Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

Tim J Wood1,2, Craig S Moore1,2, Carl J Horsfield1, John R Saunderson1,2, Andrew W Beavis1,2,3.   

Abstract

OBJECTIVE: The purpose of this study was to develop size-based radiotherapy kilovoltage cone beam CT (CBCT) protocols for the pelvis.
METHODS: Image noise was measured in an elliptical phantom of varying size for a range of exposure factors. Based on a previously defined "small pelvis" reference patient and CBCT protocol, appropriate exposure factors for small, medium, large and extra-large patients were derived which approximate the image noise behaviour observed on a Philips CT scanner (Philips Medical Systems, Best, Netherlands) with automatic exposure control (AEC). Selection criteria, based on maximum tube current-time product per rotation selected during the radiotherapy treatment planning scan, were derived based on an audit of patient size.
RESULTS: It has been demonstrated that 110 kVp yields acceptable image noise for reduced patient dose in pelvic CBCT scans of small, medium and large patients, when compared with manufacturer's default settings (125 kVp). Conversely, extra-large patients require increased exposure factors to give acceptable images. 57% of patients in the local population now receive much lower radiation doses, whereas 13% require higher doses (but now yield acceptable images).
CONCLUSION: The implementation of size-based exposure protocols has significantly reduced radiation dose to the majority of patients with no negative impact on image quality. Increased doses are required on the largest patients to give adequate image quality. ADVANCES IN KNOWLEDGE: The development of size-based CBCT protocols that use the planning CT scan (with AEC) to determine which protocol is appropriate ensures adequate image quality whilst minimizing patient radiation dose.

Entities:  

Mesh:

Year:  2015        PMID: 26419892      PMCID: PMC4743457          DOI: 10.1259/bjr.20150364

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  27 in total

1.  IGRT induced dose burden for a variety of imaging protocols at two different anatomical sites.

Authors:  Markus Stock; Asa Palm; Andreas Altendorfer; Elisabeth Steiner; Dietmar Georg
Journal:  Radiother Oncol       Date:  2011-11-16       Impact factor: 6.280

2.  Estimation of organ doses from kilovoltage cone-beam CT imaging used during radiotherapy patient position verification.

Authors:  Daniel E Hyer; David E Hintenlang
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

3.  Inclusion of the dose from kilovoltage cone beam CT in the radiation therapy treatment plans.

Authors:  Parham Alaei; George Ding; Huaiqun Guan
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

4.  Image-guided radiation therapy: what is our Utopia?

Authors:  A W Beavis
Journal:  Br J Radiol       Date:  2010-03       Impact factor: 3.039

5.  Daily kV cone-beam CT and deformable image registration as a method for studying dosimetric consequences of anatomic changes in adaptive IMRT of head and neck cancer.

Authors:  Ulrik Vindelev Elstrøm; Barbara A Wysocka; Ludvig P Muren; Jørgen B B Petersen; Cai Grau
Journal:  Acta Oncol       Date:  2010-10       Impact factor: 4.089

6.  Reducing radiation exposure to patients from kV-CBCT imaging.

Authors:  George X Ding; Peter Munro; Jason Pawlowski; Arnold Malcolm; Charles W Coffey
Journal:  Radiother Oncol       Date:  2010-09-16       Impact factor: 6.280

7.  Assessment of patient organ doses and effective doses using the VIP-Man adult male phantom for selected cone-beam CT imaging procedures during image guided radiation therapy.

Authors:  Jianwei Gu; Bryan Bednarz; X George Xu; Steve B Jiang
Journal:  Radiat Prot Dosimetry       Date:  2008-07-29       Impact factor: 0.972

8.  Simple methods to reduce patient dose in a Varian cone beam CT system for delivery verification in pelvic radiotherapy.

Authors:  P Roxby; T Kron; F Foroudi; A Haworth; C Fox; A Mullen; J Cramb
Journal:  Br J Radiol       Date:  2009-03-16       Impact factor: 3.039

9.  An organ and effective dose study of XVI and OBI cone-beam CT systems.

Authors:  Daniel E Hyer; Christopher F Serago; Siyong Kim; Jonathan G Li; David E Hintenlang
Journal:  J Appl Clin Med Phys       Date:  2010-04-17       Impact factor: 2.102

10.  Investigation of the usability of conebeam CT data sets for dose calculation.

Authors:  Anne Richter; Qiaoqiao Hu; Doreen Steglich; Kurt Baier; Jürgen Wilbert; Matthias Guckenberger; Michael Flentje
Journal:  Radiat Oncol       Date:  2008-12-16       Impact factor: 3.481

View more
  6 in total

1.  Patient-based low dose cone beam CT acquisition settings for prostate image-guided radiotherapy treatments on a Varian TrueBeam linear accelerator.

Authors:  Maria Antonietta Piliero; Margherita Casiraghi; Davide Giovanni Bosetti; Simona Cima; Letizia Deantonio; Stefano Leva; Francesco Martucci; Marino Tettamanti; Francesco Pupillo; Luca Bellesi; Antonella Richetti; Stefano Presilla
Journal:  Br J Radiol       Date:  2020-08-27       Impact factor: 3.039

2.  PCXMC cone beam computed tomography dosimetry investigations.

Authors:  Aaron Fetin; Lucy Cartwright; Jonathan Sykes; Alicja Wach
Journal:  Phys Eng Sci Med       Date:  2022-01-24

3.  Optimizing the target detectability of cone beam CT performed in image-guided radiation therapy for patients of different body sizes.

Authors:  Ching-Ching Yang; Pei-Chieh Yu; Jau-Ming Ruan; Yu-Cheng Chen
Journal:  J Appl Clin Med Phys       Date:  2018-03-08       Impact factor: 2.102

4.  Monte Carlo-based software for 3D personalized dose calculations in image-guided radiotherapy.

Authors:  Coralie Le Deroff; Lucie Berger; Julien Bellec; Guillaume Boissonnat; Héléna Chesneau; Sophie Chiavassa; Julie Desrousseaux; Stéphanie Gempp; Olivier Henry; Jimmy Jarril; Delphine Lazaro; Ronan Lefeuvre; Vincent Passal; Fanny Solinhac; Caroline Lafond; Gregory Delpon
Journal:  Phys Imaging Radiat Oncol       Date:  2022-02-26

5.  Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study.

Authors:  Sara Chan; Eileen Giles; Lyndal Newmarch; Michala Short
Journal:  J Med Radiat Sci       Date:  2019-10-16

6.  CBCT imaging: a simple approach for optimising and evaluating concomitant imaging doses, based on patient-specific attenuation, during radiotherapy pelvis treatment.

Authors:  Caroline Ordóñez-Sanz; Mark Cowen; Neda Shiravand; Niall D MacDougall
Journal:  Br J Radiol       Date:  2021-08-01       Impact factor: 3.039

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.