and neck cancer.

Authors:  Wan Jeon; Hong-Gyun Wu; Sang Hyuk Song; Jung-In Kim
Journal:  Radiat Oncol J       Date:  2012-03-31

8.  Recent advances in image-guided radiotherapy for head and neck carcinoma.

Authors:  Sameer K Nath; Daniel R Simpson; Brent S Rose; Ajay P Sandhu
Journal:  J Oncol       Date:  2009-07-29       Impact factor: 4.375

9.  Verification of setup errors in external beam radiation therapy using electronic portal imaging.

Authors:  K Krishna Murthy; Zakiya Al-Rahbi; S S Sivakumar; C A Davis; R Ravichandran; Kamal El Ghamrawy
Journal:  J Med Phys       Date:  2008-04

10.  Immobilization versus no immobilization for pelvic external beam radiotherapy.

Authors:  Avinash H Udayashankar; Shibina Noorjahan; Nirmala Srikantia; K Ravindra Babu; Sandeep Muzumder
Journal:  Rep Pract Oncol Radiother       Date:  2018-05-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.