Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT: image quality and system performance.

Daniel Létourneau1, Rebecca Wong, Douglas Moseley, Michael B Sharpe, Stephen Ansell, Mary Gospodarowicz, David A Jaffray.   

Abstract

PURPOSE: To assess the feasibility of an online strategy for palliative radiotherapy (RT) of spinal bone metastasis, which integrates imaging, planning, and treatment delivery in a single step at the treatment unit. The technical challenges of this approach include cone-beam CT (CBCT) image quality for target definition, online planning, and efficient process integration. METHODS AND MATERIALS: An integrated imaging, planning, and delivery system was constructed and tested with phantoms. The magnitude of CBCT image artifacts following the use of an antiscatter grid and a nonlinear scatter correction was quantified using phantom data and images of patients receiving conventional palliative RT of the spine. The efficacy of online planning was then assessed using corrected CBCT images. Testing of the complete process was performed on phantoms with assessment of timing and dosimetric accuracy.
RESULTS: The use of image corrections reduced the cupping artifact from 30% to 4.5% on CBCT images of a body phantom and improved the accuracy of CBCT numbers (water: +/- 20 Hounsfield unit [HU], and lung and bone: to within +/- 130 HU). Bony anatomy was clearly visible and was deemed sufficient for target definition. The mean total time (n = 5) for application of the online approach was 23.1 min. Image-guided dose placement was assessed using radiochromic film measurements with good agreement (within 5% of dose difference and 2 mm of distance to agreement).
CONCLUSIONS: The technical feasibility of CBCT-guided online planning and delivery for palliative single treatment has been demonstrated. The process was performed in one session equivalent to an initial treatment slot (<30 min) with dosimetric accuracy satisfying accepted RT standards.

Entities:  

Mesh:

Year:  2007        PMID: 17336223     DOI: 10.1016/j.ijrobp.2006.09.058

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  20 in total

1.  Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom.

Authors:  H-L Chao; W-L Chen; C-C Hu; J-K Wu; C-J Wu; J C-H Cheng
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

2.  Validation of a deformable image registration technique for cone beam CT-based dose verification.

Authors:  M Moteabbed; G C Sharp; Y Wang; A Trofimov; J A Efstathiou; H-M Lu
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

3.  Target-specific optimization of four-dimensional cone beam computed tomography.

Authors:  Moiz Ahmad; Tinsu Pan
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

4.  Four-dimensional volume-of-interest reconstruction for cone-beam computed tomography-guided radiation therapy.

Authors:  Moiz Ahmad; Peter Balter; Tinsu Pan
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

5.  Use of kilovoltage X-ray volume imaging in patient dose calculation for head-and-neck and partial brain radiation therapy.

Authors:  Weigang Hu; Jinsong Ye; Jiazhou Wang; Xuejun Ma; Zhen Zhang
Journal:  Radiat Oncol       Date:  2010-04-19       Impact factor: 3.481

6.  DEGRO practice guidelines for palliative radiotherapy of metastatic breast cancer: bone metastases and metastatic spinal cord compression (MSCC).

Authors:  Rainer Souchon; Frederik Wenz; Felix Sedlmayer; Wilfried Budach; Jürgen Dunst; Petra Feyer; Wulf Haase; Wolfgang Harms; Marie-Luise Sautter-Bihl; Rolf Sauer
Journal:  Strahlenther Onkol       Date:  2009-08-28       Impact factor: 3.621

7.  Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study.

Authors:  Markus Stock; Marlies Pasler; Wolfgang Birkfellner; Peter Homolka; Richard Poetter; Dietmar Georg
Journal:  Radiother Oncol       Date:  2009-08-18       Impact factor: 6.280

8.  Demons deformable registration for CBCT-guided procedures in the head and neck: convergence and accuracy.

Authors:  S Nithiananthan; K K Brock; M J Daly; H Chan; J C Irish; J H Siewerdsen
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

9.  The effects of target motion in kV-CBCT imaging.

Authors:  Sriram Padmanaban; Raghavendiran Boopathy; Bhuvana Kunjithapatham; Prabakar Sukumar; Vivekanandan Nagarajan
Journal:  Pol J Radiol       Date:  2010-01

10.  Investigation of the usability of conebeam CT data sets for dose calculation.

Authors:  Anne Richter; Qiaoqiao Hu; Doreen Steglich; Kurt Baier; Jürgen Wilbert; Matthias Guckenberger; Michael Flentje
Journal:  Radiat Oncol       Date:  2008-12-16       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.