| Literature DB >> 28585724 |
Abstract
Mobile fluoroscopy (c-arm) units offering 3D image reconstruction are becoming more common in surgical settings. Although these images are "CT-like" and sometimes replace the postoperative CT, the acquisition is technically very different from a traditional CT acquisition. Dose assessment is complicated by a large beam width, automatic exposure rate control, and a rotation of less than 360°. The purpose of this work was to explore the impact of these factors on the volumetric dose calculation and to provide practical recommendations for clinical physicists assessing dose from these units using commonly available equipment. CTDIW was calculated using the IAEA method for dosimetry of wide beams and compared to scans of the 32-cm CTDI phantom using the full beam width and a 20-mm collimated beam width. The impact of the partial rotation on the CTDIW calculation was assessed by acquiring measurements at four and twelve positions on the phantom periphery. For the system tested, the CTDIW was calculated to be 16.1 mGy using the IAEA method with default clinical protocol. Results showed that measuring CTDIW with the full beam width or a collimated beam width alone resulted in CTDI values of 19.0 mGy and 19.5 mGy, respectively. Using four peripheral measurements instead of 12 resulted in a difference of 4% for a collimated beam and 6% for an open beam. Variations in positioning on the order of a few centimeters resulted in a variation of only 4% with an open beam. The excellent reproducibility of the measurements using the full beam width suggests that this simple method is adequate for year-to-year comparisons. In contrast, the IAEA method is difficult to employ, particularly with 180° acquisitions. Use of peripheral measurements in excess of the usual four is time-consuming and not necessary for most applications obtained with the geometry specific to this system.Entities:
Keywords: zzm321990CTDIzzm321990; 3D fluoroscopy; cone beam CT; mobile c-arm; volumetric dose
Mesh:
Year: 2017 PMID: 28585724 PMCID: PMC5874898 DOI: 10.1002/acm2.12108
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Postsurgical sagittal reconstruction of patient hardware from the (a) Ziehm c‐arm and (b) multislice CT scanner.
Figure 2The centering of the phantom or chamber at isocenter was verified from the reconstructed image. This image demonstrates a slight misalignment of the pencil chamber, indicated by the location of the chamber cross‐section superior to the isocenter crosshairs. This image was acquired with the chamber free‐in‐air, but it is equally visible in the phantom.
Figure 3Photograph of the lead pieces used to collimate the x‐ray beam.
Figure 4The experimental set‐up, showing (a) a diagram of the positions at which the measurements were acquired and (b) a photograph of the phantom with ion chambers and angle indicator in place.
CTDIW measurements made, with 95% confidence intervals. The number of positions indicated refers to the number of measurement points at the phantom periphery: either 4 (spaced every 90°) or 12 (spaced every 30°). Trials 1–3 with the open beam involved repositioning the phantom and c‐arm between trials, to assess the impact of small differences in positioning
| Measurement method | CTDIW (mGy) |
|---|---|
| IAEA method | 16.1 (10.6, 21.6) |
| Open beam in phantom (12 positions) | 19.0 (17.3, 20.8) |
| Collimated beam in phantom (12 positions) | 19.5 (17.9, 21.0) |
| Open beam trial #1 (4 positions) | 21.7 (19.9, 23.4) |
| Open beam trial #2 (4 positions) | 21.5 (19.8, 23.3) |
| Open beam trial #3 (4 positions) | 20.1 (18.4, 21.8) |
| 12 positions (collimated) | 19.5 (17.9, 21.0) |
| 4 positions (collimated) | 20.2 (18.7, 21.8) |
| 12 positions (open) | 19.0 (17.3, 20.8) |
| 4 positions (open) | 20.1 (18.4, 21.8) |
Figure 5Phantom dose measurements (mGy) as a function of angle in the phantom, using a 20‐mm collimated beam width.