Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction.

Jongduk Baek1, Norbert J Pelc.   

Abstract

PURPOSE: The authors examine the nonstationary noise behavior of a cone-beam CT system with FDK reconstruction.
METHODS: To investigate the nonstationary noise behavior, an analytical expression for the NPS of local volumes and an entire volume was derived and quantitatively compared to the NPS estimated from experimental air and water images.
RESULTS: The NPS of local volumes at different locations along the z-axis showed radial symmetry in the f(x)-f(y) plane and different missing cone regions in the f(z) direction depending on the tilt angle of rays through the local volumes. For local volumes away from the z-axis, the NPS of air and water images showed sharp transitions in the f(x)-f(y) and f(y)-f(z) planes and lack of radial symmetry in the f(x)-f(y) plane. These effects are mainly caused by varying magnification and different noise levels from view to view. In the NPS of the entire volume, the f(x)-f(y) plane showed radial symmetry because the nonstationary noise behaviors of local volumes were averaged out. The nonstationary sharp transitions were manifested as a high-frequency roll-off.
CONCLUSIONS: The results from noise power analysis for local volumes and an entire volume demonstrate the spatially varying noise behavior in the reconstructed cone-beam CT images.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21626945      PMCID: PMC3078160          DOI: 10.1118/1.3556590

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  An experimental comparison of detector performance for computed radiography systems.

Authors:  Ehsan Samei; Michael J Flynn
Journal:  Med Phys       Date:  2002-04       Impact factor: 4.071

2.  The noise power spectrum in CT with direct fan beam reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

3.  Experimental optical fan beam tomography.

Authors:  K E Bennett; G W Faris; R L Byer
Journal:  Appl Opt       Date:  1984-08-15       Impact factor: 1.980

4.  Estimation of the local statistical noise in emission computed tomography.

Authors:  N M Alpert; D A Chesler; J A Correia; R H Ackerman; J Y Chang; S Finklestein; S M Davis; G L Brownell; J M Taveras
Journal:  IEEE Trans Med Imaging       Date:  1982       Impact factor: 10.048

5.  Optical tomography: experimental verification of noise theory.

Authors:  K Bennett; R L Byer
Journal:  Opt Lett       Date:  1984-07-01       Impact factor: 3.776

6.  Image covariance and lesion detectability in direct fan-beam x-ray computed tomography.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  Phys Med Biol       Date:  2008-04-18       Impact factor: 3.609

7.  Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; I A Cunningham
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

8.  Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function.

Authors:  K Faulkner; B M Moores
Journal:  Phys Med Biol       Date:  1984-11       Impact factor: 3.609

9.  The noise power spectrum in computed X-ray tomography.

Authors:  S J Riederer; N J Pelc; D A Chesler
Journal:  Phys Med Biol       Date:  1978-05       Impact factor: 3.609

10.  Detectability in computed tomographic images.

Authors:  K M Hanson
Journal:  Med Phys       Date:  1979 Sep-Oct       Impact factor: 4.071

View more
  15 in total

1.  Beyond noise power in 3D computed tomography: the local NPS and off-diagonal elements of the Fourier domain covariance matrix.

Authors:  Angel R Pineda; Daniel J Tward; Antonio Gonzalez; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

2.  Investigation of Noise and Contrast Sensitivity of an Electron Multiplying Charge-Coupled Device (EMCCD) based Cone Beam Micro-CT System.

Authors:  Sumukh Bysani Krishnakumar; Alexander R Podgorsak; Sv Setlur Nagesh; Amit Jain; Stephen Rudin; Daniel R Bednarek; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-25

3.  Characterization of imaging performance in differential phase contrast CT compared with the conventional CT--noise power spectrum NPS(k).

Authors:  Xiangyang Tang; Yi Yang; Shaojie Tang
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

4.  Effect of detector lag on CT noise power spectra.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

5.  Fundamental relationship between the noise properties of grating-based differential phase contrast CT and absorption CT: theoretical framework using a cascaded system model and experimental validation.

Authors:  Ke Li; Nicholas Bevins; Joseph Zambelli; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

7.  Task-based optimization of dedicated breast CT via Hotelling observer metrics.

Authors:  Adrian A Sanchez; Emil Y Sidky; Xiaochuan Pan
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

8.  Impact of bowtie filter and object position on the two-dimensional noise power spectrum of a clinical MDCT system.

Authors:  Daniel Gomez-Cardona; Juan Pablo Cruz-Bastida; Ke Li; Adam Budde; Jiang Hsieh; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

9.  A multi-source inverse-geometry CT system: initial results with an 8 spot x-ray source array.

Authors:  Jongduk Baek; Bruno De Man; Jorge Uribe; Randy Longtin; Daniel Harrison; Joseph Reynolds; Bogdan Neculaes; Kristopher Frutschy; Louis Inzinna; Antonio Caiafa; Robert Senzig; Norbert J Pelc
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

10.  Characterization of imaging performance in differential phase contrast CT compared with the conventional CT: spectrum of noise equivalent quanta NEQ(k).

Authors:  Xiangyang Tang; Yi Yang; Shaojie Tang
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.