Monte Carlo calculation of beam quality correction for solid-state detectors and phantom scatter correction at 137Cs energy.

T Palani Selvam1, Subhalaxmi Mishra, R S Vishwakarma.   

Abstract

Beam quality correction kQQo (r), which reflects the absorbed energy dependence of the detector, is calculated for solid state detector materials diamond, LiF, Li2B4O7 and Al2O3 for the 137Cs RTR brachytherapy source using the Monte Carlo-based EGSnrc code system. The study also includes calculation of detector-specific phantom scatter corrections kphant (r) for solid phantoms such as PMMA, polystyrene, RW1, solid water, virtual water and plastic water. Above corrections are calculated as a function of distance r along the transverse axis of the source. kQQo (r) is about unity for the Li2B4O7 detector. LiF detector shows a gradual decrease in kQQo (r) with r (decrease is about 2% over the distance range of 1 - 15 cm). Diamond detector shows a gradual increase in kQQo (r)with r (about 3% larger than unity at 15 cm). In the case of Al2O3 detector, kQQo (r)decreases with r steeply (about 14% over the distance range of 1 - 15 cm). The study shows that some solid state detectors demonstrate distance-dependent kphant (r)values, but the degree deviation from unity depends on the type of solid phantom and the detector.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24423840      PMCID: PMC5711252          DOI: 10.1120/jacmp.v15i1.4445

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

American Association of Physicists in Medicine (AAPM) Task Group reports AAPM TG43 and TG43U 1 recommend water as a reference medium for dosimetry of interstitial brachytherapy sources. Due to high‐dose gradients near brachytherapy sources and specification of the dose parameters within few centimeters of the source, source‐detector distance should be specified very accurately for dosimetric measurements. Precise positioning of detectors, reproducibility of source and detectors in reference liquid water medium, and water proofing of detectors posses a practical problem. Solid phantom materials can be easily machined to accommodate the source and detectors in a precise geometrical configuration, facilitating an accurate measurement and reproducibility in source‐detector geometry. In a previously published article, relative absorbed‐dose energy response corrections R for detector materials such as air, LiF, , Si diode, diamond, and were presented for and brachytherapy sources. The corrections were calculated using the EGSnrc‐based Monte Carlo code system for liquid water, PMMA, and polystyrene phantom materials. The present study is aimed at investigating absorbed‐dose energy dependence of solid‐state detector materials such as diamond, LiF, , and at the energy. This investigation also includes calculation of detector‐specific phantom scatter correction for different solid phantoms such as PMMA, polystyrene, RW1, solid water, virtual water, and plastic water. The EGSnrc‐based user‐codes DOSRZnrc and FLURZnrc are used in the study.

MATERIALS AND METHODS

RTR source

The geometric details and material data of the RTR are from the published work. The active length and active radius (active material is gold) of the source are 1.5 cm and 0.04 cm, respectively. The outer radius of the source is 1.5 mm. For Monte Carlo calculations, we have considered only the 662 keV gamma energy of emission, as in a previously published study by Selvam et al., it was demonstrated that X‐rays were not important.

Phantom materials

Elemental composition, mass fraction, mass density, , and effective atomic number () of water and solid phantom materials are presented in Table 1. The atomic composition and density details of the phantoms are taken from the literature. , , , values are calculated at 662 keV using the software by Taylor et al. Elemental composition, mass fraction, mass density, , and effective atomic number () of water and solid phantom materials a Hubble & Seltzer (Ref. 11); b ICRU‐44 ( Ref. 8); c Meigooni et al. (Ref. 9); d Reniers et al. (Ref. 10); e Taylor et al. (Ref. 12).

Theoretical background of measurement of absorbed dose to water at brachytherapy energies

Dose measurements in water phantom

Following discussion is based on the published study by Adolfsson et al. Primary standards of absolute measurements of absorbed dose to water are based on water calorimetry. or megavoltage (MV) photon beam serves as a reference beam quality for this purpose. A dosimeter, for example, ionization chamber calibrated to measure at the primary or secondary standards can be used in other beam quality Q (example, other clinical MV photon beams) by using the beam quality correction factor . , The other dosimeters, such as solid‐state dosimeters, can therefore be calibrated to measure at Q traceable to the primary standard. Note that may be calculated at a brachytherapy beam quality, Q, involving a solid‐state detector. Consider a solid‐state detector is used for measuring at . This quantity is denoted by The output measured by the solid‐state detector is denoted by . An absorbed dose‐to‐water calibration coefficient can be obtained by using the following the relation: The absorbed dose to the material of the sensitive detector element at , , and are related as follows: , , where the function (called intrinsic energy‐dependence , ) relates and as below: Let us now consider a cylindrical photon emitting brachytherapy source (beam quality Q, in this study it is ) is immersed in a liquid water phantom. The absorbed dose to water in the liquid water phantom at r along the transverse axis of the source is denoted by . The output measured by the detector at r is . Likewise Eq. (3), absorbed dose to the detector at Q, and are related by is obtained by using the following relation: where is the beam quality correction and is given by Using Eq. (1) in Eq. (6) gives Using Eqs. (3) and (4) in Eq. (7) gives is relative absorbed dose energy response correction. , , As described in the previously published work, , , , absorbed‐dose dependence at relates absorbed dose to medium of interest (usually water), and absorbed dose to detector, , as below: Similarly at : Equation (12) is therefore written as: Equation (10) has two components: (a) f , relative intrinsic energy dependence of the detector which can only be determined experimentally, and (b) , inverse of relative absorbed‐dose energy response correction. Investigations on photon energy dependence of LiF:Mg, Ti TLDs were published in the 1960s and 1970s, with a summary of the results presented by Budd et al. Most of the studies measured an intrinsic energy dependence that was greater than unity for photon energies below about 150 keV, relative to TLDs that had been calibrated using photons. On average, the measured light output was about 10% higher than would be expected based solely on the absorbed‐dose energy dependence. For detailed discussion on intrinsic energy dependence of TLDs, readers may consult the literature. As mentioned by Adolfsson et al., when an ion chamber is used, where W is the mean energy imparted to air to form an ion pair in air at Q, and is the corresponding quantity at . The value of W is usually considered to be independent of the beam quality in MV photon and electron beams, but may take other values in beams of protons and heavier charged particles due to the increased ion density along the tracks of the heavy charged particles compared to that along electron tracks. Note that if the yield of radiation‐induced products in the detector is independent of the radiation beam quality (i.e., yield is constant), then Therefore Eq. (9) becomes

Brachytherapy dose measurements in a solid phantom

Generally, in brachytherapy, absorbed dose measurements involving solid‐state detectors are carried out in solid phantoms. The absorbed dose to detector at r in the solid phantom at Q is denoted by . It is recalled that is absorbed dose to detector at Q at r in the liquid water phantom. and are related as follows: where accounts for influence of solid phantom on the response of the detector, which is known as phantom scatter correction at beam quality Q. Therefore, when measurements are carried out in solid phantoms at Q, in addition to the application of (Eq. (16)), the detector response is required to be corrected for to account of phantom scatter. The final expression for obtaining absorbed dose to water in the liquid water phantom is given by where is output measured by the solid detector at Q in a solid phantom at r.

Monte Carlo calculations

FLURZnrc simulations of collision kerma and mean energies for RTR source

The approach adapted for the Monte Carlo calculations of dose ratio of detector to water is as described in the published study. The source is positioned at the centre of a 40 cm diameter by 40 cm height cylindrical phantoms (liquid water and solid phantoms). The photon fluence spectrum in 10 keV energy intervals is scored along the transverse axis of the source () in 2 mm high and 0.5 mm thick cylindrical shells. The fluence spectrum is converted to collision kerma to water and collision kerma to detector materials by using the mass energy‐absorption coefficients of water and detector materials, respectively.

Calculations of dose ratios at

In the published study, it was demonstrated that the detector‐to‐water dose ratio calculated at the reference beam quality ( beam) at 0.5 mm depth in water phantom was independent of the detector thickness (). In the present study, we calculated the above dose ratio for depths 5 cm and 10 cm along the central axis of the water phantom. We used detector dimensions of . In the Monte Carlo calculations, a parallel beam is incident on a cylindrical water phantom. The beam has a radius of 5.64 cm at the front face of the phantom (field size is 100 cm2). A realistic spectrum from a telecobalt unit distributed along with the EGSnrc code system is used in the calculations. This investigation produced similar dose ratios as obtained at 5 mm depth. This suggests that is independent of depth in the water phantom. We also calculated the dose ratio at depths 5 mm, 5 cm, and 10 cm in the PMMA phantom using the detector dimensions . The results obtained from the PMMA phantom compare well with the results of water phantom. We have therefore used the values of published in the previous work for deriving .

Monte Carlo parameters

Up to 2.5 × 109 photon histories are simulated. The 1 g statistical uncertainties on the calculated absorbed dose and collision kerma values are about 0.2%. The statistical uncertainties on the calculated values of and are less than 0.5%. The values of Monte Carlo parameters AE, AP, ECUT, PCUT, and ESAVE used in the FLURRZnrc calculations are 0.521, 0.01, 0.01, 2, and 2 MeV, respectively. In the case of DOSRZnrc calculations, the value of ECUT used is 0.521 MeV (10 keV kinetic energy of electrons) and the values of other parameters are as that used in the FLURRZnrc simulations. The parameters AE and AP are the low‐energy thresholds for the production of knock‐on electrons and secondary bremsstrahlung photons, respectively. The parameters ECUT and PCUT electron and photon transport cutoff, respectively. ESAVE is a parameter related to range rejection technique.

RESULTS & DISCUSSION

Fluence‐weighted mean energy,

Table 2 presents the values of as a function of r for the RTR source in various phantoms. As r increases decreases, but the degree of decrease depends on the type of phantom. For the phantoms such as water, virtual water, RW1, and solid water, decreases from about 565 keV to 260 keV when the distance is increased from 1 cm to 15 cm. In the case of plastic water phantom, decreases from 570 keV to 285 keV in the above distance range. The values of at 15 cm are 228 keV and 239 keV, respectively, for PMMA and polystyrene phantoms. Monte Carlo‐calculated values of fluence‐weighted mean energy for different phantoms. The values are presented as a function of distance along the transverse axis of the RTR brachytherapy source

Phantom scatter correction

The investigation of phantom scatter also included water as a detector material. Values of calculated for the phantoms polystyrene, PMMA, virtual water, RW1, solid water, and plastic water are presented in Figs. 1 to 6. A solid phantom may be termed as water‐equivalent when the value of is unity. The investigation suggests that some solid‐state detectors demonstrate distance‐dependent values, but the degree of dependence depends on the type of solid phantom and the type of detector. For example, the phantoms such as RW1, virtual water, and solid water almost behave like water‐equivalent at all distances (1‐15 cm) for all the investigated detectors (with a maximum deviation of about 2% from unity for the detector in RW1 phantom). Polystyrene, virtual water, RW1, and solid water phantoms are water‐equivalent for the diamond detector as is about unity, independent of distance (maximum deviation is about 1% in the distance range of 1‐15 cm for polystyrene phantom). Whereas, for the phantoms PMMA and plastic water, increases with r for the diamond detector. The value increases to 1.0607 in PMMA and 1.0212 in plastic water at 15 cm for the diamond detector. For the LiF, detectors, virtual water, RW1, and solid water are water‐equivalent (within 1%). Note that detector behaves like water detector at all distances for all the solid phantom materials investigated. For the detector, the phantoms such as Polystyrene, PMMA, and RW1 show decrease in with r and the degree of decrease is higher for polystyrene phantom. For example, the value decreases to 0.9075, 0.9697, and 0.9794 at 15 cm for the phantoms polystyrene, PMMA, and RW1, respectively. The degree of decrease is higher in polystyrene phantom. Phantom scatter correction presented for polystyrene phantom as a function of distance along the transverse axis of the RTR brachytherapy source. The values are presented for detector materials LiF, , diamond, , and water. Same as Fig. 1, but for PMMA phantom. Same as Fig. 1, but for virtual water phantom. Same as Fig. 1, but for RW1 phantom. Same as Fig. 1, but for solid water phantom. Same as Fig. 1, but for plastic water phantom.

Beam quality correction

Figure 7 presents the values of for the RTR source obtained using Eq. (16). The numerical values of this figure are given in Table 3. For the detector, is about unity, and is independent of r. The LiF detector shows a gradual decrease in with r. The decrease is 2% over the distance range of 1‐15 cm. Diamond detector shows a gradual increase in with r (about 3% larger than unity at 15 cm). For the detector, decreases with r steeply (about 14% over the distance range of 1–15 cm). Monte Carlo‐calculated beam quality correction for RTR brachytherapy source. The values are presented as a function of distance along the transverse axis of the source for detector materials LiF, , diamond, and . Monte Carlo‐calculated values of beam quality correction . The data are presented as a function of distance along the transverse axis of the RTR source

Influence of detector dimensions on detector response

Dimensions of TLD‐100 (LiF:Mg,Ti) chips reported in the literature , are , and , and . Carbon‐doped cylindrical discs of detectors () are used in radiotherapy photon beams. : C chips (2 mm long and in cross‐sectional area) are used in 192Ir high‐dose‐rate dosimetry. The sensitive volume of the PTW/diamond detector is a disk made from natural diamond (density 3.51 g/cm3) with a radius ranging from 1.0 to 2.2 mm and a thickness ranging from 0. 2 to 0.4 mm. cylindrical pellets () are used in radiotherapy dose measurements. In order to quantify the influence of detector thicknesses on the calculated response, we adapted an approach as applied in a previously published work due to limitations associated with the DOSRZnrc user‐code. LiF, , and detectors are modeled as cylindrical shells of thickness 1 mm and height 2 mm along the transverse axis of the source. The phantoms considered are water, polystyrene, and plastic water. Absorbed dose and collision kerma to these detectors are calculated at and 15 cm. The DOSRZnrc‐based collision kerma values are statistically identical to the FLURZnrc‐based collision kerma values. This suggests that detector dimensions do not affect the calculated values. In the case of diamond detector, the calculations are carried out for 0.2 mm and 0.4 mm thicknesses separately (height is 2 mm). DOSRZnrc calculations using these thicknesses show collision kerma values comparable to those obtained using the FLURZnrc user‐code. Whereas the absorbed dose calculated for the 0.2 mm thick diamond detector is smaller by about 1% when compared to the collision kerma. In the case of 0.4 mm thick diamond detector, both collision kerma and absorbed dose are statistically identical.

CONCLUSIONS

Absorbed‐dose energy dependence of solid‐state detector materials such as diamond, LiF, , and for the RTR brachytherapy source is studied using the Monte Carlo‐based EGSnrc code system. Beam quality correction , which reflects absorbed‐dose energy dependence of the detector, shows a gradual decrease with r for the LiF detector (decrease is about 2% over the distance range of 1‐15 cm). Diamond detector shows a gradual increase in with r (about 3% larger than unity at 15 cm). For detector, decreases with r steeply (about 14% over the distance range of 1‐15 cm). does not show energy dependence. The study shows that some solid‐state detectors demonstrate distance‐dependent values, but the degree of dependence depends on the type of solid phantom and the detector.

ACKNOWLEDGMENTS

The authors would like to thank Dr. D. N. Sharma, Director, Health, Safety & Environment Group, Bhabha Atomic Research Centre (BARC), and Mr. D. A. R. Babu, Head, Radiological Physics & Advisory Division, BARC for their encouragement and support throughout the study.
Table 1

Elemental composition, mass fraction, mass density, , and effective atomic number () of water and solid phantom materials

Element Z A Water a Solid water b RW1 b Plastic Water c Virtual Watef d PMMA a Polystyrene a
Composition and mass fraction
H11.0080.11190.0810.1320.09300.0770.080540.07742
C612.010.6720.7940.62820.6870.599850.92258
N714.010.0240.01000.023
O815.990.88810.1990.0380.17940.1890.31961
Mg1224.310.009
Cl1735.460.0010.0270.00960.001
Ca2040.080.0230.07950.023
Br3579.900.0003
Mass density (g/cm3)1.0001.0360.9701.0131.0301.1901.060
<Z/A> 0.5550.5400.5650.5450.5380.5390.538
Zeff e 3.343.572.813.443.643.603.50

a Hubble & Seltzer (Ref. 11);

b ICRU‐44 ( Ref. 8);

c Meigooni et al. (Ref. 9);

d Reniers et al. (Ref. 10);

e Taylor et al. (Ref. 12).

Table 2

Monte Carlo‐calculated values of fluence‐weighted mean energy for different phantoms. The values are presented as a function of distance along the transverse axis of the RTR brachytherapy source

Fluence‐weighted mean energy (MeV)
Distance, r (cm) Water PMMA Polystyrene Plastic Water Virtual Water RW1 Solid Water
10.5660.5570.5630.5680.5670.5670.566
20.5160.4990.5100.5210.5170.5170.516
30.4710.4490.4620.4800.4730.4720.472
40.4330.4070.4210.4450.4360.4350.433
50.4010.3710.3870.4140.4040.4010.402
60.3740.3420.3580.3900.3750.3740.375
70.3510.3180.3340.3690.3530.3510.351
80.3320.2990.3130.3510.3340.3300.333
90.3150.2820.2950.3340.3180.3150.316
100.3020.2670.2810.3220.3040.3000.302
110.2890.2560.2670.3110.2920.2880.291
120.2790.2460.2580.3020.2820.2770.281
130.2710.2390.2490.2950.2750.2690.273
140.2650.2320.2430.2890.2690.2640.268
150.2600.2280.2390.2850.2630.2590.264
Table 3

Monte Carlo‐calculated values of beam quality correction . The data are presented as a function of distance along the transverse axis of the RTR source

kQQ0(r)
Distance, r (cm) Diamond Al2O3 Li2B4O7 LiF
11.0010.9931.0000.999
21.0020.9871.0000.998
31.0040.9781.0000.997
41.0050.9691.0000.996
51.0070.9581.0010.994
61.0100.9471.0010.993
71.0120.9361.0010.991
81.0140.9251.0020.989
91.0160.9141.0020.987
101.0180.9031.0020.986
111.0210.8941.0030.984
121.0220.8861.0030.983
131.0240.8791.0030.982
141.0250.8731.0040.981
151.0260.8691.0040.980
  19 in total

1.  AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams.

Authors:  P R Almond; P J Biggs; B M Coursey; W F Hanson; M S Huq; R Nath; D W Rogers
Journal:  Med Phys       Date:  1999-09       Impact factor: 4.071

2.  Room scatter studies in the air kerma strength standardization of the Amersham CDCS-J-type 137Cs source: a Monte Carlo study.

Authors:  T Palani Selvam; K N Govinda Rajan; P S Nagarajan; B C Bhatt; P Sethulakshmi
Journal:  Phys Med Biol       Date:  2002-05-07       Impact factor: 3.609

3.  The radial dose function of low-energy brachytherapy seeds in different solid phantoms: comparison between calculations with the EGSnrc and MCNP4C Monte Carlo codes and measurements.

Authors:  B Reniers; F Verhaegen; S Vynckier
Journal:  Phys Med Biol       Date:  2004-04-21       Impact factor: 3.609

4.  Robust calculation of effective atomic numbers: the Auto-Z(eff) software.

Authors:  M L Taylor; R L Smith; F Dossing; R D Franich
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

5.  Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

Authors:  J G H Sutherland; D W O Rogers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

6.  Calculation of stopping power ratios for carbon ion dosimetry.

Authors:  Oksana Geithner; P Andreo; N Sobolevsky; G Hartmann; O Jäkel
Journal:  Phys Med Biol       Date:  2006-04-19       Impact factor: 3.609

7.  Absorbed dose water calorimeter.

Authors:  S R Domen
Journal:  Med Phys       Date:  1980 Mar-Apr       Impact factor: 4.071

8.  Response of lithium formate EPR dosimeters at photon energies relevant to the dosimetry of brachytherapy.

Authors:  Emelie Adolfsson; Gudrun Aim Carlsson; Jan-Erik Grindborg; Hakan Gustafsson; Eva Lund; Asa Carlsson Tedgren
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

9.  A comparative study of dosimetric properties of Plastic Water and Solid Water in brachytherapy applications.

Authors:  A S Meigooni; Z Li; V Mishra; J F Williamson
Journal:  Med Phys       Date:  1994-12       Impact factor: 4.071

10.  Accuracy of Monte Carlo photon transport simulation in characterizing brachytherapy dosimeter energy-response artefacts.

Authors:  R K Das; Z Li; H Perera; J F Williamson
Journal:  Phys Med Biol       Date:  1996-06       Impact factor: 3.609

View more
  4 in total

1.  Phantom scatter corrections of radiochromic films in high-energy brachytherapy dosimetry: a Monte Carlo study.

Authors:  Mishra Subhalaxmi; T Palani Selvam
Journal:  Radiol Phys Technol       Date:  2015-03-10

2.  Monte Carlo Calculation of Beam Quality and Phantom Scatter Corrections for Lithium Formate Electron Paramagnetic Resonance Dosimeter for High-energy Brachytherapy Dosimetry.

Authors:  Subhalaxmi Mishra; T Palani Selvam
Journal:  J Med Phys       Date:  2017 Apr-Jun

3.  An energy minimization method for the correction of cupping artifacts in cone-beam CT.

Authors:  Shipeng Xie; Wenqin Zhuang; Haibo Li
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

4.  Monte Carlo-based beam quality and phantom scatter corrections for solid-state detectors in 60Co and 192Ir brachytherapy dosimetry.

Authors:  Mishra Subhalaxmi; T Palani Selvam
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102