| Literature DB >> 16723023 |
Christoph Thilmann1, Simeon Nill, Thomas Tücking, Angelika Höss, Bernd Hesse, Lars Dietrich, Rolf Bendl, Bernhard Rhein, Peter Häring, Christian Thieke, Uwe Oelfke, Juergen Debus, Peter Huber.
Abstract
BACKGROUND: The purpose of the study was the clinical implementation of a kV cone beam CT (CBCT) for setup correction in radiotherapy. PATIENTS AND METHODS: For evaluation of the setup correction workflow, six tumor patients (lung cancer, sacral chordoma, head-and-neck and paraspinal tumor, and two prostate cancer patients) were selected. All patients were treated with fractionated stereotactic radiotherapy, five of them with intensity modulated radiotherapy (IMRT). For patient fixation, a scotch cast body frame or a vacuum pillow, each in combination with a scotch cast head mask, were used. The imaging equipment, consisting of an x-ray tube and a flat panel imager (FPI), was attached to a Siemens linear accelerator according to the in-line approach, i.e. with the imaging beam mounted opposite to the treatment beam sharing the same isocenter. For dose delivery, the treatment beam has to traverse the FPI which is mounted in the accessory tray below the multi-leaf collimator. For each patient, a predefined number of imaging projections over a range of at least 200 degrees were acquired. The fast reconstruction of the 3D-CBCT dataset was done with an implementation of the Feldkamp-David-Kress (FDK) algorithm. For the registration of the treatment planning CT with the acquired CBCT, an automatic mutual information matcher and manual matching was used. RESULTS AND DISCUSSION: Bony landmarks were easily detected and the table shifts for correction of setup deviations could be automatically calculated in all cases. The image quality was sufficient for a visual comparison of the desired target point with the isocenter visible on the CBCT. Soft tissue contrast was problematic for the prostate of an obese patient, but good in the lung tumor case. The detected maximum setup deviation was 3 mm for patients fixated with the body frame, and 6 mm for patients positioned in the vacuum pillow. Using an action level of 2 mm translational error, a target point correction was carried out in 4 cases. The additional workload of the described workflow compared to a normal treatment fraction led to an extra time of about 10-12 minutes, which can be further reduced by streamlining the different steps.Entities:
Mesh:
Year: 2006 PMID: 16723023 PMCID: PMC1557518 DOI: 10.1186/1748-717X-1-16
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Patient characteristics
| #1 | lung cancer cT2cN0 right lower lobe | primary tumor (boost) | fractionated stereotactic radiotherapy | vacuum pillow |
| #2 | oropharyngeal cancer pT2cN0 | primary and locoregional lymph nodes | fractionated IMRT | vacuum pillow and head mask |
| #3 | prostate cancer T3c Gleason score 6 PSA 5.6 | prostate and seminal vesicles | fractionated IMRT | stereotactic body cast and head mask |
| #4 | prostate cancer T2c Gleason score 7 PSA 12.0 | prostate and seminal vesicles | fractionated IMRT | stereotactic body cast and head mask |
| #5 | unresectable chordoma | lumbosacral spine | fractionated IMRT | stereotactic body cast and head mask |
| #6 | recurrence of soft tissue sarcoma | lumbal spine and right m. psoas | fractionated IMRT | stereotactic body cast and head mask |
Figure 1Linear accelerator equipped with an x-Ray tube mounted at the opposite side of the MV-beam source. The flat panel detector is attached right below the multi-leaf collimator. Single kV-images or cone beam CT sequences of patient in treatment position can be acquired for image guide radiotherapy.
Figure 2Schematic description of the workflow applied for automatic patient positioning.
Figure 3Clinical examples of cone beam (right side) compared to diagnostic treatment planning CT (left side).
Setup deviations evaluated with CBCT
| #1 | 3.1 mm | 0.1 mm | 6.0 mm | 0° | yes | good |
| #2 | -0.6 mm | 0.5 mm | -0.7 mm | 0.6° | no | good |
| #3 | -0.7 mm | -1.2 mm | 2.3 mm | 0.7° | yes | poor |
| #4 | 1.2 mm | 3.6 mm | 0.1 mm | 1.1° | yes | sufficient |
| #5 | 0.3 mm | 0.1 mm | 0.1 mm | 0° | no | sufficient |
| #6 | -2.6 mm | -1.7 mm | -1.7 mm | 1.5° | yes | sufficient |
Mean time intervals needed for cone-beam CT setup evaluation
| #1 | 02:10 | 03:20 | 04:20 | 01:40 | 11:30 |
| #2 | 02:02 | 03:12 | 04:10 | 02:20 | 11:44 |
| #3 | 02:20 | 02:46 | 04:00 | 01:10 | 10:16 |
| #4 | 02:00 | 02:40 | 02:30 | 00:20 | 07:30 |
| #5 | 02:00 | 02:37 | 04:10 | 01:50 | 10:37 |
| #6 | 02:00 | 03:15 | 03:10 | 01:40 | 10:05 |