Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Khaled Soliman1, Marouf Adili1, Abdullah Alrushoud1.   

Abstract

AIM: Blood irradiators (BI) initial acceptance testing and routine annual dosimetry checks require radiation dose measurements in order to comply with regulatory requirements.
BACKGROUND: Traditionally thermo-luminescence dosimeters (TLD) have been used to measure the dose. The EBT3 film is reported to be a better dosimeter for low energy X-rays than its predecessors EBT2 and EBT. To the best of our knowledge, the use of EBT3 films to perform dosimetry on X-ray based BI has not been reported yet.
MATERIALS AND METHODS: We performed routine radiation dosimetry checks using EBT3 films on a new X-ray based BI and compared the results with TLD dosimetry. Calibration films were irradiated with radiation beam from a Co-60 Gamma Knife (GK) radiosurgery machine and, alternatively, using an Ir-192 high dose rate (HDR) brachytherapy device. The films were calibrated to cover a wide dose range from 1 to 40 Gy. Such a wide dose range has not been reported yet in BI film dosimetry.
RESULTS: We obtained a relative difference of about 6.6% between doses measured using TLD and those measured using EBT3 films. Both irradiation methods using GK or HDR were found to be adequate for the calibration of the EBT3 Gafchromic films.
CONCLUSIONS: We recommend the use of EBT3 films in routine X-ray based BI dosimetry checks. The presented method takes advantage of available radiotherapy equipment that can be efficiently used for EBT3 films calibration. The method is fast, reproducible and saves valuable medical physicist's time.

Entities:  

Keywords:  Blood irradiation; Gafchromic EBT3; Gamma Knife; Quality assurance; X-ray based irradiators

Year:  2019        PMID: 31198412      PMCID: PMC6556754          DOI: 10.1016/j.rpor.2019.05.006

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  23 in total

1.  Dosimetry of gamma chamber blood irradiator using Gafchromic EBT film.

Authors:  S Vandana; V S Shaiju; S D Sharma; S Mhatre; S Shinde; G Chourasiya; Y S Mayya
Journal:  Appl Radiat Isot       Date:  2010-09-17       Impact factor: 1.513

2.  Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

Authors:  J G H Sutherland; D W O Rogers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

3.  Post-irradiation colouration of Gafchromic EBT radiochromic film.

Authors:  Tsang Cheung; Martin J Butson; Peter K N Yu
Journal:  Phys Med Biol       Date:  2005-10-04       Impact factor: 3.609

4.  Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range.

Authors:  Martin J Butson; Tsang Cheung; Peter K N Yu
Journal:  Appl Radiat Isot       Date:  2005-08-18       Impact factor: 1.513

5.  Dosimetric characterization of an image-guided stereotactic small animal irradiator.

Authors:  R Pidikiti; S Stojadinovic; M Speiser; K H Song; F Hager; D Saha; T D Solberg
Journal:  Phys Med Biol       Date:  2011-03-28       Impact factor: 3.609

6.  Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams.

Authors:  Thomas A D Brown; Kenneth R Hogstrom; Diane Alvarez; Kenneth L Matthews; Kyungmin Ham; Joseph P Dugas
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

7.  Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams.

Authors:  J Sorriaux; A Kacperek; S Rossomme; J A Lee; D Bertrand; S Vynckier; E Sterpin
Journal:  Phys Med       Date:  2012-10-26       Impact factor: 2.685

8.  Dosimetry of blood irradiation with radiochromic film.

Authors:  M J Butson; P K Yu; T Cheung; M G Carolan; K Y Quach; A Arnold; P E Metcalfe
Journal:  Transfus Med       Date:  1999-09       Impact factor: 2.019

9.  Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams.

Authors:  S Reinhardt; M Hillbrand; J J Wilkens; W Assmann
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

10.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

View more
  2 in total

1.  Diacetylene-Based Colorimetric Radiation Sensors for the Detection and Measurement of γ Radiation during Blood Irradiation.

Authors:  Apoorva Mittal; Shalini Verma; Gopishankar Natanasabapathi; Pratik Kumar; Akhilesh K Verma
Journal:  ACS Omega       Date:  2021-03-29

2.  Development of custom lead shield and strainer for targeted irradiation for mice in the gamma cell chamber.

Authors:  Nurhaslina Hasan; Nur Fatihah Ronny Sham; Muhammad Khalis Abdul Karim; Syed Baharom Syed Ahmad Fuad; Narimah Abdul Hamid Hasani; Effat Omar; Mohammad Johari Ibahim
Journal:  Sci Rep       Date:  2021-07-15       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.