Energy dependence of EBT-1 radiochromic film response for photon (10 kvp-15 MVp) and electron beams (6-18 MeV) readout by a flatbed scanner.

Christian Richter1, Jörg Pawelke, Leonhard Karsch, Julia Woithe.   

Abstract

PURPOSE: The aim of this article is to investigate the energy dependence of the radiochromic film type, Gafchromic EBT-1, when scanned with a flatbed scanner for film readout.
METHODS: Dose response curves were determined for 12 different beam qualities ranging from a 10 kVp x-ray beam to a 15 MVp x-ray beam and include also two high energy electron beam qualities (6 and 18 MeV). The dose responses measured as net optical density (netOD) for the different beam qualities were normalized to the response of a reference beam quality (6 MVp).
RESULTS: A strong systematic energy dependence of the film response was found. The lower the effective beam energy, the less sensitive the EBT-1 films get. The maximum decrease in dose for the same film response between the 25 kVp and 6 MVp beam qualities was 44%. Additionally, a difference in energy dependence for different doses was discovered, meaning that higher doses show a smaller dependency on energy than lower doses. The maximum decrease in the normalized netOD was found to be 25% for a dose of 0.5 Gy relative to the normalized netOD for 10 Gy. Moreover, a scaling procedure is introduced, allowing the correction of the energy dependence for the investigated beam qualities and also for comparable x-ray beam qualities within the energy range studied.
CONCLUSIONS: A strong energy dependence for EBT-1 radiochromic films was found. The films were readout with a flatbed scanner. If the effective beam energy is known, the energy dependence can be corrected with the introduced scaling procedure. Further investigation of the influence of the spectral band of the readout device on energy dependence is needed to understand the reason for the different energy dependences found in this and previous works.

Entities:  

Mesh:

Year:  2009        PMID: 20095263     DOI: 10.1118/1.3253902

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

Authors:  J G H Sutherland; D W O Rogers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector.

Authors:  Frédéric Lacroix; Mathieu Guillot; Malcolm McEwen; Claudiu Cojocaru; Luc Gingras; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

4.  Percutaneous transthoracic needle biopsy of small (≤ 1 cm) lung nodules under C-arm cone-beam CT virtual navigation guidance.

Authors:  Ji Yung Choo; Chang Min Park; Nyoung Keun Lee; Sang Min Lee; Hyun-Ju Lee; Jin Mo Goo
Journal:  Eur Radiol       Date:  2012-09-14       Impact factor: 5.315

5.  Radiobiological influence of megavoltage electron pulses of ultra-high pulse dose rate on normal tissue cells.

Authors:  Lydia Laschinsky; Leonhard Karsch; Elisabeth Leßmann; Melanie Oppelt; Jörg Pawelke; Christian Richter; Michael Schürer; Elke Beyreuther
Journal:  Radiat Environ Biophys       Date:  2016-05-19       Impact factor: 1.925

6.  Comparison study of in vivo dose response to laser-driven versus conventional electron beam.

Authors:  Melanie Oppelt; Michael Baumann; Ralf Bergmann; Elke Beyreuther; Kerstin Brüchner; Josefin Hartmann; Leonhard Karsch; Mechthild Krause; Lydia Laschinsky; Elisabeth Leßmann; Maria Nicolai; Maria Reuter; Christian Richter; Alexander Sävert; Michael Schnell; Michael Schürer; Julia Woithe; Malte Kaluza; Jörg Pawelke
Journal:  Radiat Environ Biophys       Date:  2015-01-20       Impact factor: 1.925

7.  Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA.

Authors:  Grzegorz Zwierzchowski; Grzegorz Bielęda; Janusz Skowronek; Magdalena Mazur
Journal:  J Contemp Brachytherapy       Date:  2016-08-16

8.  Dosimetric evaluation of internal shielding in a high dose rate skin applicator.

Authors:  Françoise Lliso; Domingo Granero; Jose Perez-Calatayud; Vicente Carmona; M Carmen Pujades; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2011-03-31

9.  Establishment of a small animal tumour model for in vivo studies with low energy laser accelerated particles.

Authors:  Kerstin Brüchner; Elke Beyreuther; Michael Baumann; Mechthild Krause; Melanie Oppelt; Jörg Pawelke
Journal:  Radiat Oncol       Date:  2014-02-18       Impact factor: 3.481

10.  Testing the methodology for dosimetry audit of heterogeneity corrections and small MLC-shaped fields: Results of IAEA multi-center studies.

Authors:  Joanna Izewska; Paulina Wesolowska; Godfrey Azangwe; David S Followill; David I Thwaites; Mehenna Arib; Amalia Stefanic; Claudio Viegas; Luo Suming; Daniela Ekendahl; Wojciech Bulski; Dietmar Georg
Journal:  Acta Oncol       Date:  2016-03-03       Impact factor: 4.089

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.