Chunxiao Wang1, Xueying Xia2, Bishan Tian1, Shiyou Zhou1.
Abstract
Objective. To compare fourier-domain optical coherence tomography (FD-OCT) and time-domain OCT (TD-OCT) in the determination of thinnest corneal thickness (TCT). Methods. This study included 55 keratoconus patients and 50 healthy volunteers. The RTVue-OCT (FD-OCT) and Visante-OCT (TD-OCT) were used for the measurement of the TCT. Three consecutive scans were performed. The comparison and agreement between the two modalities were analyzed by paired t-test, the Pearson correlation, intraclass correlation coefficient (ICC), and Bland-Altman plots. intraobserver repeatability was analyzed by the intraobserver within-subject standard deviation (S (w)), coefficient of variation (CV(w)), and ICC. Results. The TCT value of normal corneas was higher by RTVue-OCT (530.4 ± 19.7 μm) than Visante-OCT (521.5 ± 18.3 μm) (p < 0.001). For keratoconus eyes, the TCT was 425.0 ± 58.2 μm and 424.4 ± 55.7 μm (difference being 0.6 ± 10.2 μm, p = 0.604). Strong correlation (r = 0.938∼0.985) (ICC = 0.915-0.984) was observed between the two OCTs, and each OCT exhibited excellent repeatability in determining the TCT in all subjects (ICC = 0.984-0.994). Conclusions. The values of TCT obtained from RTVue-OCT and Visante-OCT were highly correlated; however, the two values were different. Both OCT instruments exhibited good intraobserver reliability. The existence of systematic differences suggested that the two instruments cannot be used interchangeably.Entities:
Year: 2015 PMID: 26199737 PMCID: PMC4496494 DOI: 10.1155/2015/402925
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Corneal pachymetry of a patient with keratoconus generated by the (a) RTVue-OCT and (b) Visante-OCT. (a) Minimum of corneal thickness can be acquired from the table of “keratoconus analysis” on the upper right and the thinnest point is also marked on the pachymetric map on the lower right. A real-time monitoring on the upper left is used for pupil centering. (b) Minimum of corneal thickness in different regions is demonstrated on the table on the lower right and the thinnest point is also marked on the pachymetric map on the left.
Difference and correlation analysis of the TCT from the RTVue-OCT and Visante-OCT.
| RTVue-OCT | Visante-OCT | Difference | Correlation | ICC | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | SD | 95% CI-U | 95% CI-L | Mean | SD | 95% CI-U | 95% CI-L | Mean | SD |
|
|
| (95% LoA) | |
| Normal eyes | 530.4 | 19.7 | 500.2 | 564.5 | 521.5 | 18.3 | 493.4 | 555.2 | 8.9 | 6.9 | <0.001 | 0.938 | <0.001 | 0.915 (0.906–0.952) |
| Keratoconus | 425.0 | 58.2 | 307.7 | 504.6 | 424.4 | 55.7 | 310.7 | 499.8 | 0.6 | 10.2 | 0.604 | 0.985 | <0.001 | 0.984 (0.976–0.990) |
| Subgroup 1 | 469.5 | 26.5 | 433.4 | 518.6 | 466.2 | 25.9 | 430.7 | 517.3 | 3.2 | 7.7 | 0.009 | 0.957 | <0.001 | 0.950 (0.899–0.974) |
| Subgroup 2 | 376.0 | 41.7 | 286.7 | 425.7 | 378.3 | 41.3 | 291.3 | 423.0 | −2.3 | 11.8 | 0.223 | 0.960 | <0.001 | 0.959 (0.924–0.978) |
Figure 2Scatterplot showing the relationship of the TCT values between RTVue-OCT and Visante-OCT in keratoconic and normal eyes.
Figure 3Bland-Altman plots showing the agreement between the RTVue-OCT and Visante-OCT measurements of the TCT. (a) Normal eyes. (b) Keratoconic eyes.
Analysis of intraobserver repeatability of each OCT for normal and keratoconic eyes.
| Normal eyes | Keratoconus | |||
|---|---|---|---|---|
| RTVue-OCT | Visante-OCT | RTVue-OCT | Visante-OCT | |
|
| 1.23 | 1.76 | 5.17 | 4.65 |
| CV( | 0.2% | 0.3% | 1.2% | 1.1% |
| ICC (95% LoA) | 0.994 (0.991–0.997) | 0.989 (0.983–0.993) | 0.984 (0.976–0.989) | 0.987 (0.981–0.991) |