Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.

Sou-Tung Chiu-Tsao1, Yunsil Ho, Ravi Shankar, Lin Wang, Louis B Harrison.   

Abstract

The purpose of this paper is to evaluate the energy dependence of the response of two new high sensitivity models of radiochromic films EBT and XR-QA. We determined the dose response curves of these films for four different radiation sources, namely, 6 MV photon beams (6 MVX), Ir-192, I-125, and Pd-103. The first type (EBT) is designed for intensity modulated radiation therapy (IMRT) dosimetry, and the second type (XR-QA) is designed for kilovoltage dosimetry. All films were scanned using red (665 nm) and green (520 nm) light sources in a charge-coupled device-based densitometer. The dose response curves [net optical density (NOD) versus dose] were plotted and compared for different radiation energies and light sources. Contrary to the early GAFCHROMIC film types (such as models XR, HS, MD55-2, and HD810), the net optical densities of both EBT and XR-QA were higher with a green (520 nm) than those with a red (665 nm) light source due to the different absorption spectrum of the new radiochromic emulsion. Both film types yield measurable optical densities for doses below 2 Gy. EBT film response is nearly independent of radiation energy, within the uncertainty of measurement. The NOD values of EBT film at 1 and 2 Gy are 0.13 and 0.25 for green, and 0.1 and 0.17 for red, respectively. In contrast, the XR-QA film sensitivity varies with radiation energy. The doses required to produce NOD of 0.5 are 6.9, 5.4, 0.7, and 0.9 Gy with green light and 19, 13, 1.7, and 1.5 Gy with red light, for 6 MVX, Ir-192, I -125, and Pd-103, respectively. EBT film was found to have minimal photon energy dependence of response for the energies tested and is suitable for dosimetry of radiation with a wide energy spectrum, including primary and scattered radiation. XR-QA film is promising for kilovoltage sources with a narrow energy spectra. The new high sensitivity radiochromic films are promising tools in radiation dosimetry.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16370422     DOI: 10.1118/1.2065467

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  39 in total

1.  Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

Authors:  J G H Sutherland; D W O Rogers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector.

Authors:  Frédéric Lacroix; Mathieu Guillot; Malcolm McEwen; Claudiu Cojocaru; Luc Gingras; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  Technical note: Heterogeneity dose calculation accuracy in IMRT: study of five commercial treatment planning systems using an anthropomorphic thorax phantom.

Authors:  Scott E Davidson; Richard A Popple; Geoffrey S Ibbott; David S Followill
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

4.  Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium.

Authors:  Alexandra Rink; David F Lewis; Sangya Varma; I Alex Vitkin; David A Jaffray
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

5.  Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

Authors:  Somayeh Gholami; Hamid Reza Mirzaei; Ali Jabbary Arfaee; Ramin Jaberi; Hassan Ali Nedaie; Seied Rabi Mahdavi; Eftekhar Rajab Bolookat; Ali S Meigooni
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-18

6.  Tumour bed irradiation of human tumour xenografts in a nude rat model using a common X-ray tube.

Authors:  S V Tokalov; W Enghardt; N Abolmaali
Journal:  J Biosci       Date:  2010-06       Impact factor: 1.826

7.  Characterization of a 137Cs irradiator from a new perspective with modern dosimetric tools.

Authors:  Samuel L Brady; Greta Toncheva; Mark W Dewhirst; Terry T Yoshizumi
Journal:  Health Phys       Date:  2009-09       Impact factor: 1.316

8.  Monte Carlo simulations and radiation dosimetry measurements of peripherally applied HDR 192Ir breast brachytherapy D-shaped applicators.

Authors:  Yun Yang; Mark J Rivard
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

9.  Commissioning of a novel microCT/RT system for small animal conformal radiotherapy.

Authors:  Manuel Rodriguez; Hu Zhou; Paul Keall; Edward Graves
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

10.  High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities.

Authors:  John Wong; Elwood Armour; Peter Kazanzides; Iulian Iordachita; Erik Tryggestad; Hua Deng; Mohammad Matinfar; Christopher Kennedy; Zejian Liu; Timothy Chan; Owen Gray; Frank Verhaegen; Todd McNutt; Eric Ford; Theodore L DeWeese
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-08-01       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.