A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization.

W Zbijewski1, P De Jean, P Prakash, Y Ding, J W Stayman, N Packard, R Senn, D Yang, J Yorkston, A Machado, J A Carrino, J H Siewerdsen.   

Abstract

PURPOSE: This paper reports on the design and initial imaging performance of a dedicated cone-beam CT (CBCT) system for musculoskeletal (MSK) extremities. The system complements conventional CT and MR and offers a variety of potential clinical and logistical advantages that are likely to be of benefit to diagnosis, treatment planning, and assessment of therapy response in MSK radiology, orthopaedic surgery, and rheumatology.
METHODS: The scanner design incorporated a host of clinical requirements (e.g., ability to scan the weight-bearing knee in a natural stance) and was guided by theoretical and experimental analysis of image quality and dose. Such criteria identified the following basic scanner components and system configuration: a flat-panel detector (FPD, Varian 3030+, 0.194 mm pixels); and a low-power, fixed anode x-ray source with 0.5 mm focal spot (SourceRay XRS-125-7K-P, 0.875 kW) mounted on a retractable C-arm allowing for two scanning orientations with the capability for side entry, viz. a standing configuration for imaging of weight-bearing lower extremities and a sitting configuration for imaging of tensioned upper extremity and unloaded lower extremity. Theoretical modeling employed cascaded systems analysis of modulation transfer function (MTF) and detective quantum efficiency (DQE) computed as a function of system geometry, kVp and filtration, dose, source power, etc. Physical experimentation utilized an imaging bench simulating the scanner geometry for verification of theoretical results and investigation of other factors, such as antiscatter grid selection and 3D image quality in phantom and cadaver, including qualitative comparison to conventional CT.
RESULTS: Theoretical modeling and benchtop experimentation confirmed the basic suitability of the FPD and x-ray source mentioned above. Clinical requirements combined with analysis of MTF and DQE yielded the following system geometry: a -55 cm source-to-detector distance; 1.3 magnification; a 20 cm diameter bore (20 x 20 x 20 cm3 field of view); total acquisition arc of -240 degrees. The system MTF declines to 50% at -1.3 mm(-1) and to 10% at -2.7 mm(-1), consistent with sub-millimeter spatial resolution. Analysis of DQE suggested a nominal technique of 90 kVp (+0.3 mm Cu added filtration) to provide high imaging performance from -500 projections at less than -0.5 kW power, implying -6.4 mGy (0.064 mSv) for low-dose protocols and -15 mGy (0.15 mSv) for high-quality protocols. The experimental studies show improved image uniformity and contrast-to-noise ratio (without increase in dose) through incorporation of a custom 10:1 GR antiscatter grid. Cadaver images demonstrate exquisite bone detail, visualization of articular morphology, and soft-tissue visibility comparable to diagnostic CT (10-20 HU contrast resolution).
CONCLUSIONS: The results indicate that the proposed system will deliver volumetric images of the extremities with soft-tissue contrast resolution comparable to diagnostic CT and improved spatial resolution at potentially reduced dose. Cascaded systems analysis provided a useful basis for system design and optimization without costly repeated experimentation. A combined process of design specification, image quality analysis, clinical feedback, and revision yielded a prototype that is now awaiting clinical pilot studies. Potential advantages of the proposed system include reduced space and cost, imaging of load-bearing extremities, and combined volumetric imaging with real-time fluoroscopy and digital radiography.

Entities:  

Mesh:

Year:  2011        PMID: 21928644      PMCID: PMC3172864          DOI: 10.1118/1.3611039

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  A method for including redundant data in computed tomography.

Authors:  M D Silver
Journal:  Med Phys       Date:  2000-04       Impact factor: 4.071

2.  Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography).

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

3.  Micro-angiography for neuro-vascular imaging. II. Cascade model analysis.

Authors:  Arundhuti Ganguly; Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2003-11       Impact factor: 4.071

Review 4.  Image guidance in spine surgery.

Authors:  Langston T Holly; Kevin T Foley
Journal:  Orthop Clin North Am       Date:  2007-07       Impact factor: 2.472

5.  Assessment of patient and occupational dose in established and new applications of MDCT fluoroscopy.

Authors:  Raoul M S Joemai; Dirk Zweers; Wim R Obermann; Jacob Geleijns
Journal:  AJR Am J Roentgenol       Date:  2009-04       Impact factor: 3.959

6.  Cascaded systems analysis of the 3D noise transfer characteristics of flat-panel cone-beam CT.

Authors:  Daniel J Tward; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

7.  Radiation exposure from musculoskeletal computerized tomographic scans.

Authors:  Debdut Biswas; Jesse E Bible; Michael Bohan; Andrew K Simpson; Peter G Whang; Jonathan N Grauer
Journal:  J Bone Joint Surg Am       Date:  2009-08       Impact factor: 5.284

8.  Digital radiology using active matrix readout of amorphous selenium: theoretical analysis of detective quantum efficiency.

Authors:  W Zhao; J A Rowlands
Journal:  Med Phys       Date:  1997-12       Impact factor: 4.071

9.  Clinical utility of tomosynthesis in suspected scaphoid fracture. A pilot study.

Authors:  Mats Geijer; Annika M Börjesson; Jan H Göthlin
Journal:  Skeletal Radiol       Date:  2010-11-06       Impact factor: 2.199

10.  Comparison of 1.0-T extremity MR and 1.5-T conventional high-field-Strength MR in patients with rheumatoid arthritis.

Authors:  Ali M Naraghi; Lawrence M White; Chirag Patel; George Tomlinson; Edward C Keystone
Journal:  Radiology       Date:  2009-04-20       Impact factor: 11.105

View more
  55 in total

1.  Antiscatter grids in mobile C-arm cone-beam CT: effect on image quality and dose.

Authors:  S Schafer; J W Stayman; W Zbijewski; C Schmidgunst; G Kleinszig; J H Siewerdsen
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging.

Authors:  P Prakash; W Zbijewski; G J Gang; Y Ding; J W Stayman; J Yorkston; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

4.  Fiducial marker-based correction for involuntary motion in weight-bearing C-arm CT scanning of knees. Part I. Numerical model-based optimization.

Authors:  Jang-Hwan Choi; Rebecca Fahrig; Andreas Keil; Thor F Besier; Saikat Pal; Emily J McWalter; Gary S Beaupré; Andreas Maier
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

5.  Three-dimensional computed tomography evaluation of anterior cruciate ligament footprint for anatomic single-bundle reconstruction.

Authors:  Guilherme Moreira de Abreu-e-Silva; Mcbrite H G Castro Nunes de Oliveira; Gustavo Silame Maranhão; Lucas de Melo Castro Deligne; Rudolf Moreira Pfeilsticker; Eduardo Nilo Vasconcellos Novais; Tarcizo Afonso Nunes; Marco Antônio Percope de Andrade
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-10-22       Impact factor: 4.342

6.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

7.  Extremity cone-beam CT for evaluation of medial tibiofemoral osteoarthritis: Initial experience in imaging of the weight-bearing and non-weight-bearing knee.

Authors:  Gaurav K Thawait; Shadpour Demehri; Abdullah AlMuhit; Wojciech Zbijweski; John Yorkston; Filippo Del Grande; Bashir Zikria; John A Carrino; Jeffrey H Siewerdsen
Journal:  Eur J Radiol       Date:  2015-09-12       Impact factor: 3.528

Review 8.  Recent and future directions in CT imaging.

Authors:  Norbert J Pelc
Journal:  Ann Biomed Eng       Date:  2014-01-17       Impact factor: 3.934

9.  Direct comparison of conventional radiography and cone-beam CT in small bone and joint trauma.

Authors:  E De Smet; G De Praeter; K L A Verstraete; K Wouters; Luc De Beuckeleer; F M H M Vanhoenacker
Journal:  Skeletal Radiol       Date:  2015-03-12       Impact factor: 2.199

10.  Multiresolution iterative reconstruction in high-resolution extremity cone-beam CT.

Authors:  Qian Cao; Wojciech Zbijewski; Alejandro Sisniega; John Yorkston; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Phys Med Biol       Date:  2016-10-03       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.