Practical patient dosimetry for partial rotation cone beam CT.

E C Podnieks1, I S Negus.   

Abstract

OBJECTIVES: This work investigates the validity of estimating effective dose for cone beam CT (CBCT) exposures from the weighted CT dose index (CTDIW) and irradiated length.
METHODS: Measurements were made within cylindrical poly(methyl methacrylate) (PMMA) phantoms measuring 14 cm and 28 cm in length and 32 cm in diameter for the 200° DynaCT acquisition on the Siemens Artis zee fluoroscopy unit (Siemens Medical Solutions, Erlangen, Germany). An interpolated average dose was calculated to account for the partial rotation. Organ and effective doses were estimated by modelling projections in the Monte Carlo software programme PCXMC (STUK, Helsinki, Finland).
RESULTS: The CTDIW was found to closely approximate the interpolated average dose if the positions of the measured doses reflected the X-ray beam rotation. The average dose was found to increase by 8% when the phantom length was increased from 14 to 28 cm. Using the interpolated average dose and the irradiated length for effective dose calculations gave similar values to PCXMC when a double-length (28-cm) CT dose index phantom was irradiated. Simplifying the estimation of effective dose with PCXMC by modelling just 4 projections around the abdomen gave effective doses that were only 7% different to those given when 41 projections were modelled. Calculated doses to key organs within the beam varied by as much as 27%.
CONCLUSION: Estimating effective dose from the CTDIW and the irradiated length is sufficiently accurate for CBCT if the chamber positions are considered carefully. A conversion factor can be used only if a single CT dose index phantom is available. The estimation of organ doses requires a large number of modelled projections in PCXMC.

Entities:  

Mesh:

Year:  2011        PMID: 21304011      PMCID: PMC3473941          DOI: 10.1259/bjr/18118287

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  8 in total

1.  Dose and image quality for a cone-beam C-arm CT system.

Authors:  Rebecca Fahrig; Robert Dixon; Thomas Payne; Richard L Morin; Arundhuti Ganguly; Norbert Strobel
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

2.  Estimation of organ and effective doses resulting from cone beam CT imaging for radiotherapy treatment planning.

Authors:  L J Sawyer; S A Whittle; E S Matthews; H C Starritt; T P Jupp
Journal:  Br J Radiol       Date:  2009-03-02       Impact factor: 3.039

3.  Converting dose-length product to effective dose at CT.

Authors:  Walter Huda; Kent M Ogden; Mohammad R Khorasani
Journal:  Radiology       Date:  2008-09       Impact factor: 11.105

4.  Concepts for dose determination in flat-detector CT.

Authors:  Yiannis Kyriakou; Paul Deak; Oliver Langner; Willi A Kalender
Journal:  Phys Med Biol       Date:  2008-06-13       Impact factor: 3.609

5.  Enlarged longitudinal dose profiles in cone-beam CT and the need for modified dosimetry.

Authors:  Shinichiro Mori; Masahiro Endo; Kanae Nishizawa; Takanori Tsunoo; Takahiko Aoyama; Hideaki Fujiwara; Kenya Murase
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

6.  Effective dose analysis of three-dimensional rotational angiography during catheter ablation procedures.

Authors:  J-Y Wielandts; K Smans; J Ector; S De Buck; H Heidbüchel; H Bosmans
Journal:  Phys Med Biol       Date:  2010-01-07       Impact factor: 3.609

7.  Imaging doses from the Elekta Synergy X-ray cone beam CT system.

Authors:  A Amer; T Marchant; J Sykes; J Czajka; C Moore
Journal:  Br J Radiol       Date:  2007-06       Impact factor: 3.039

8.  The trouble with CTD100.

Authors:  John M Boone
Journal:  Med Phys       Date:  2007-04       Impact factor: 4.071

  8 in total
  4 in total

1.  Comparison of air kerma between C-arm CT and 64-multidetector-row CT using a phantom.

Authors:  Seiki Hosokawa; Nobuyuki Kawai; Morio Sato; Hiroki Minamiguchi; Motoki Nakai; Kazuhiro Murotani; Tadayoshi Nishioku; Shintaro Shirai; Tetsuo Sonomura
Journal:  Jpn J Radiol       Date:  2014-02-20       Impact factor: 2.374

2.  Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition.

Authors:  Stephanie Leon
Journal:  J Appl Clin Med Phys       Date:  2017-06-06       Impact factor: 2.102

3.  Dose assessment in dental cone-beam computed tomography: Comparison of optically stimulated luminescence dosimetry with Monte Carlo method.

Authors:  Chena Lee; Jeongmin Yoon; Sang-Sun Han; Ji Yeon Na; Jeong-Hee Lee; Young Hyun Kim; Jae Joon Hwang
Journal:  PLoS One       Date:  2020-03-31       Impact factor: 3.240

4.  Estimation of the effective dose of dental cone-beam computed tomography using personal computer-based Monte Carlo software.

Authors:  Eun-Kyung Kim; Won-Jeong Han; Jin-Woo Choi; Bulgan Battulga
Journal:  Imaging Sci Dent       Date:  2018-03-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.