Concepts for dose determination in flat-detector CT.

Yiannis Kyriakou1, Paul Deak, Oliver Langner, Willi A Kalender.   

Abstract

Flat-detector computed tomography (FD-CT) scanners provide large irradiation fields of typically 200 mm in the cranio-caudal direction. In consequence, dose assessment according to the current definition of the computed tomography dose index CTDI(L=100 mm), where L is the integration length, would demand larger ionization chambers and phantoms which do not appear practical. We investigated the usefulness of the CTDI concept and practical dosimetry approaches for FD-CT by measurements and Monte Carlo (MC) simulations. An MC simulation tool (ImpactMC, VAMP GmbH, Erlangen, Germany) was used to assess the dose characteristics and was calibrated with measurements of air kerma. For validation purposes measurements were performed on an Axiom Artis C-arm system (Siemens Medical Solutions, Forchheim, Germany) equipped with a flat detector of 40 cm x 30 cm. The dose was assessed for 70 kV and 125 kV in cylindrical PMMA phantoms of 160 mm and 320 mm diameter with a varying phantom length from 150 to 900 mm. MC simulation results were compared to the values obtained with a calibrated ionization chambers of 100 mm and 250 mm length and to thermoluminesence (TLD) dose profiles. The MCs simulations were used to calculate the efficiency of the CTDI(L) determination with respect to the desired CTDI(infinity). Both the MC simulation results and the dose distributions obtained by MC simulation were in very good agreement with the CTDI measurements and with the reference TLD profiles, respectively, to within 5%. Standard CTDI phantoms which have a z-extent of 150 mm underestimate the dose at the center by up to 55%, whereas a z-extent of 600 mm appears to be sufficient for FD-CT; the baseline value of the respective profile was within 1% to the reference baseline. As expected, the measurements with ionization chambers of 100 mm and 250 mm offer a limited accuracy, whereas an increased integration length of 600 mm appeared to be necessary to approximate CTDI(infinity) in within 1%. MC simulations appear to offer a practical and accurate way of assessing conversion factors for arbitrary dosimetry setups using a standard pencil chamber to provide estimates of CTDI(infinity). This would eliminate the need for extra-long phantoms and ionization chambers or excessive amounts of TLDs.

Entities:  

Mesh:

Year:  2008        PMID: 18552416     DOI: 10.1088/0031-9155/53/13/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  8 in total

1.  Volume-of-interest imaging of the inner ear in a human temporal bone specimen using a robot- driven C-arm flat panel detector CT system.

Authors:  D Kolditz; T Struffert; Y Kyriakou; A Bozzato; A Dörfler; W A Kalender
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-11       Impact factor: 3.825

Review 2.  [Basic principles of flat detector computed tomography (FD-CT)].

Authors:  Y Kyriakou; T Struffert; A Dörfler; W A Kalender
Journal:  Radiologe       Date:  2009-09       Impact factor: 0.635

3.  Dose exposure of patients undergoing comprehensive stroke imaging by multidetector-row CT: comparison of 320-detector row and 64-detector row CT scanners.

Authors:  S Diekmann; E Siebert; R Juran; M Roll; W Deeg; H-C Bauknecht; F Diekmann; R Klingebiel; G Bohner
Journal:  AJNR Am J Neuroradiol       Date:  2010-01-28       Impact factor: 3.825

4.  The value of flat-detector computed tomography during catheterisation of congenital heart disease.

Authors:  Martin Glöckler; Andreas Koch; Verena Greim; Amira Shabaiek; Andrè Rüffer; Robert Cesnjevar; Stephan Achenbach; Sven Dittrich
Journal:  Eur Radiol       Date:  2011-07-28       Impact factor: 5.315

5.  Practical patient dosimetry for partial rotation cone beam CT.

Authors:  E C Podnieks; I S Negus
Journal:  Br J Radiol       Date:  2011-02-08       Impact factor: 3.039

6.  Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition.

Authors:  Stephanie Leon
Journal:  J Appl Clin Med Phys       Date:  2017-06-06       Impact factor: 2.102

7.  Evaluation of cumulative dose for cone-beam computed tomography (CBCT) scans within phantoms made from different compositions using Monte Carlo simulations.

Authors:  Abdullah Abuhaimed; Colin J Martin; Marimuthu Sankaralingam; Kurian Oomen; David J Gentle
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

8.  Diagnostic value of a spiral breast computed tomography system equipped with photon counting detector technology in patients with implants: An observational study of our initial experiences.

Authors:  Lisa Ruby; Sojin Shim; Nicole Berger; Magda Marcon; Thomas Frauenfelder; Andreas Boss
Journal:  Medicine (Baltimore)       Date:  2020-07-24       Impact factor: 1.817

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.