Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

Jehn-Yu Huang1, Melike Pekmezci, Stephanie Yaplee, Shan Lin.   

Abstract

BACKGROUND: To assess the intra-examiner repeatability and agreement of corneal pachymetry maps measured by time-domain (Visante OCT) and Fourier-domain optical coherence tomography (RTVue OCT).
METHODS: This observational cross-sectional study enrolled 72 eyes from 72 volunteers. Only one eye of each patient was chosen randomly to receive repeated scanning with both devices by the same examiner, in order to test the intra-observer repeatability. The first scan by each device from all enrolled eyes was used to analyze the difference and agreement between the two devices. The agreement between the two devices was analyzed by the Bland-Altman method. Intra-observer repeatability of each OCT device was analyzed by intra-class correlation (ICC).
RESULTS: The mean corneal thickness of the central 2 mm zone was 524.3 ± 35.7 µm and 525.4 ± 35.3 µm by the Visante OCT and the RTVue OCT respectively (p = 0.089). The intra-observer repeatability of the RTVue OCT (ICC = 0.994) was superior to that of the Visante OCT (ICC = 0.989) in the central 2 mm zone. However, the intra-observer repeatability of the RTVue OCT in the pericentral 2 to 5 mm zone was not superior to that of the Visante OCT. Both OCT devices had similar repeatability in the pericentral 2 to 5 mm zone [ICC of the Visante OCT = 0.991; 95% confidence interval (CI) 0.986-0.995), ICC of the RTVue OCT = 0.991; 95% CI 0.985-0.994)).
CONCLUSIONS: The difference in CCT measurement by the Visante OCT and the RTVue OCT is probably too small to influence clinical decision making for refractive surgery and glaucoma management. The RTVue OCT demonstrated better intra-observer repeatability in the central 2 mm zone, which probably was related to its rapid image acquisition capability.

Entities:  

Mesh:

Year:  2010        PMID: 20352443     DOI: 10.1007/s00417-010-1360-7

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  31 in total

1.  Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings.

Authors:  N Feltgen; D Leifert; J Funk
Journal:  Br J Ophthalmol       Date:  2001-01       Impact factor: 4.638

2.  Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry.

Authors:  Hassan Hashemi; Maryam Roshani; Shiva Mehravaran; Hiva Parsafar; Kamran Yazdani
Journal:  J Cataract Refract Surg       Date:  2007-10       Impact factor: 3.351

3.  Comparison of fourier-domain and time-domain optical coherence tomography for assessment of corneal thickness and intersession repeatability.

Authors:  Gaurav Prakash; Amar Agarwal; Soosan Jacob; Dhivya Ashok Kumar; Athiya Agarwal; Rumki Banerjee
Journal:  Am J Ophthalmol       Date:  2009-05-13       Impact factor: 5.258

4.  Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes.

Authors:  Claudia Maria Prospero Ponce; Karolinne Maia Rocha; Scott D Smith; Ronald R Krueger
Journal:  J Cataract Refract Surg       Date:  2009-06       Impact factor: 3.351

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  The effect of corneal thickness on applanation tonometry.

Authors:  M M Whitacre; R A Stein; K Hassanein
Journal:  Am J Ophthalmol       Date:  1993-05-15       Impact factor: 5.258

7.  Central corneal thickness of Caucasians and African Americans in glaucomatous and nonglaucomatous populations.

Authors:  F A La Rosa; R L Gross; S Orengo-Nania
Journal:  Arch Ophthalmol       Date:  2001-01

8.  The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma.

Authors:  Mae O Gordon; Julia A Beiser; James D Brandt; Dale K Heuer; Eve J Higginbotham; Chris A Johnson; John L Keltner; J Philip Miller; Richard K Parrish; M Roy Wilson; Michael A Kass
Journal:  Arch Ophthalmol       Date:  2002-06

9.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

10.  A pilot study of Fourier-domain optical coherence tomography of retinal dystrophy patients.

Authors:  Jennifer I Lim; Ou Tan; Amani A Fawzi; J Jill Hopkins; John H Gil-Flamer; David Huang
Journal:  Am J Ophthalmol       Date:  2008-07-17       Impact factor: 5.258

View more
  12 in total

1.  Optical coherence tomography measurements with the LENTIS Mplus multifocal intraocular lens.

Authors:  Eirini Skiadaresi; Colm McAlinden; Giuseppe Ravalico; Jonathan Moore
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-01-04       Impact factor: 3.117

2.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

3.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

4.  Evaluation of rigid gas permeable lens fitting in keratoconic patients with optical coherence tomography.

Authors:  Amal M Elbendary; Waleed Abou Samra
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-08       Impact factor: 3.117

5.  Repeatability of corneal epithelial thickness measurements using Fourier-domain optical coherence tomography in normal and post-LASIK eyes.

Authors:  Xingxuan Jack Ma; Li Wang; Douglas D Koch
Journal:  Cornea       Date:  2013-12       Impact factor: 2.651

6.  Reliability of RTVue, Visante, and slit-lamp adapted ultrasonic pachymetry for central corneal thickness measurement.

Authors:  Jin Pyo Hong; Sang Min Nam; Tae-Im Kim; Kyoung Yul Seo; Sang Yeop Lee; Alessandro Meduri; Eung Kweon Kim
Journal:  Yonsei Med J       Date:  2012-05       Impact factor: 2.759

7.  Intraoperative use of spectral-domain optical coherence tomography during Descemet's stripping automated endothelial keratoplasty.

Authors:  Chelvin Ca Sng; Federico Luengo Gimeno; Jodhbir S Mehta; Hla Myint Htoon; Donald T Tan
Journal:  Clin Ophthalmol       Date:  2012-03-23

8.  Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus.

Authors:  Chunxiao Wang; Xueying Xia; Bishan Tian; Shiyou Zhou
Journal:  J Ophthalmol       Date:  2015-06-25       Impact factor: 1.909

9.  Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.

Authors:  Haya M Al-Farhan; Wafa'a Majed Al-Otaibi
Journal:  Clin Ophthalmol       Date:  2012-07-06

10.  Limitations in imaging common conjunctival and corneal pathologies with fourier-domain optical coherence tomography.

Authors:  Hakan Demirci; Daniel W Steen
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.