Mapping of low flip angles in magnetic resonance.

Fabien Balezeau1, Pierre-Antoine Eliat, Alejandro Bordelois Cayamo, Hervé Saint-Jalmes.   

Abstract

Errors in the flip angle have to be corrected in many magnetic resonance imaging applications, especially for T1 quantification. However, the existing methods of B1 mapping fail to measure lower values of the flip angle despite the fact that these are extensively used in dynamic acquisition and 3D imaging. In this study, the nonlinearity of the radiofrequency (RF) transmit chain, especially for very low flip angles, is investigated and a simple method is proposed to accurately determine both the gain of the RF transmitter and the B1 field map for low flip angles. The method makes use of the spoiled gradient echo sequence with long repetition time (TR), such as applied in the double-angle method. It uses an image acquired with a flip angle of 90° as a reference image that is robust to B1 inhomogeneity. The ratio of the image at flip angle alpha to the image at a flip angle of 90° enables us to calculate the actual value of alpha. This study was carried out at 1.5 and 4.7 T, showing that the linearity of the RF supply system is highly dependent on the hardware. The method proposed here allows us to measure the flip angle from 1° to 60° with a maximal uncertainty of 10% and to correct T1 maps based on the variable flip angle method.

Mesh:

Year:  2011        PMID: 21941028      PMCID: PMC3391187          DOI: 10.1088/0031-9155/56/20/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  23 in total

1.  Effects of RF amplifier distortion on selective excitation and their correction by prewarping.

Authors:  F Chan; J Pauly; A Macovski
Journal:  Magn Reson Med       Date:  1992-02       Impact factor: 4.668

2.  An analysis of the accuracy of magnetic resonance flip angle measurement methods.

Authors:  Glen R Morrell; Matthias C Schabel
Journal:  Phys Med Biol       Date:  2010-09-29       Impact factor: 3.609

3.  Rapid high-resolution T(1) mapping by variable flip angles: accurate and precise measurements in the presence of radiofrequency field inhomogeneity.

Authors:  Hai-Ling Margaret Cheng; Graham A Wright
Journal:  Magn Reson Med       Date:  2006-03       Impact factor: 4.668

4.  Saturated double-angle method for rapid B1+ mapping.

Authors:  Charles H Cunningham; John M Pauly; Krishna S Nayak
Journal:  Magn Reson Med       Date:  2006-06       Impact factor: 4.668

5.  Accurate determination of spin-density and T1 in the presence of RF-field inhomogeneities and flip-angle miscalibration.

Authors:  R Venkatesan; W Lin; E M Haacke
Journal:  Magn Reson Med       Date:  1998-10       Impact factor: 4.668

6.  Imaging of the active B1 field in vivo.

Authors:  R Stollberger; P Wach
Journal:  Magn Reson Med       Date:  1996-02       Impact factor: 4.668

7.  Snapshot FLASH MRI. Applications to T1, T2, and chemical-shift imaging.

Authors:  A Haase
Journal:  Magn Reson Med       Date:  1990-01       Impact factor: 4.668

8.  Radiofrequency map of an NMR coil by imaging.

Authors:  S Akoka; F Franconi; F Seguin; A Le Pape
Journal:  Magn Reson Imaging       Date:  1993       Impact factor: 2.546

9.  Rapid B1 mapping using orthogonal, equal-amplitude radio-frequency pulses.

Authors:  Yulin V Chang
Journal:  Magn Reson Med       Date:  2011-06-28       Impact factor: 4.668

10.  High resolution MRI relaxation measurements of water in the articular cartilage of the meniscectomized rat knee at 4.7 T.

Authors:  Yiannis Spandonis; Frank P Heese; Laurance D Hall
Journal:  Magn Reson Imaging       Date:  2004-09       Impact factor: 2.546

View more
  9 in total

1.  Modification of population based arterial input function to incorporate individual variation.

Authors:  Harrison Kim
Journal:  Magn Reson Imaging       Date:  2017-09-27       Impact factor: 2.546

2.  Variability in Quantitative DCE-MRI: Sources and Solutions.

Authors:  Harrison Kim
Journal:  J Nat Sci       Date:  2018

3.  Steady-state double-angle method for rapid B1 mapping.

Authors:  Mustapha Bouhrara; Richard G Spencer
Journal:  Magn Reson Med       Date:  2019-03-03       Impact factor: 4.668

4.  Simultaneous measurement of T1 /B1 and pharmacokinetic model parameters using active contrast encoding (ACE)-MRI.

Authors:  Jin Zhang; Kerryanne Winters; Olivier Reynaud; Sungheon Gene Kim
Journal:  NMR Biomed       Date:  2017-05-22       Impact factor: 4.044

5.  Advances in Diffusion and Perfusion MRI for Quantitative Cancer Imaging.

Authors:  Mehran Baboli; Jin Zhang; Sungheon Gene Kim
Journal:  Curr Pathobiol Rep       Date:  2019-12-02

6.  Fast B1 Mapping Based on Double-Angle Method with T1 Correction Using Standard Pulse Sequence.

Authors:  Yoshiyuki Ishimori; Takeshi Shimanuki; Tomoya Kobayashi; Masahiko Monma
Journal:  J Med Phys       Date:  2022-03-31

7.  Quantitative multi-parameter mapping of R1, PD(*), MT, and R2(*) at 3T: a multi-center validation.

Authors:  Nikolaus Weiskopf; John Suckling; Guy Williams; Marta M Correia; Becky Inkster; Roger Tait; Cinly Ooi; Edward T Bullmore; Antoine Lutti
Journal:  Front Neurosci       Date:  2013-06-10       Impact factor: 4.677

8.  Optimization of steady-state free precession MRI for lung ventilation imaging with 19 F C3 F8 at 1.5T and 3T.

Authors:  Adam Maunder; Madhwesha Rao; Fraser Robb; Jim M Wild
Journal:  Magn Reson Med       Date:  2018-11-02       Impact factor: 4.668

9.  Unbalanced SSFP for super-resolution in MRI.

Authors:  Peter J Lally; Paul M Matthews; Neal K Bangerter
Journal:  Magn Reson Med       Date:  2020-11-17       Impact factor: 3.737

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.