| Literature DB >> 4000085 |
Abstract
The availability of computerized radiation therapy treatment planning systems that utilize computed tomography (CT) scan data requires testing additional to that routinely needed for non-CT systems. These additional items include dimensioning verification, establishing CT number-to-tissue property conversions, verifying the accuracy of heterogeneity corrected dose predictions and autocontouring. One testing protocol is presented and sample results from an Atomic Energy of Canada Theraplan L system are presented.Mesh:
Year: 1985 PMID: 4000085 DOI: 10.1118/1.595713
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071