Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

J G H Sutherland1, D W O Rogers.   

Abstract

PURPOSE: The absorbed-dose energy dependence of GAFCHROMIC EBT and EBT2 film irradiated in photon beams is studied to understand the shape of the curves and the physics behind them.
METHODS: The absorbed-dose energy dependence is calculated using the EGSnrc-based EGS_chamber and DOSRZnrc codes by calculating the ratio of dose to water to dose to active film layers at photon energies ranging from 3 keV to 18 MeV. These data are compared to the mass energy absorption coefficient ratios and the restricted stopping power ratios of water to active film materials as well as to previous experimental results.
RESULTS: In the photon energy range of 100 keV to 18 MeV the absorbed-dose energy dependence is found to be energy independent within +/- 0.6%. However, below 100 keV, the absorbed-dose energy dependence of EBT varies by approximately 10% due to changes in mass energy absorption coefficient ratios of water to film materials, as well as an increase in the number of electrons being created and scattered in the central surface layer of the film. Results are found to disagree with previous experimental studies suggesting the possibility of an intrinsic energy dependence at lower photon energies. For EBT2 film the absorbed-dose energy dependence at low photon energies varies by 50% or 10% depending on the manufacturing lot due to changes in the ratio of mass energy absorption coefficients of the active emulsion layers to water.
CONCLUSIONS: Caution is recommended when using GAFCHROMIC EBT/EBT2 films at photon energies below 100 keV. It is recommended that the effective atomic number of future films be produced as close to that of water and that thicker active layers are advantageous.

Entities:  

Mesh:

Year:  2010        PMID: 20384246      PMCID: PMC2837726          DOI: 10.1118/1.3301574

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code.

Authors:  Daryoush Sheikh-Bagheri; D W O Rogers
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

2.  Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.

Authors:  Sou-Tung Chiu-Tsao; Yunsil Ho; Ravi Shankar; Lin Wang; Louis B Harrison
Journal:  Med Phys       Date:  2005-11       Impact factor: 4.071

3.  Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range.

Authors:  Martin J Butson; Tsang Cheung; Peter K N Yu
Journal:  Appl Radiat Isot       Date:  2005-08-18       Impact factor: 1.513

4.  Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter.

Authors:  Alexandra Rink; I Alex Vitkin; David A Jaffray
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

5.  MicroRT-small animal conformal irradiator.

Authors:  S Stojadinovic; D A Low; A J Hope; M Vicic; J O Deasy; J Cui; D Khullar; P J Parikh; K T Malinowski; E W Izaguirre; S Mutic; P W Grigsby
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

6.  Energy dependence of EBT-1 radiochromic film response for photon (10 kvp-15 MVp) and electron beams (6-18 MeV) readout by a flatbed scanner.

Authors:  Christian Richter; Jörg Pawelke; Leonhard Karsch; Julia Woithe
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

7.  Efficiency improvements for ion chamber calculations in high energy photon beams.

Authors:  J Wulff; K Zink; I Kawrakow
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

8.  Acceptance testing computerized radiation therapy treatment planning systems: direct utilization of CT scan data.

Authors:  E C McCullough; T W Holmes
Journal:  Med Phys       Date:  1985 Mar-Apr       Impact factor: 4.071

9.  Commissioning of a novel microCT/RT system for small animal conformal radiotherapy.

Authors:  Manuel Rodriguez; Hu Zhou; Paul Keall; Edward Graves
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

10.  The use of new GAFCHROMIC EBT film for 125I seed dosimetry in Solid Water phantom.

Authors:  Sou-Tung Chiu-Tsao; David Medich; John Munro
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

  10 in total
  26 in total

1.  Evaluation of the incident directional dependence of radiochromic film by use of Monte Carlo simulation and measurement.

Authors:  Yuki Yuasa; Shinji Kawamura; Takehiro Shiinoki; Takuya Uehara; Masahiro Koike; Ryuji Kanzaki; Hideki Hanazawa; Shotaro Takahashi; Keiko Shibuya
Journal:  Radiol Phys Technol       Date:  2016-03-24

2.  Phantom scatter corrections of radiochromic films in high-energy brachytherapy dosimetry: a Monte Carlo study.

Authors:  Mishra Subhalaxmi; T Palani Selvam
Journal:  Radiol Phys Technol       Date:  2015-03-10

3.  Protocol for the measurement of the absorbed dose rate to water for a planar 32P beta emitting brachytherapy source: A multi-institutional validation.

Authors:  Liana Mulet; Izabella Barreto; Gil'ad N Cohen; Antonio L Damato; Thomas Mauceri; Jennifer Pursley; Christopher L Deufel
Journal:  Brachytherapy       Date:  2021-10-24       Impact factor: 2.362

4.  Monte Carlo model of the scanning beam digital x-ray (SBDX) source.

Authors:  M Bazalova; M D Weil; B Wilfley; E E Graves
Journal:  Phys Med Biol       Date:  2012-10-24       Impact factor: 3.609

5.  Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Authors:  Khaled Soliman; Marouf Adili; Abdullah Alrushoud
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-07

6.  Linearization of dose-response curve of the radiochromic film dosimetry system.

Authors:  Slobodan Devic; Nada Tomic; Saad Aldelaijan; Francois Deblois; Jan Seuntjens; Maria F Chan; Dave Lewis
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.506

7.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

8.  Is It Possible to Publish a Calibration Function for Radiochromic Film?

Authors:  Maria F Chan; David Lewis; Xiang Yu
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2014-01-29

9.  Monte Carlo investigation of energy response of various detector materials in ¹²⁵I and ¹⁶⁹Yb brachytherapy dosimetry.

Authors:  T Palani Selvam; Biju Keshavkumar
Journal:  J Appl Clin Med Phys       Date:  2010-07-28       Impact factor: 2.102

10.  Monte Carlo calculation of beam quality correction for solid-state detectors and phantom scatter correction at 137Cs energy.

Authors:  T Palani Selvam; Subhalaxmi Mishra; R S Vishwakarma
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.