BACKGROUND: To evaluate the accuracy of using kilovoltage x-ray cone-beam computed tomography (kV-CBCT) imaging for in vivo dose calculations.Entities:
Mesh:
Year: 2010 PMID: 20403191 PMCID: PMC2874569 DOI: 10.1186/1748-717X-5-29
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
The densities and CBCT numbers.
| Region of interest (ROI) | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CBCT Numbers (HU) | 0 | 1379 | 1500 | 1950 | 1990 | 2000 | 2103 | 2158 | 2468 | 2500 | 2670 | 3293 | 3847 |
| Density | 0.0 | 0.0 | 0.9 | 0.9 | 1.02 | 1.03 | 1.06 | 1.09 | 1.30 | 1.50 | 1.62 | 1.84 | 1.86 |
The densities and CBCT numbers for generating the CBCT calibration curve
Figure 1Calibration curves for kV CT and kV CBCT. The calibration curves for kV-CT and kV-CBCT based dose calculations in the treatment planning system.
Figure 2DVH displays of three clinical cases. The DVHs of three cases: one NK/T lymphoma (a), one NPC (b) and one Brain (c). The solid lines represent the dose based on conventional CT and the dash lines represent the dose based on kV CBCT.
Figure 3Dose distributions of three clinical cases. The transverse views of dose distributions of the NK/T lymphoma (a), NPC (b) and brain tumor (c). Left: calculated dose based on CBCT images; Right: calculated dose based on planning CT images. They show good agreement on both relative high and low isodoses.
The comparison of iso-plane dose distributions.
| Patients | Beam index | DTA | Gamma index analysis |
|---|---|---|---|
| NK/T | 1 | 99.0 | 100.0 |
| 2 | 100.0 | 100.0 | |
| 3 | 99.4 | 100.0 | |
| 4 | 99.5 | 100.0 | |
| 5 | 99.2 | 100.0 | |
| NPC | 1 | 97.2 | 99.5 |
| 2 | 98.8 | 99.8 | |
| 3 | 99.8 | 99.8 | |
| 4 | 98.9 | 100.0 | |
| 5 | 96.3 | 99.6 | |
| 6 | 94.5 | 99.9 | |
| 7 | 99.8 | 100.0 | |
| 8 | 99.5 | 100.0 | |
| 9 | 99.0 | 100.0 | |
| Brain Tumor | 1 | 99.7 | 100.0 |
| 2 | 99.8 | 100.0 | |
| 3 | 96.3 | 100.0 | |
| 4 | 100.0 | 100.0 | |
| 5 | 100.0 | 100.0 |
The comparison of iso-plane dose distributions based on conventional CT and KVCBCT for 3 clinical cases using distance to agreement (DTA) and gamma index analysis in all the beams.
Dose comparisons in targets and normal tissues.
| Patients | 1st kVCBCT (Gy) | kVCT | Difference | |
|---|---|---|---|---|
| NK-T lymphoma | 1PTV: mean dose | 51.03 | 51.28 | 0.49% |
| 1Right-eye: mean dose | 11.68 | 11.76 | 0.70% | |
| 1Left-eye: mean dose | 10.54 | 10.58 | 0.37% | |
| 1Spinal cord: max dose | 47.14 | 47.65 | 1.07% | |
| 1Whole body: max dose | 59.18 | 59.44 | -0.43% | |
| NPC | 2GTV: mean dose | 67.8 | 67.78 | -0.03% |
| 2CTV: mean dose | 64.9 | 64.89 | -0.02% | |
| 2PTV: mean dose | 63.76 | 63.78 | 0.03% | |
| 2Brain Stem: max dose | 53.83 | 54.36 | 0.98% | |
| 2Spinal cord: max dose | 32.85 | 32.73 | -0.34% | |
| 2L-parotid: mean dose | 34.92 | 34.11 | -2.37% | |
| 2R-parotid: mean dose | 33.38 | 32.91 | -1.41% | |
| 2Whole body: max dose | 70.73 | 70.12 | 0.87% | |
| Brain Tumor | 3CTV: mean dose | 61.85 | 61.84 | -0.02% |
| 3PTV: mean dose | 60.64 | 60.57 | -0.11% | |
| 3Brain: max dose | 50.13 | 50.02 | -0.22% | |
| 3Left-eye: mean dose | 19.42 | 19.4 | -0.13% | |
| 3Right-eye: mean dose | 9.94 | 9.64 | -3.15% | |
| 3Whole body: max dose | 67.21 | 67.66 | 0.66% | |
Dose comparisons of the first kVCBCT to the planning kVCT in targets and normal tissues for the three clinical cases (NK-T lymphoma, NPC and Brain Tumor). The difference of maximal dose in tumor is 0.49% and in normal structures are -3.15%. The numbers before the contours are the indexes of patients, 1PTV means the PTV in patient1.
Figure 4The transverse views of CT and the 2nd CBCT. The transverse views of the reference CT (a) and cone beam CT (b) after two weeks of the treatment. A slight change happened in the external contour and air cavity.
Dose comparisons of the first and second CBCT images.
| Mean dose (Gy) | Maximal dose (Gy) | |||||||
|---|---|---|---|---|---|---|---|---|
| GTV | CTV | PTV | Left-parotid | Right-parotid | Brain-stem | Spinal-cord | Whole-body | |
| kV CT | 50.72 | 51.04 | 51.02 | 32.14 | 30.40 | 50.92 | 31.76 | 55.78 |
| 1st kV CBCT | 50.23 | 50.77 | 50.74 | 31.88 | 30.57 | 51.21 | 31.17 | 55.63 |
| Dose difference between 1st kV CBCT and kV CT(%) | 0.98% | 0.54% | 0.54% | 0.8% | -0.56% | 0.57% | -1.87% | -0.26% |
| 2nd kV CBCT | 50.37 | 50.77 | 50.76 | 32.56 | 28.73 | 51.36 | 32.96 | 55.71 |
| Dose difference between 2nd kV CBCT and kV CT(%) | 0.7% | 0.53% | 0.52% | -1.3% | 5.81% | 0.86% | 3.77% | -0.11% |
Dose comparison of the first and second CBCTs to the reference planning kVCT in a patient (patient 4) with marked anatomical changes. The largest changes in dose difference between CBCT and planning CT are found in the spinal cord (from -1.87% to 3.77%) and in the right parotid (from -0.56% to 5.81%).