Linac parameters such as the multi-leaf collimator (MLC) position and jaw position, cumulative monitor units (MUs), and the corresponding gantry angle were recorded during the clinical delivery of volumetric modulated arc therapy for prostate, lung, and head/neck cancer patients. Then, linac parameters were converted into the beam-data format used in the treatment planning system, and the dose distribution was reconstructed. The dose-volume histogram and the dose difference (DD) were compared with the corresponding values in the treatment plan. A reproducible error of in-treatment linac parameters was observed when a sudden change of beam intensity or MLC/jaw speed occurred. The maximum cumulative MU error was more than 4 MU for lung cancer cases, and the maximum MLC position exceeded 5 mm for prostate and head/neck cancer patients. However, these errors were quickly compensated for at the next control point. All treatments analyzed in the present study were delivered within 0.4% accuracy at the planning target volume. The cumulative dose agreed with that of the plan within 3% of the prescribed dose. The 1% DD was 93.9, 99.9, and 93.4% of the prescription dose for prostate, lung, and head/neck cancer patients, respectively.

Entities:  

Mesh:

Year:  2013        PMID: 23479401     DOI: 10.1007/s12194-013-0205-6

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  22 in total

1.  Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files.

Authors:  Jianguo Qian; Louis Lee; Wu Liu; Karen Chu; Edward Mok; Gary Luxton; Quynh-Thu Le; Lei Xing
Journal:  Phys Med Biol       Date:  2010-06-04       Impact factor: 3.609

2.  Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation.

Authors:  Yong Yang; Eduard Schreibmann; Tianfang Li; Chuang Wang; Lei Xing
Journal:  Phys Med Biol       Date:  2007-01-12       Impact factor: 3.609

3.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

4.  Commissioning of volumetric modulated arc therapy (VMAT).

Authors:  James L Bedford; Alan P Warrington
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-01       Impact factor: 7.038

5.  Quality assurance of volumetric modulated arc therapy using Elekta Synergy.

Authors:  Akihiro Haga; Keiichi Nakagawa; Kenshiro Shiraishi; Saori Itoh; Atsuro Terahara; Hideomi Yamashita; Kuni Ohtomo; Shigeki Saegusa; Toshikazu Imae; Kiyoshi Yoda; Roberto Pellegrini
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

6.  First in-situ dose calculation report using in-treatment kilovoltage cone-beam ct and in-treatment linac parameters during volumetric modulated arc therapy.

Authors:  Akira Sakumi; Akihiko Haga; Satoshi Kida; Naoya Saotome; Yukari Okano; Kenshiro Shiraishi; Takeshi Onoe; Kiyoshi Yoda; Kuni Ohtomo; Keiichi Nakagawa
Journal:  J Radiat Res       Date:  2011       Impact factor: 2.724

7.  Use of kilovoltage X-ray volume imaging in patient dose calculation for head-and-neck and partial brain radiation therapy.

Authors:  Weigang Hu; Jinsong Ye; Jiazhou Wang; Xuejun Ma; Zhen Zhang
Journal:  Radiat Oncol       Date:  2010-04-19       Impact factor: 3.481

8.  Patient-specific quality assurance method for VMAT treatment delivery.

Authors:  Eduard Schreibmann; Anees Dhabaan; Eric Elder; Tim Fox
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

9.  A study on adaptive IMRT treatment planning using kV cone-beam CT.

Authors:  George X Ding; Dennis M Duggan; Charles W Coffey; Matthew Deeley; Dennis E Hallahan; Anthony Cmelak; Arnold Malcolm
Journal:  Radiother Oncol       Date:  2007-08-20       Impact factor: 6.280

10.  Treatment of lung cancer using volumetric modulated arc therapy and image guidance: a case study.

Authors:  James L Bedford; Vibeke Nordmark Hansen; Helen A McNair; Alexandra H Aitken; Juliet E C Brock; Alan P Warrington; Michael Brada
Journal:  Acta Oncol       Date:  2008       Impact factor: 4.089

View more
  2 in total

1.  A model for preemptive maintenance of medical linear accelerators-predictive maintenance.

Authors:  Charles M Able; Alan H Baydush; Callistus Nguyen; Jacob Gersh; Alois Ndlovu; Igor Rebo; Jeremy Booth; Mario Perez; Benjamin Sintay; Michael T Munley
Journal:  Radiat Oncol       Date:  2016-03-10       Impact factor: 3.481

2.  Comparison of DVH-based plan verification methods for VMAT: ArcCHECK-3DVH system and dynalog-based dose reconstruction.

Authors:  Masahide Saito; Noriyuki Kadoya; Kiyokazu Sato; Kengo Ito; Suguru Dobashi; Ken Takeda; Hiroshi Onishi; Keiichi Jingu
Journal:  J Appl Clin Med Phys       Date:  2017-06-26       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.