An experimental comparison of detector performance for computed radiography systems.

Ehsan Samei1, Michael J Flynn.   

Abstract

The intrinsic resolution, noise, and signal-to-noise transfer characteristics of five commercial digital computed radiography (CR) systems were compared using identical experimental methods. The reader/screen combinations evaluated were Agfa ADC-Compact/MD-10, Agfa ADC-Compact/MD-30, Agfa ADC-Solo/MD-10, Agfa ADC-Solo/MD-30, Lumisys CR-2000/MD-10, Fuji FCR-9501 (HQ)/ST-Va, Kodak CR-400/GP-25, and Kodak CR-400/HR. Measurements were made at 70 and 115 kVp with 19 mm added aluminum filtration. The presampled modulation transfer functions (MTFs) of the systems were measured using an edge method. The noise power spectra (NPS) were determined by 2D Fourier analysis of uniformly exposed radiographs. The frequency-dependent detective quantum efficiencies (DQEs) were computed from the MTF, NPS, exposure measurements, and computational estimates of the ideal signal-to-noise ratios. Using 70 kVp and 0.1-0.12 mm pixel sizes, spatial frequencies of 2.1, 2.0, 2.2, 1.9, 2.0, 2.0, 2.3, 2.3, and 3.5 cycles/mm were measured at 0.2 MTF for the eight reader/screen combinations, respectively. Using 70 kVp, 7.74 x 10(-8) C/kg (0.3 mR), and 0.1-0.12 mm pixel sizes, DQE(0.15) values of 20.3%, 22.9%, 24.6%, 28.6%, 22.2%, 30.0%, 29.5%, and 17.3% were obtained for the eight combinations, respectively. The corresponding values at 115 kVp were 15.9%, 18.5%, 21.5%, 21.8%, 15.3%, 23.1%, 22.3%, and 13.8%, respectively. The findings of the study demonstrate the pixel size, orientation, beam quality, screen, and reader dependencies of image quality in CR systems. The physical performance of the systems having standard-resolution screens demonstrated similar resolution performance but more notable variations in DQE. The one high-resolution screen tested had reduced DQE and increased MTF at high frequencies.

Entities:  

Mesh:

Year:  2002        PMID: 11991117     DOI: 10.1118/1.1449873

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography.

Authors:  Patrik Sund; Magnus Båth; Susanne Kheddache; Lars Gunnar Månsson
Journal:  Eur Radiol       Date:  2003-10-16       Impact factor: 5.315

2.  Solid-state fluoroscopic imager for high-resolution angiography: parallel-cascaded linear systems analysis.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

3.  Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan; Steven K Onishi
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

4.  Relationship between radiation dose reduction and image quality change in photostimulable phosphor luminescence X-ray imaging systems.

Authors:  T Sakurai; R Kawamata; Y Kozai; Y Kaku; K Nakamura; M Saito; H Wakao; I Kashima
Journal:  Dentomaxillofac Radiol       Date:  2010-05       Impact factor: 2.419

5.  The noise power spectrum in CT with direct fan beam reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

6.  The gold standard: PACS alchemy and the gold standard.

Authors:  Kenneth Burgess; Thomas L Slovis
Journal:  Pediatr Radiol       Date:  2004-10-23

7.  Assessment of detective quantum efficiency: intercomparison of a recently introduced international standard with prior methods.

Authors:  Nicole T Ranger; Ehsan Samei; James T Dobbins; Carl E Ravin
Journal:  Radiology       Date:  2007-06       Impact factor: 11.105

8.  Single-exposure dual-energy subtraction chest radiography: detection of pulmonary nodules and masses in clinical practice.

Authors:  Zsolt Szucs-Farkas; Michael A Patak; Seyran Yuksel-Hatz; Thomas Ruder; Peter Vock
Journal:  Eur Radiol       Date:  2007-09-27       Impact factor: 5.315

9.  Effective DQE (eDQE) and speed of digital radiographic systems: an experimental methodology.

Authors:  Ehsan Samei; Nicole T Ranger; Alistair MacKenzie; Ian D Honey; James T Dobbins; Carl E Ravin
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

10.  Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.