| Literature DB >> 22536040 |
Chelvin Ca Sng1, Federico Luengo Gimeno, Jodhbir S Mehta, Hla Myint Htoon, Donald T Tan.
Abstract
PURPOSE: To evaluate the intraoperative changes in the donor lenticule, recipient cornea, and the reduction of interface fluid thickness during Descemet's stripping and automated endothelial keratoplasty with EndoGlide™ (Angiotech Pharmaceuticals Inc, Vancouver, Canada) donor insertion, using intraoperative spectral-domain optical coherence tomography.Entities:
Keywords: Descemet’s stripping and automated endothelial keratoplasty; spectral-domain optical coherence tomography
Year: 2012 PMID: 22536040 PMCID: PMC3334228 DOI: 10.2147/OPTH.S28971
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Photograph of the spectral-domain optical coherence tomography device used intraoperatively during Descemet’s stripping and automated endothelial keratoplasty.
Estimated mean thickness of the donor, recipient, and interface fluid at each surgical step during Descemet’s stripping and endothelial keratoplasty
| Surgical step | Estimated mean thickness (μm) | 95% confidence interval for estimated mean thickness (μm) | ||
|---|---|---|---|---|
|
| ||||
| Lower | Upper | |||
| 1. Immediately prior to air tamponade | 239.5 | 194.0 | 285.0 | NA |
| 2. After fluid expression through venting incisions | 266.1 | 220.6 | 311.5 | 0.102 (1 versus 2) |
| 3. 6 minutes of air tamponade | 282.9 | 237.5 | 328.4 | 0.004 (1 versus 3) |
| 4. 10 minutes of air tamponade | 220.7 | 175.2 | 266.2 | 0.352 (1 versus 4) |
| 1. Immediately prior to air tamponade | 674.7 | 624.9 | 724.5 | NA |
| 2. After fluid expression through venting incisions | 606.3 | 556.1 | 656.5 | <0.001 (1 versus 2) |
| 3. 6 minutes of air tamponade | 588.4 | 538.3 | 638.6 | <0.001 (1 versus 3) |
| 4. 10 minutes of air tamponade | 581.9 | 532.1 | 631.7 | 0.001 (1 versus 4) |
| 1. Immediately prior to air tamponade | 495.5 | 475.9 | 515.0 | NA |
| 2. After fluid expression through venting incisions | 67.5 | 37.3 | 97.6 | <0.001 (1 versus 2) |
| 3. 6 minutes of air tamponade | 33.0 | 24.6 | 41.4 | <0.001 (1 versus 3) |
| 4. 10 minutes of air tamponade | 2.3 | 0.9 | 3.7 | <0.001 (1 versus 4) |
Figure 2Intraoperative estimated mean thickness of the donor lenticule, recipient cornea, and interface fluid during Descemet’s stripping and automated endothelial keratoplasty.
Note: *Error bars indicate standard error.
Figure 3Intraoperative spectral-domain optical coherence tomography scans at the following surgical stages of Descemet’s stripping and automated endothelial keratoplasty. (A) After donor insertion and immediately prior to full air tamponade. (B) After air tamponade and immediately prior to the expression of fluid from the venting incisions. (C) After the expression of fluid from the venting incisions. (D) At 6 minutes of air tamponade. No interface fluid can be detected in this patient at this stage.