Commissioning of a novel microCT/RT system for small animal conformal radiotherapy.

Manuel Rodriguez1, Hu Zhou, Paul Keall, Edward Graves.   

Abstract

The purpose of this work was to commission a 120 kVp photon beam produced by a micro-computed tomography (microCT) scanner for use in irradiating mice to therapeutic doses. A variable-aperture collimator has been integrated with a microCT scanner to allow the delivery of beams with pseudocircular profiles of arbitrary width between 0.1 and 6.0 cm. The dose rate at the isocenter of the system was measured using ion chamber and gafchromic EBT film as 1.56-2.13 Gy min(-1) at the water surface for field diameters between 0.2 and 6.0 cm. The dose rate decreases approximately 10% per every 5 mm depth in water for field diameters between 0.5 and 1.0 cm. The flatness, symmetry and penumbra of the beam are 3.6%, 1.0% and 0.5 mm, respectively. These parameters are sufficient to accurately conform the radiation dose delivered to target organs on mice. The irradiated field size is affected principally by the divergence of the beam. In general, the beam has appropriate dosimetric characteristics to accurately deliver the dose to organs inside the mice's bodies. Using multiple beams delivered from a variety of angular directions, targets as small as 2 mm may be irradiated while sparing surrounding tissue. This microCT/RT system is a feasible tool to irradiate mice using treatment planning and delivery methods analogous to those applied to humans.

Entities:  

Mesh:

Year:  2009        PMID: 19478377      PMCID: PMC2810127          DOI: 10.1088/0031-9155/54/12/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  23 in total

1.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  The volume effect of detectors in the dosimetry of small fields used in IMRT.

Authors:  Wolfram U Laub; Tony Wong
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

Review 3.  Emergent technologies for 3-dimensional image-guided radiation delivery.

Authors:  David A Jaffray
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

4.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine.

Authors:  A Niroomand-Rad; C R Blackwell; B M Coursey; K P Gall; J M Galvin; W L McLaughlin; A S Meigooni; R Nath; J E Rodgers; C G Soares
Journal:  Med Phys       Date:  1998-11       Impact factor: 4.071

5.  A cesium-137 irradiator for small laboratory animals.

Authors:  E B Hranitzky; P R Almond; H D Suit; E B Moore
Journal:  Radiology       Date:  1973-06       Impact factor: 11.105

Review 6.  AAPM code of practice for radiotherapy accelerators: report of AAPM Radiation Therapy Task Group No. 45.

Authors:  R Nath; P J Biggs; F J Bova; C C Ling; J A Purdy; J van de Geijn; M S Weinhous
Journal:  Med Phys       Date:  1994-07       Impact factor: 4.071

7.  Partial volume rat lung irradiation: an evaluation of early DNA damage.

Authors:  M A Khan; R P Hill; J Van Dyk
Journal:  Int J Radiat Oncol Biol Phys       Date:  1998-01-15       Impact factor: 7.038

8.  Histological and elemental changes in the rat brain after local irradiation with carbon ion beams.

Authors:  Sentaro Takahashi; Xue-Zhi Sun; Yoshihisa Kubota; Nobuhiko Takai; Kumie Nojima
Journal:  J Radiat Res       Date:  2002-06       Impact factor: 2.724

9.  Dosimetry in mice exposed to 1.6 GHz microwaves in a carrousel irradiator.

Authors:  M Swicord; J Morrissey; D Zakharia; M Ballen; Q Balzano
Journal:  Bioelectromagnetics       Date:  1999       Impact factor: 2.010

10.  Use of the Leksell Gamma Knife for localized small field lens irradiation in rodents.

Authors:  Colleen DesRosiers; Marc S Mendonca; Craig Tyree; Vadim Moskvin; Morris Bank; Leo Massaro; Robert M Bigsby; Andrea Caperall-Grant; Shailaja Valluri; Joseph R Dynlacht; Robert Timmerman
Journal:  Technol Cancer Res Treat       Date:  2003-10
View more
  16 in total

1.  Dynamic contrast-enhanced micro-CT on mice with mammary carcinoma for the assessment of antiangiogenic therapy response.

Authors:  Fabian Eisa; Robert Brauweiler; Martin Hupfer; Tristan Nowak; Laura Lotz; Inge Hoffmann; David Wachter; Ralf Dittrich; Matthias W Beckmann; Gregor Jost; Hubertus Pietsch; Willi A Kalender
Journal:  Eur Radiol       Date:  2011-11-10       Impact factor: 5.315

2.  Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film.

Authors:  J G H Sutherland; D W O Rogers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

3.  Investigation of the effects of treatment planning variables in small animal radiotherapy dose distributions.

Authors:  Amy R Motomura; Magdalena Bazalova; Hu Zhou; Paul J Keall; Edward E Graves
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

4.  The importance of tissue segmentation for dose calculations for kilovoltage radiation therapy.

Authors:  Magdalena Bazalova; Edward E Graves
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

5.  Kilovoltage beam Monte Carlo dose calculations in submillimeter voxels for small animal radiotherapy.

Authors:  Magdalena Bazalova; Hu Zhou; Paul J Keall; Edward E Graves
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

6.  Novel technique for high-precision stereotactic irradiation of mouse brains.

Authors:  J Hartmann; J Wölfelschneider; C Stache; R Buslei; A Derer; M Schwarz; T Bäuerle; R Fietkau; U S Gaipl; C Bert; A Hölsken; B Frey
Journal:  Strahlenther Onkol       Date:  2016-07-11       Impact factor: 3.621

7.  Perfusion measurements by micro-CT using prior image constrained compressed sensing (PICCS): initial phantom results.

Authors:  Brian E Nett; Robert Brauweiler; Willi Kalender; Howard Rowley; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2010-04-01       Impact factor: 3.609

8.  Development of a micro-computed tomography-based image-guided conformal radiotherapy system for small animals.

Authors:  Hu Zhou; Manuel Rodriguez; Fred van den Haak; Geoffrey Nelson; Rahil Jogani; Jiali Xu; Xinzhi Zhu; Yongjiang Xian; Phuoc T Tran; Dean W Felsher; Paul J Keall; Edward E Graves
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-13       Impact factor: 7.038

9.  Modality comparison for small animal radiotherapy: a simulation study.

Authors:  Magdalena Bazalova; Geoff Nelson; John M Noll; Edward E Graves
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

10.  Proposal for a Simple and Efficient Monthly Quality Management Program Assessing the Consistency of Robotic Image-Guided Small Animal Radiation Systems.

Authors:  N Patrik Brodin; Chandan Guha; Wolfgang A Tomé
Journal:  Health Phys       Date:  2015-11       Impact factor: 1.316

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.