Imaging doses from the Elekta Synergy X-ray cone beam CT system.

A Amer1, T Marchant, J Sykes, J Czajka, C Moore.   

Abstract

The Elekta Synergy is a radiotherapy treatment machine with integrated kilovoltage (kV) X-ray imaging system capable of producing cone beam CT (CBCT) images of the patient in the treatment position. The aim of this study is to assess the additional imaging dose. Cone beam CT dose index (CBDI) is introduced and measured inside standard CTDI phantoms for several sites (head: 100 kV, 38 mAs, lung: 120 kV, 152 mAs and pelvis: 130 kV, 456 mAs). The measured weighted doses were compared with thermoluminescent dosimeter (TLD) measurements at various locations in a Rando phantom and at patients' surfaces. The measured CBDIs in-air at the isocentre were 9.2 mGy 100 mAs(-1), 7.3 mGy 100 mAs(-1) and 5.3 mGy 100 mAs(-1) for 130 kV, 120 kV and 100 kV, respectively. The body phantom weighted CBDI were 5.5 mGy 100 mAs(-1) and 3.8 mGy 100 mAs(-1 )for 130 kV and 120 kV. The head phantom weighted CBDI was 4.3 mGy 100 mAs(-1) for 100 kV. The weighted doses for the Christie Hospital CBCT imaging techniques were 1.6 mGy, 6 mGy and 22 mGy for the head, lung and pelvis. The measured CBDIs were used to estimate the total effective dose for the Synergy system using the ImPACT CT Patient Dosimetry Calculator. Measured CBCT doses using the Christie Hospital protocols are low for head and lung scans whether compared with electronic portal imaging (EPI), commonly used for treatment verification, or single and multiple slice CT. For the pelvis, doses are similar to EPI but higher than CT. Repeated use of CBCT for treatment verification is likely and hence the total patient dose needs to be carefully considered. It is important to consider further development of low dose CBCT techniques to keep additional doses as low as reasonably practicable.

Entities:  

Mesh:

Year:  2007        PMID: 17684077     DOI: 10.1259/bjr/80446730

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  27 in total

1.  Dosimetric assessment of the exposure of radiotherapy patients due to cone-beam CT procedures.

Authors:  Mariana Baptista; Salvatore Di Maria; Sandra Vieira; Joana Santos; Joana Pereira; Miguel Pereira; Pedro Vaz
Journal:  Radiat Environ Biophys       Date:  2018-11-03       Impact factor: 1.925

2.  Dosimetric properties and commissioning of cone-beam CT image beam line with a carbon target.

Authors:  Y Dzierma; F G Nuesken; N P Licht; C Ruebe
Journal:  Strahlenther Onkol       Date:  2013-05-30       Impact factor: 3.621

Review 3.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

4.  Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy.

Authors:  B Loutfi-Krauss; J Köhn; N Blümer; K Freundl; T Koch; E Kara; C Scherf; C Rödel; U Ramm; J Licher
Journal:  Strahlenther Onkol       Date:  2014-09-20       Impact factor: 3.621

5.  [Organizational models of inpatient emergency and acute care].

Authors:  A Kaltwasser; C Dodt
Journal:  Med Klin Intensivmed Notfmed       Date:  2014-10       Impact factor: 0.840

6.  Image quality and dose distributions of three linac-based imaging modalities.

Authors:  Yvonne Dzierma; Evemarie Ames; Frank Nuesken; Jan Palm; Norbert Licht; Christian Rübe
Journal:  Strahlenther Onkol       Date:  2014-12-20       Impact factor: 3.621

7.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

8.  4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction.

Authors:  Cong Zhao; Yuncheng Zhong; Xinhui Duan; You Zhang; Xiaokun Huang; Jing Wang; Mingwu Jin
Journal:  Phys Med Biol       Date:  2018-05-29       Impact factor: 3.609

9.  Improved cone-beam computed tomography in supine and prone breast radiotherapy. Surface reconstruction, radiation exposure, and clinical workflow.

Authors:  A De Puysseleyr; T Mulliez; A Gulyban; E Bogaert; T Vercauteren; T Van Hoof; J Van de Velde; R Van Den Broecke; C De Wagter; W De Neve
Journal:  Strahlenther Onkol       Date:  2013-10-03       Impact factor: 3.621

10.  Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.

Authors:  Jürgen Wilbert; Matthias Guckenberger; Bülent Polat; Otto Sauer; Michael Vogele; Michael Flentje; Reinhart A Sweeney
Journal:  Radiat Oncol       Date:  2010-05-26       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.