Effect of detector lag on CT noise power spectra.

Jongduk Baek1, Norbert J Pelc.   

Abstract

PURPOSE: The authors examined the effect of detector lag on the noise power spectrum (NPS) of CT images reconstructed with filtered backprojection (FBP).
METHODS: The authors derived an analytical expression of the NPS with detector lag, and then verified it using computer simulations with parallel beam and fan beam geometries. The dependence of the NPS on the amount of lag, location within the scanned field of view (FOV), and the number of views used in the reconstruction (samples per rotation) was investigated using constant and view dependent noise in the raw data.
RESULTS: Detector lag introduces noise correlation in the azimuthal direction. The effect on the NPS is a frequency dependent reduction in amplitude. In small regions of the image, the effect is primarily in the frequencies corresponding to the azimuthal direction. The noise blurring and NPS filtering increases with increasing radial distance, and therefore regions at larger radial distances have lower noise power. With the same detector lag response function, the amount of noise correlation and NPS filtering decreases with increasing number of views.
CONCLUSIONS: The shape of the NPS depends on the detector lag coefficients, location of the region, and the number of views used in the reconstruction. In general, the noise correlation caused by detector lag decreased the amplitude of the NPS.

Mesh:

Year:  2011        PMID: 21815373      PMCID: PMC3162307          DOI: 10.1118/1.3589135

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications.

Authors:  L E Antonuk; K W Jee; Y El-Mohri; M Maolinbay; S Nassif; X Rong; Q Zhao; J H Siewerdsen; R A Street; K S Shah
Journal:  Med Phys       Date:  2000-02       Impact factor: 4.071

2.  Cone-beam computed tomography with a flat-panel imager: effects of image lag.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

3.  Additive noise properties of active matrix flat-panel imagers.

Authors:  M Maolinbay; Y El-Mohri; L E Antonuk; K W Jee; S Nassif; X Rong; Q Zhao
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

4.  The noise power spectrum in CT with direct fan beam reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

5.  Systematic investigation of the signal properties of polycrystalline HgI2 detectors under mammographic, radiographic, fluoroscopic and radiotherapy irradiation conditions.

Authors:  Zhong Su; Larry E Antonuk; Youcef El-Mohri; Larry Hu; Hong Du; Amit Sawant; Yixin Li; Yi Wang; Jin Yamamoto; Qihua Zhao
Journal:  Phys Med Biol       Date:  2005-06-01       Impact factor: 3.609

6.  Investigation of the signal behavior at diagnostic energies of prototype, direct detection, active matrix, flat-panel imagers incorporating polycrystalline HgI2.

Authors:  Hong Du; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Zhong Su; Jin Yamamoto; Yi Wang
Journal:  Phys Med Biol       Date:  2008-02-14       Impact factor: 3.609

7.  An empirical method for lag correction in cone-beam CT.

Authors:  N Mail; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

8.  Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

9.  The noise power spectrum in computed X-ray tomography.

Authors:  S J Riederer; N J Pelc; D A Chesler
Journal:  Phys Med Biol       Date:  1978-05       Impact factor: 3.609

10.  Polycrystalline Mercuric Iodide Films on CMOS Readout Arrays.

Authors:  Neal E Hartsough; Jan S Iwanczyk; Einar Nygard; Nail Malakhov; William C Barber; Thulasidharan Gandhi
Journal:  IEEE Trans Nucl Sci       Date:  2009-08-01       Impact factor: 1.679

View more
  2 in total

1.  Beyond noise power in 3D computed tomography: the local NPS and off-diagonal elements of the Fourier domain covariance matrix.

Authors:  Angel R Pineda; Daniel J Tward; Antonio Gonzalez; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

2.  Impact of bowtie filter and object position on the two-dimensional noise power spectrum of a clinical MDCT system.

Authors:  Daniel Gomez-Cardona; Juan Pablo Cruz-Bastida; Ke Li; Adam Budde; Jiang Hsieh; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.