Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?

Frank P DiFilippo1, Richard C Brunken.   

Abstract

UNLABELLED: Artifacts related to metallic implants are an established limitation of CT-based attenuation correction (CT-AC) in PET/CT. However, the impact of metallic components of pacemaker leads and implantable cardioverter defibrillator (ICD) leads on the accuracy of cardiac PET has not been evaluated. The goal of this study was to investigate the magnitude of artifacts related to pacing and defibrillation leads in both phantom and patient studies.
METHODS: Images were acquired on a PET/CT scanner using CT-AC and were compared with those obtained on a dedicated PET scanner using transmission source-based attenuation correction. Phantoms consisting of pacemaker leads and ICD leads submerged in uniform background activity solution were imaged, and regions were analyzed to measure radionuclide concentrations at known lead locations relative to background. In addition, 15 cardiac 18F-FDG patients (having either pacing leads, defibrillation leads, or both) were imaged on both PET/CT and PET scanners. Images were visually and quantitatively assessed to determine whether artifact related to the implanted leads was present and, if so, its severity relative to surrounding myocardium.
RESULTS: In phantom studies, artifacts caused by pacing lead electrodes were barely noticeable, but artifacts arising from highly radioopaque ICD shock coil electrodes were clearly apparent. In the patient studies, no artifacts from pacing leads were identified. However, significant artifact was observed in 50% of the patient studies with ICD leads. In the affected areas, local myocardial uptake in PET/CT images using CT-AC was, on average, 30% higher than that in the corresponding PET images.
CONCLUSION: Although pacemaker leads do not appear to cause artifact in cardiac PET/CT images, ICD leads frequently do result in artifacts of sufficient magnitude to impact clinical image interpretation. Accordingly, software-based corrections in CT-AC algorithms appear necessary for accurate cardiac imaging with PET/CT.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15750156

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  27 in total

1.  Calcified Lymph Nodes Causing Clinically Relevant Attenuation Correction Artifacts on PET/CT Imaging.

Authors:  Amar Mehta; Ajeet Mehta; Charles Laymon; Todd M Blodgett
Journal:  J Radiol Case Rep       Date:  2010-02-01

2.  Comparative assessment of energy-mapping approaches in CT-based attenuation correction for PET.

Authors:  Mohammad R Ay; Maryam Shirmohammad; Saeed Sarkar; Arman Rahmim; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2011-02       Impact factor: 3.488

3.  Is metal artefact reduction mandatory in cardiac PET/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?

Authors:  Pardis Ghafarian; S M R Aghamiri; Mohammad R Ay; Arman Rahmim; Thomas H Schindler; Osman Ratib; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-10-20       Impact factor: 9.236

4.  Qualitative and quantitative assessment of metal artifacts arising from implantable cardiac pacing devices in oncological PET/CT studies: a phantom study.

Authors:  Mohammad R Ay; Abolfazl Mehranian; Mehrsima Abdoli; Pardis Ghafarian; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2011-12       Impact factor: 3.488

5.  Assessment of errors caused by X-ray scatter and use of contrast medium when using CT-based attenuation correction in PET.

Authors:  Mohammad Reza Ay; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-04-19       Impact factor: 9.236

6.  Infected pacemaker causing septic lung emboli detected on FDG PET/CT.

Authors:  Gad Abikhzer; Sophie Turpin; Jean-Luc Bigras
Journal:  J Nucl Cardiol       Date:  2010-06       Impact factor: 5.952

7.  Effects of MR surface coils on PET quantification.

Authors:  Lawrence R MacDonald; Steve Kohlmyer; Chi Liu; Thomas K Lewellen; Paul E Kinahan
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

Review 8.  Recommendations for 18F-fluorodeoxyglucose positron emission tomography imaging for diagnosis of cardiac sarcoidosis-2018 update: Japanese Society of Nuclear Cardiology recommendations.

Authors:  Shinichiro Kumita; Keiichiro Yoshinaga; Masao Miyagawa; Mitsuru Momose; Keisuke Kiso; Tokuo Kasai; Masanao Naya
Journal:  J Nucl Cardiol       Date:  2019-08       Impact factor: 5.952

Review 9.  PET/CT Myocardial Perfusion Imaging Acquisition and Processing: Ten Tips and Tricks to Help You Succeed.

Authors:  Talal Alnabelsi; Akanksha Thakkar; Ahmed Ibrahim Ahmed; Yushui Han; Mouaz H Al-Mallah
Journal:  Curr Cardiol Rep       Date:  2021-03-11       Impact factor: 2.931

Review 10.  Morphology supporting function: attenuation correction for SPECT/CT, PET/CT, and PET/MR imaging.

Authors:  Tzu C Lee; Adam M Alessio; Robert M Miyaoka; Paul E Kinahan
Journal:  Q J Nucl Med Mol Imaging       Date:  2015-11-17       Impact factor: 2.346

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.