| Literature DB >> 26766975 |
Changsheng Ma1, Jianping Cao2, Yong Yin1, Jian Zhu1.
Abstract
On-board kilovoltage (KV) cone-beam computed tomography (CBCT) images are used predominantly for the setup of patients' positioning. The image data can also potentially be used for dose calculation with the precise calibration of Hounsfield units (HU) to electron density (HU-density). CBCT calibration was analyzed in this study. A clinical treatment planning system was employed for CT and KV CBCT image to dose calculations and subsequent comparisons. Two HU-density tables were generated using the Computerized Imaging Reference Systems (CIRS) phantom. The results showed that a maximum ∼4% dose discrepancy was observed for inserts. The single field isodose curves were very close. The lung clinical patient study indicated that the volume of lung tumor that achieved the prescribed dose in CBCT was lower than in the CT plan. Our study showed that the dosimetric accuracy of CBCT-based dose calculation for lung tumor is acceptable only for the purpose of dosimetric checks with calibration applied. KV CBCT images cannot replace traditional CT images for dose calculation accuracy.Entities:
Keywords: Cone beam CT; Hounsfield units; lung tumor; radiotherapy
Year: 2014 PMID: 26766975 PMCID: PMC4704274 DOI: 10.1111/1759-7714.12055
Source DB: PubMed Journal: Thorac Cancer ISSN: 1759-7706 Impact factor: 3.500