Youn Yi Jo1, Jong Wook Song, Young Chul Yoo, Ji Young Park, Jae Kwang Shim, Young Lan Kwak.
Abstract
BACKGROUND: We compared the continuous cardiac index measured by the FloTrac/Vigileo™ system (FCI) to that measured by a pulmonary artery catheter (CCI) with emphasis on the accuracy of the FCI in patients with a decreased left ventricular ejection fraction (LVEF) and a low cardiac output status during off-pump coronary bypass surgery (OPCAB). We also assessed the influence of several factors affecting the pulse contour, such as the mean arterial pressure (MAP), the systemic vascular resistance index (SVRI) and the use of norepinephrine.Entities:
Keywords: Cardiac output; Measurement techniques; Off pump coronary artery bypass
Year: 2011 PMID: 21602972 PMCID: PMC3092957 DOI: 10.4097/kjae.2011.60.4.237
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Patients' Characteristics
Values are medians (range), means (SD) or number of patients. RAS blocker: rennin-angiotensin system antagonist.
Hemodynamic Data
The table presents the hemodynamic data. The data is presented as means ± SDs. T1: 15 min after induction of anesthesia, T2: during Y-graft construction with opened pericardium, T3: 5 min after applying stabilizer for creating an obtuse marginalis branch anastomosis, T4: 15 min after completion of grafting with opened pericardium, T5: 15 min after sternum closure, HR: heart rate, MAP: mean arterial pressure, CVP: central venous pressure, SvO2: mixed venous oxygen saturation, CCI: continuous cardiac index measured by a pulmonary artery catheter, FCI: cardiac index measured by the FloTrac/Vigileo™ system, SVRI: systemic vascular resistance index. *P < 0.05 compared to the values at T1.
Bland-Altman Analyses of the Cardiac Index
The table presents the Bland-Altman analyses of the cardiac index as measured by the FloTrac/Vigileo™ system and the pulmonary artery catheter. T1: 15 min after induction of anesthesia, T2: during Y-graft construction with opened pericardium, T3: 5 min after applying stabilizer for creating an obtuse marginalis branch anastomosis, T4: 15 min after completion of grafting with opened pericardium, T5: 15 min after sternum closure.
Fig. 1Modified Bland-Altman analysis for repeated measures of agreement between the continuous cardiac index as measured by the FloTrac/Vigileo™ system (FCI) and a pulmonary artery catheter (CCI). (A) Overall, 250 data pairs, (B) 20 patients with a left ventricular ejection fraction < 40%, 100 data pairs, (C) 30 patients with left ventricular ejection fraction ≥ 40%, 150 data pairs. The unit of bias and the limit of agreement are L/min/m2.
Subgroup Bland-Altman Analyses of the Cardiac Index
The table presents the subgroup Bland-Altman analyses of the cardiac index as measured by the FloTrac/Vigileo™ system (FCI, L/min/m2) and the pulmonary artery catheter (CCI, L/min/m2) according to the left ventricular ejection fraction (LVEF, %), the CCI, the mean arterial pressure (MAP, mmHg), the systemic vascular resistance index (SVRI, dyne·s/cm5/m2) and the use of norepinephrine (NE). n corresponds to the number of obtained data sets. FCI and CCI are presented as means ± SDs. The unit of bias and the limit of agreement are L/min/m2.