Optimization of image acquisition techniques for dual-energy imaging of the chest.

N A Shkumat1, J H Siewerdsen, A C Dhanantwari, D B Williams, S Richard, N S Paul, J Yorkston, R Van Metter.   

Abstract

Experimental and theoretical studies were conducted to determine optimal acquisition techniques for a prototype dual-energy (DE) chest imaging system. Technique factors investigated included the selection of added x-ray filtration, kVp pair, and the allocation of dose between low- and high-energy projections, with total dose equal to or less than that of a conventional chest radiograph. Optima were computed to maximize lung nodule detectability as characterized by the signal-difference-to-noise ratio (SDNR) in DE chest images. Optimal beam filtration was determined by cascaded systems analysis of DE image SDNR for filter selections across the periodic table (Z(filter) = 1-92), demonstrating the importance of differential filtration between low- and high-kVp projections and suggesting optimal high-kVp filters in the range Z(filter) = 25-50. For example, added filtration of approximately 2.1 mm Cu, approximately 1.2 mm Zr, approximately 0.7 mm Mo, and approximately 0.6 mm Ag to the high-kVp beam provided optimal (and nearly equivalent) soft-tissue SDNR. Optimal kVp pair and dose allocation were investigated using a chest phantom presenting simulated lung nodules and ribs for thin, average, and thick body habitus. Low- and high-energy techniques ranged from 60-90 kVp and 120-150 kVp, respectively, with peak soft-tissue SDNR achieved at [60/120] kVp for all patient thicknesses and all levels of imaging dose. A strong dependence on the kVp of the low-energy projection was observed. Optimal allocation of dose between low- and high-energy projections was such that approximately 30% of the total dose was delivered by the low-kVp projection, exhibiting a fairly weak dependence on kVp pair and dose. The results have guided the implementation of a prototype DE imaging system for imaging trials in early-stage lung nodule detection and diagnosis.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17985636     DOI: 10.1118/1.2777278

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging.

Authors:  P Prakash; W Zbijewski; G J Gang; Y Ding; J W Stayman; J Yorkston; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration.

Authors:  A N Primak; J C Ramirez Giraldo; X Liu; L Yu; C H McCollough
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

3.  Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images.

Authors:  Samuel Richard; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

4.  Multiscale deformable registration for dual-energy x-ray imaging.

Authors:  G J Gang; C A Varon; H Kashani; S Richard; N S Paul; R Van Metter; J Yorkston; J H Siewerdsen
Journal:  Med Phys       Date:  2009-02       Impact factor: 4.071

5.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

6.  Evaluation of a template-based algorithm for markerless lung tumour localization on single- and dual-energy kilovoltage images.

Authors:  Alec M Block; Rakesh Patel; Murat Surucu; Matthew M Harkenrider; John C Roeske
Journal:  Br J Radiol       Date:  2016-10-12       Impact factor: 3.039

7.  Optimization of dual-energy subtraction chest radiography by use of a direct-conversion flat-panel detector system.

Authors:  Mari Fukao; Kiyosumi Kawamoto; Hiroaki Matsuzawa; Osamu Honda; Takeshi Iwaki; Tsukasa Doi
Journal:  Radiol Phys Technol       Date:  2014-08-15

8.  Dual-energy cone-beam CT with a flat-panel detector: effect of reconstruction algorithm on material classification.

Authors:  W Zbijewski; G J Gang; J Xu; A S Wang; J W Stayman; K Taguchi; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

9.  Dual-source dual-energy CT with additional tin filtration: Dose and image quality evaluation in phantoms and in vivo.

Authors:  Andrew N Primak; Juan Carlos Ramirez Giraldo; Christian D Eusemann; Bernhard Schmidt; Birgit Kantor; Joel G Fletcher; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2010-11       Impact factor: 3.959

10.  Development of a high-performance dual-energy chest imaging system: initial investigation of diagnostic performance.

Authors:  Hany Kashani; Jianan Grace Gang; Nicholas A Shkumat; Carlos A Varon; John Yorkston; Richard Van Metter; Narinder S Paul; Jeffrey H Siewerdsen
Journal:  Acad Radiol       Date:  2009-04       Impact factor: 3.173

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.