Radiation dose from cone beam computed tomography for image-guided radiation therapy.

Monica W K Kan1, Lucullus H T Leung, Wicger Wong, Nelson Lam.   

Abstract

PURPOSE: To perform a comprehensive study on organ absorbed doses and effective doses from cone beam computed tomography (CBCT) for three different treatment sites. METHODS AND MATERIALS: An extensive set of dosimetric measurements were performed using a widely used CBCT system, the On-Board Imager (OBI). Measurements were performed using a female anthropomorphic phantom with thermoluminescent dosimeters (TLD). The effective doses to the body and the absorbed doses to 26 organs were reported using two different technical settings, namely, the standard mode and the low-dose mode. The measurements were repeated for three different scan sites: head and neck, chest, and pelvis. Comparisons of patient doses as well as image quality were performed among the standard mode CBCT, low-dose mode CBCT, and fan beam CT.
RESULTS: The mean skin doses from standard mode CBCT to head and neck, chest and pelvis were 6.7, 6.4, and 5.4 cGy per scan, respectively. The effective doses to the body from standard mode CBCT for imaging of head and neck, chest, and pelvis were 10.3, 23.7, and 22.7 mSv per scan, respectively. Patient doses from low-dose mode CBCT were approximately one fifth of those from standard mode CBCT.
CONCLUSIONS: Patient position verification by standard mode CBCT acquired by OBI on a daily basis could increase the secondary cancer risk by up to 2% to 4%. Therefore lower mAs settings for daily CBCT should be considered, especially when bony anatomy is the main interest.

Entities:  

Mesh:

Year:  2007        PMID: 17980510     DOI: 10.1016/j.ijrobp.2007.08.062

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  43 in total

1.  Progressive cone beam CT dose control in image-guided radiation therapy.

Authors:  Hao Yan; Xin Zhen; Laura Cerviño; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

2.  Accelerated barrier optimization compressed sensing (ABOCS) reconstruction for cone-beam CT: phantom studies.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

3.  Reconstructing cone-beam CT with spatially varying qualities for adaptive radiotherapy: a proof-of-principle study.

Authors:  Wenting Lu; Hao Yan; Xuejun Gu; Zhen Tian; Ouyang Luo; Liu Yang; Linghong Zhou; Laura Cervino; Jing Wang; Steve Jiang; Xun Jia
Journal:  Phys Med Biol       Date:  2014-09-26       Impact factor: 3.609

4.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

Review 5.  Image-guided radiation therapy (IGRT): practical recommendations of Italian Association of Radiation Oncology (AIRO).

Authors:  Paola Franzone; Alba Fiorentino; Salvina Barra; Domenico Cante; Laura Masini; Elena Cazzulo; Liana Todisco; Pietro Gabriele; Elisabetta Garibaldi; Anna Merlotti; Maria Grazia Ruo Redda; Filippo Alongi; Renzo Corvò
Journal:  Radiol Med       Date:  2016-09-06       Impact factor: 3.469

6.  Multi-system verification of registrations for image-guided radiotherapy in clinical trials.

Authors:  Yunfeng Cui; James M Galvin; William L Straube; Walter R Bosch; James A Purdy; X Allen Li; Ying Xiao
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-01-13       Impact factor: 7.038

7.  Daily patient setup error in prostate image guided radiation therapy with fiducial-based kilovoltage onboard imaging and conebeam computed tomography.

Authors:  Jason C Ye; Muhammad M Qureshi; Pauline Clancy; Lauren N Dise; John Willins; Ariel E Hirsch
Journal:  Quant Imaging Med Surg       Date:  2015-10

Review 8.  Cone Beam CT vs. Fan Beam CT: A Comparison of Image Quality and Dose Delivered Between Two Differing CT Imaging Modalities.

Authors:  Lawrence Lechuga; Georg A Weidlich
Journal:  Cureus       Date:  2016-09-12

9.  Improved cone-beam computed tomography in supine and prone breast radiotherapy. Surface reconstruction, radiation exposure, and clinical workflow.

Authors:  A De Puysseleyr; T Mulliez; A Gulyban; E Bogaert; T Vercauteren; T Van Hoof; J Van de Velde; R Van Den Broecke; C De Wagter; W De Neve
Journal:  Strahlenther Onkol       Date:  2013-10-03       Impact factor: 3.621

10.  Recent advances in image-guided radiotherapy for head and neck carcinoma.

Authors:  Sameer K Nath; Daniel R Simpson; Brent S Rose; Ajay P Sandhu
Journal:  J Oncol       Date:  2009-07-29       Impact factor: 4.375

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.