Evaluation of the accuracy and response time of STAT-mode continuous cardiac output.

M A Lazor1, E T Pierce, G D Stanley, J L Cass, E F Halpern, R H Bode.   

Abstract

OBJECTIVES: This study was conducted to compare continuous cardiac output (CCO) with bolus thermodilution cardiac output (BTD) at steady state, and to compare the response time of STAT CCO with that of trend CCO, mean arterial pressure, and mixed venous oxygen saturation [SvO2] during an acute hemodynamic change.
DESIGN: Prospective study.
SETTING: University hospital. PARTICIPANTS: Twenty-nine patients undergoing cardiac surgery or liver transplantation.
INTERVENTIONS: STAT and trend CCO were compared with BTD cardiac output during steady state intraoperatively and postoperatively in the intensive care unit. Ten patients, who required epicardial pacing after cardiac surgery, were studied to compare the response time of STAT CCO with that of trend CCO, mean arterial pressure, and BvO2 after a 10% to 20% increase in pacing rate.
MEASUREMENTS AND MAIN RESULTS: A total of 108 cardiac output data sets were analyzed at steady state. Steady state was defined as stable heart rate and mean arterial pressure (+/- 5%) and stable central venous pressure (+/- 2 mmHg) measured immediately before and after each data set. Cardiac output ranged from 2.3 to 8.5 L/min. The correlation between STAT CCO and BTD was r = 0.94, and for trend CCO and BTD was r = 0.94. The bies and precision for STAT CCO versus BTD were 0.06 L/min (Cl 95%: -0.08 to 0.18) and 0.61 L/min. The bias and precision for trend CCO versus BTD were 0.06 L/min (Cl 95%: -0.04 to 0.16) and 0.49 L/min. Eleven data sets were analyzed to study response time of STAT CCO, which was defined as the first time the percent change of the mean of each variable was significantly increased from baseline. Significant increases in mean arterial pressure and SvO2 were detected after 30 seconds (2.5%, p = 0.01) and 90 seconds (2.0%, p = 0.04), respectively. A significant increase in STAT CCO was reached at 270 seconds (4.4%, p = 0.005). Trend CCO tended to increase but did not reach statistical significance within 6 minutes.
CONCLUSIONS: STAT and trend CCO are accurate and precise and show close agreement with BTD cardiac output at steady state. The faster algorithm of STAT CCO offers some advantage over trend CCO during an acute hemodynamic change. However, because of the averaging process for determining CCO, the response time of STAT CCO is slower than that of mean arterial pressure and SvO2.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9187990     DOI: 10.1016/s1053-0770(97)90050-1

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  8 in total

1.  Response time of the Opti-Q continuous cardiac output pulmonary artery catheter in the urgent mode to a step change in cardiac output.

Authors:  L J Goldstein
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

2.  A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors.

Authors:  Hironori Ishihara; Hirobumi Okawa; Ken Tanabe; Toshihito Tsubo; Yoshihiro Sugo; Takeshi Akiyama; Sunao Takeda
Journal:  J Clin Monit Comput       Date:  2004-12       Impact factor: 2.502

3.  Dynamic device properties of pulse contour cardiac output during transcatheter aortic valve implantation.

Authors:  Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter
Journal:  J Clin Monit Comput       Date:  2014-10-30       Impact factor: 2.502

4.  Letter to the editor in response to "Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis".

Authors:  Fellery de Lange; Inge T Bootsma; E Christiaan Boerma
Journal:  Crit Care       Date:  2021-06-21       Impact factor: 9.097

5.  The uncalibrated pulse contour cardiac output during off-pump coronary bypass surgery: performance in patients with a low cardiac output status and a reduced left ventricular function.

Authors:  Youn Yi Jo; Jong Wook Song; Young Chul Yoo; Ji Young Park; Jae Kwang Shim; Young Lan Kwak
Journal:  Korean J Anesthesiol       Date:  2011-04-26

6.  Continuously assessed right ventricular end-diastolic volume as a marker of cardiac preload and fluid responsiveness in mechanically ventilated cardiac surgical patients.

Authors:  Christoph Wiesenack; Christoph Fiegl; Andreas Keyser; Sven Laule; Christopher Prasser; Cornelius Keyl
Journal:  Crit Care       Date:  2005-04-01       Impact factor: 9.097

7.  Efficacy of Goal-Directed Therapy Using Bioreactance Cardiac Output Monitoring after Valvular Heart Surgery.

Authors:  Sak Lee; Seung Hyun Lee; Byung-Chul Chang; Jae-Kwang Shim
Journal:  Yonsei Med J       Date:  2015-07       Impact factor: 2.759

Review 8.  The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications.

Authors:  I T Bootsma; E C Boerma; T W L Scheeren; F de Lange
Journal:  J Clin Monit Comput       Date:  2021-03-01       Impact factor: 2.502

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.