Is metal artefact reduction mandatory in cardiac PET/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?

Pardis Ghafarian1, S M R Aghamiri, Mohammad R Ay, Arman Rahmim, Thomas H Schindler, Osman Ratib, Habib Zaidi.   

Abstract

PURPOSE: Cardiac PET/CT imaging is often performed in patients with pacemakers and implantable cardioverter defibrillator (ICD) leads. However, metallic implants usually produce artefacts on CT images which might propagate to CT-based attenuation-corrected (CTAC) PET images. The impact of metal artefact reduction (MAR) for CTAC of cardiac PET/CT images in the presence of pacemaker, ICD and ECG leads was investigated using both qualitative and quantitative analysis in phantom and clinical studies.
METHODS: The study included 14 patients with various leads undergoing perfusion and viability examinations using dedicated cardiac PET/CT protocols. The PET data were corrected for attenuation using both artefactual CT images and CT images corrected using the MAR algorithm. The severity and magnitude of metallic artefacts arising from these leads were assessed on both linear attenuation coefficient maps (μ-maps) and attenuation-corrected PET images. CT and PET emission data were obtained using an anthropomorphic thorax phantom and a dedicated heart phantom made in-house incorporating pacemaker and ICD leads attached at the right ventricle of the heart. Volume of interest-based analysis and regression plots were performed for regions related to the lead locations. Bull's eye view analysis was also performed on PET images corrected for attenuation with and without the MAR algorithm.
RESULTS: In clinical studies, the visual assessment of PET images by experienced physicians and quantitative analysis did not reveal erroneous interpretation of the tracer distribution or significant differences when PET images were corrected for attenuation with and without MAR. In phantom studies, the mean differences between tracer uptake obtained without and with MAR were 10.16±2.1% and 6.86±2.1% in the segments of the heart in the vicinity of metallic ICD or pacemaker leads, and were 4.43±0.5% and 2.98±0.5% in segments far from the leads.
CONCLUSION: Although the MAR algorithm was able to effectively improve the quality of μ-maps, its clinical impact on the interpretation of PET images was not significant. Therefore cardiac PET images corrected for attenuation using CTAC in the presence of metallic leads can be interpreted without correction for metal artefacts. It should however be emphasized that in some special cases with multiple ICD leads attached to the myocardium wall, MAR might be useful for accurate attenuation correction.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20959974     DOI: 10.1007/s00259-010-1635-6

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  24 in total

1.  Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner.

Authors:  EhabI M Kamel; Cyrill Burger; Alfred Buck; GustavI K von Schulthess; GerhardI W Goerres
Journal:  Eur Radiol       Date:  2002-08-23       Impact factor: 5.315

2.  Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot.

Authors:  T G Flohr; K Stierstorfer; S Ulzheimer; H Bruder; A N Primak; C H McCollough
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

3.  An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: experimental and clinical studies.

Authors:  Mehran Yazdi; Mehran Yazdia; Luc Gingras; Luc Beaulieu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

4.  Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections.

Authors:  K Lance Gould; Tinsu Pan; Catalin Loghin; Nils P Johnson; Ashrith Guha; Stefano Sdringola
Journal:  J Nucl Med       Date:  2007-06-15       Impact factor: 10.057

Review 5.  Advances in Attenuation Correction Techniques in PET.

Authors:  Habib Zaidi; Marie-Louise Montandon; Abass Alavi
Journal:  PET Clin       Date:  2008-02-15

6.  Effective methods to correct contrast agent-induced errors in PET quantification in cardiac PET/CT.

Authors:  Florian Büther; Lars Stegger; Mohammad Dawood; Felix Range; Michael Schäfers; Roman Fischbach; Thomas Wichter; Otmar Schober; Klaus P Schäfers
Journal:  J Nucl Med       Date:  2007-06-15       Impact factor: 10.057

7.  Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?

Authors:  Frank P DiFilippo; Richard C Brunken
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

Review 8.  Clinical myocardial perfusion PET/CT.

Authors:  Marcelo F Di Carli; Sharmila Dorbala; Jolene Meserve; Georges El Fakhri; Arkadiusz Sitek; Stephen C Moore
Journal:  J Nucl Med       Date:  2007-05       Impact factor: 10.057

9.  The reduction of artifacts due to metal hip implants in CT-attenuation corrected PET images from hybrid PET/CT scanners.

Authors:  John A Kennedy; Ora Israel; Alex Frenkel; Rachel Bar-Shalom; Haim Azhari
Journal:  Med Biol Eng Comput       Date:  2007-05-23       Impact factor: 3.079

10.  Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device.

Authors:  Siroos Mirzaei; Michel Guerchaft; Christopher Bonnier; Peter Knoll; Michel Doat; Peter Braeutigam
Journal:  BMC Nucl Med       Date:  2005-06-14
View more
  7 in total

Review 1.  Recommendations for 18F-fluorodeoxyglucose positron emission tomography imaging for diagnosis of cardiac sarcoidosis-2018 update: Japanese Society of Nuclear Cardiology recommendations.

Authors:  Shinichiro Kumita; Keiichiro Yoshinaga; Masao Miyagawa; Mitsuru Momose; Keisuke Kiso; Tokuo Kasai; Masanao Naya
Journal:  J Nucl Cardiol       Date:  2019-08       Impact factor: 5.952

2.  Recommendations on nuclear and multimodality imaging in IE and CIED infections.

Authors:  Paola Anna Erba; Patrizio Lancellotti; Isidre Vilacosta; Oliver Gaemperli; Francois Rouzet; Marcus Hacker; Alberto Signore; Riemer H J A Slart; Gilbert Habib
Journal:  Eur J Nucl Med Mol Imaging       Date:  2018-05-24       Impact factor: 9.236

3.  Metal artefact reduction algorithms prevent false positive results when assessing patients for cardiac implantable electronic device infection.

Authors:  Fozia Z Ahmed; Jacqueline James; Deborah Tout; Parthiban Arumugam; Mamas Mamas; Amir M Zaidi
Journal:  J Nucl Cardiol       Date:  2014-08-12       Impact factor: 5.952

4.  The effect of metal artefact reduction on CT-based attenuation correction for PET imaging in the vicinity of metallic hip implants: a phantom study.

Authors:  Roy Harnish; Sven Prevrhal; Abass Alavi; Habib Zaidi; Thomas F Lang
Journal:  Ann Nucl Med       Date:  2014-04-08       Impact factor: 2.668

Review 5.  [Coronary computed tomography and cardiac devices : Diagnostic results or nothing but artifacts?]

Authors:  Silvia Smolka; Stephan Achenbach
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-05

Review 6.  Assessment of cardiac sarcoidosis with advanced imaging modalities.

Authors:  Makoto Orii; Toshio Imanishi; Takashi Akasaka
Journal:  Biomed Res Int       Date:  2014-08-28       Impact factor: 3.411

7.  Parameters Influencing PET Imaging Features: A Phantom Study with Irregular and Heterogeneous Synthetic Lesions.

Authors:  Francesca Gallivanone; Matteo Interlenghi; Daniela D'Ambrosio; Giuseppe Trifirò; Isabella Castiglioni
Journal:  Contrast Media Mol Imaging       Date:  2018-09-10       Impact factor: 3.161

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.