Myra F Rodrigues1,2, Sten Veen1, Jaap van Egmond3, Mark van Hameren1, Theodorus van Oorschot1, Steven de Vet1, Jan P C van Santvoort3, Ruud G J Wiggenraad1, Mirjam E Mast1.
Abstract
Reproducible patient positioning is important in radiotherapy (RT) of head-and-neck cancer. We therefore compared set-up errors in head-and-neck RT resulting from three different patient positioning systems. Patients were either treated with a standard head support (SHS) and conventional treatment couch (SHS-3, n = 10), a SHS and rotational couch (SHS-6, n = 10), or an individual head support (IHS) and rotational couch (IHS-6, n = 10). Interfraction mean translation vector lenghts were significantly lower for IHS-6 compared to SHS-3 (0.8 ± 0.3 mm vs. 1.4 ± 0.7 mm, P = 0.001). Intrafraction displacement was comparable among cohorts. This study showed that the use of a six degrees of freedom couch combined with an IHS in head-and-neck RT resulted in better interfraction reproducibility.Entities:
Keywords: Cone-beam CT; Head support; Head-and-neck cancer; Immobilization; Set-up correction; Six degrees of freedom couch
Year: 2019 PMID: 33458274 PMCID: PMC7807734 DOI: 10.1016/j.phro.2019.07.001
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Interfraction errors: mean translation vector length of the clipbox match.
| SHS-3 | SHS-6 | IHS-6 | ||||||
|---|---|---|---|---|---|---|---|---|
| Vector [mm] | Vector [mm] | Vector [mm] | ||||||
| Patient | mean | (SD) | Patient | mean | (SD) | Patient | mean | (SD) |
| 1 | 1.9 | (1.0) | 11 | 1.0 | (0.5) | 21 | 0.8 | (0.3) |
| 2 | 2.8 | (0.9) | 12 | 1.3 | (0.5) | 22 | 0.5 | (0.3) |
| 3 | 0.5 | (0.2) | 13 | 0.6 | (0.2) | 23 | 0.8 | (0.4) |
| 4 | 1.3 | (0.6) | 14 | 0.8 | (0.4) | 24 | 1.2 | (0.9) |
| 5 | 1.4 | (1.1) | 15 | 1.8 | (1.2) | 25 | 1.3 | (0.5) |
| 6 | 0.8 | (0.4) | 16 | 0.7 | (0.4) | 26 | 0.9 | (0.3) |
| 7 | 1.1 | (0.5) | 17 | 0.5 | (0.3) | 27 | 0.7 | (0.4) |
| 8 | 1.7 | (1.3) | 18 | 0.4 | (0.1) | 28 | 0.4 | (0.2) |
| 9 | 1.6 | (0.7) | 19 | 1.2 | (0.3) | 29 | 1.1 | (1.0) |
| 10 | 0.6 | (0.4) | 20 | 1.3 | (0.6) | 30 | 0.6 | (0.4) |
| Mean [mm]: | 1.4 | (0.7) | Mean [mm]: | 1.0 | (0.5) | Mean [mm]: | 0.8 | (0.5) |
| SD [mm]: | 0.7 | SD [mm]: | 0.4 | SD [mm]: | 0.3 | |||
SHS-3, standard head support and 3 degrees of freedom couch; SHS-6, standard head support and 6 degrees of freedom couch; IHS-6, individual head support and 6 degrees of freedom couch.
Fig. 1Clipbox match: (A) Minimum translation vector length. (B) Minimum rotation angle in either LR-, AP- or CC-direction. Blue line: SHS-3; Orange line: SHS-6; Grey line: IHS-6. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)