Characterization of imaging performance in differential phase contrast CT compared with the conventional CT: spectrum of noise equivalent quanta NEQ(k).

Xiangyang Tang1, Yi Yang, Shaojie Tang.   

Abstract

PURPOSE: Differential phase contrast CT (DPC-CT) is emerging as a new technology to improve the contrast sensitivity of conventional attenuation-based CT. The noise equivalent quanta as a function over spatial frequency, i.e., the spectrum of noise equivalent quanta NEQ(k), is a decisive indicator of the signal and noise transfer properties of an imaging system. In this work, we derive the functional form of NEQ(k) in DPC-CT. Via system modeling, analysis, and computer simulation, we evaluate and verify the derived NEQ(k) and compare it with that of the conventional attenuation-based CT.
METHODS: The DPC-CT is implemented with x-ray tube and gratings. The x-ray propagation and data acquisition are modeled and simulated through Fresnel and Fourier analysis. A monochromatic x-ray source (30 keV) is assumed to exclude any system imperfection and interference caused by scatter and beam hardening, while a 360° full scan is carried out in data acquisition to avoid any weighting scheme that may disrupt noise randomness. Adequate upsampling is implemented to simulate the x-ray beam's propagation through the gratings G(1) and G(2) with periods 8 and 4 μm, respectively, while the intergrating distance is 193.6 mm (1∕16 of the Talbot distance). The dimensions of the detector cell for data acquisition are 32 × 32, 64 × 64, 96 × 96, and 128 × 128 μm(2), respectively, corresponding to a 40.96 × 40.96 mm(2) field of view in data acquisition. An air phantom is employed to obtain the noise power spectrum NPS(k), spectrum of noise equivalent quanta NEQ(k), and detective quantum efficiency DQE(k). A cylindrical water phantom at 5.1 mm diameter and complex refraction coefficient n = 1 - δ + iβ = 1 -2.5604 × 10(-7) + i1.2353 × 10(-10) is placed in air to measure the edge transfer function, line spread function and then modulation transfer function MTF(k), of both DPC-CT and the conventional attenuation-based CT. The x-ray flux is set at 5 × 10(6) photon∕cm(2) per projection and observes the Poisson distribution, which is consistent with that of a micro-CT for preclinical applications. Approximately 360 regions, each at 128 × 128 matrix, are used to calculate the NPS(k) via 2D Fourier transform, in which adequate zero padding is carried out to avoid aliasing in noise.
RESULTS: The preliminary data show that the DPC-CT possesses a signal transfer property [MTF(k)] comparable to that of the conventional attenuation-based CT. Meanwhile, though there exists a radical difference in their noise power spectrum NPS(k) (trait 1∕|k| in DPC-CT but |k| in the conventional attenuation-based CT) the NEQ(k) and DQE(k) of DPC-CT and the conventional attenuation-based CT are in principle identical.
CONCLUSIONS: Under the framework of ideal observer study, the joint signal and noise transfer property NEQ(k) and detective quantum efficiency DQE(k) of DPC-CT are essentially the same as those of the conventional attenuation-based CT. The findings reported in this paper may provide insightful guidelines on the research, development, and performance optimization of DPC-CT for extensive preclinical and clinical applications in the future.

Entities:  

Mesh:

Year:  2012        PMID: 22830779      PMCID: PMC3396709          DOI: 10.1118/1.4730287

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

1.  Radiation dose efficiency comparison between differential phase contrast CT and conventional absorption CT.

Authors:  Joseph Zambelli; Nicholas Bevins; Zhihua Qi; Guang-Hong Chen
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Fractional Montgomery effect: a self-imaging phenomenon.

Authors:  Adolf W Lohmann; Hans Knuppertz; Jürgen Jahns
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2005-08       Impact factor: 2.129

3.  A three-dimensional-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT-helical scanning.

Authors:  Xiangyang Tang; Jiang Hsieh; Roy A Nilsen; Sandeep Dutta; Dmitry Samsonov; Akira Hagiwara
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

4.  Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction.

Authors:  Jongduk Baek; Norbert J Pelc
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  Analytical evaluation of the signal and noise propagation in x-ray differential phase-contrast computed tomography.

Authors:  Rainer Raupach; Thomas G Flohr
Journal:  Phys Med Biol       Date:  2011-03-14       Impact factor: 3.609

6.  Characterization of imaging performance in differential phase contrast CT compared with the conventional CT--noise power spectrum NPS(k).

Authors:  Xiangyang Tang; Yi Yang; Shaojie Tang
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  Measurement of the PSF for a CT scanner: appropriate wire diameter and pixel size.

Authors:  E L Nickoloff
Journal:  Phys Med Biol       Date:  1988-01       Impact factor: 3.609

Review 8.  Toward consensus on quantitative assessment of medical imaging systems.

Authors:  C E Metz; R F Wagner; K Doi; D G Brown; R M Nishikawa; K J Myers
Journal:  Med Phys       Date:  1995-07       Impact factor: 4.071

9.  Absolute measures of physical image quality: measurement and application to radiographic magnification.

Authors:  J M Sandrik; R F Wagner
Journal:  Med Phys       Date:  1982 Jul-Aug       Impact factor: 4.071

10.  Detectability in computed tomographic images.

Authors:  K M Hanson
Journal:  Med Phys       Date:  1979 Sep-Oct       Impact factor: 4.071

View more
  16 in total

1.  A general theory of interference fringes in x-ray phase grating imaging.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

2.  Predicting fringe visibility in dual-phase grating interferometry with polychromatic X-ray sources.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2020       Impact factor: 1.535

3.  Polychromatic X-ray effects on fringe phase shifts in grating interferometry.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Opt Express       Date:  2017-03-20       Impact factor: 3.894

4.  Quantitative theory of X-ray interferometers based on dual phase grating: Fringe period and visibility.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Opt Express       Date:  2018-09-03       Impact factor: 3.894

5.  Fundamental relationship between the noise properties of grating-based differential phase contrast CT and absorption CT: theoretical framework using a cascaded system model and experimental validation.

Authors:  Ke Li; Nicholas Bevins; Joseph Zambelli; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  High energy x-ray phase contrast CT using glancing-angle grating interferometers.

Authors:  A Sarapata; J W Stayman; M Finkenthal; J H Siewerdsen; F Pfeiffer; D Stutman
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

7.  Predicting visibility of interference fringes in X-ray grating interferometry.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Opt Express       Date:  2016-07-11       Impact factor: 3.894

8.  High Energy X-ray Phase-Contrast Imaging Using Glancing Angle Grating Interferometers.

Authors:  D Stutman; J W Stayman; M Finkenthal; J H Siewerdsen
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-19

9.  Beam hardening correction in polychromatic x-ray grating interferometry.

Authors:  Aimin Yan; Xizeng Wu; Hong Liu
Journal:  Opt Express       Date:  2017-10-02       Impact factor: 3.894

10.  Spatial resolution characterization of differential phase contrast CT systems via modulation transfer function (MTF) measurements.

Authors:  Ke Li; Joseph Zambelli; Nicholas Bevins; Yongshuai Ge; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2013-05-17       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.