In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T.

Michael Lettau1, Annett Sauer, Sabine Heiland, Stefan Rohde, Julia Reinhardt, Martin Bendszus, Stefan Hähnel.   

Abstract

INTRODUCTION: CT angiography (CTA) and MR angiography (MRA) are increasingly used methods for evaluation of stented vessel segments. The purpose of this study was to compare CTA, contrast-enhanced MRA (CEMRA) at 1.5 T, and CEMRA at 3 T for the visualization of carotid artery stents and to define the best noninvasive imaging technique for each stent.
METHODS: CTA and CEMRA appearances of 18 carotid artery stents of different designs and sizes (4.0 to 10.0 mm) were investigated in vitro. The profile of the contrast-to-noise ratio (CNR) of the lumen of each stent was calculated semiautomatically by a pixel-by-pixel analysis using the medical imaging software OSIRIS. For each stent, artificial lumen narrowing (ALN) was calculated.
RESULTS: In all but one stents, ALN was lower on CEMRA at 3 T than at 1.5 T. With CEMRA at 3 T and at 1.5 T, ALN in most nitinol stents was lower than in the groups of stainless steel and cobalt alloy stents. In most nitinol stents, ALN on CEMRA at 3 T was lower than on CTA. In all stainless steel stents and cobalt alloy stents, ALN was lower on CTA than on CEMRA. With CTA and CEMRA, in most stents ALN decreased with increasing stent diameter.
CONCLUSION: CTA and CEMRA evaluation of vessel patency after stent placement is possible, but considerably impaired by ALN. Investigators should be informed about the method of choice for every stent.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19924409     DOI: 10.1007/s00234-009-0625-5

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  30 in total

1.  Improved in-stent magnetic resonance angiography with high flip angle excitation.

Authors:  J M Meyer; A Buecker; E Spuentrup; K Schuermann; M Huetten; R D Hilgers; J J van Vaals; R W Guenther
Journal:  Invest Radiol       Date:  2001-11       Impact factor: 6.016

2.  Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents.

Authors:  M Lenhart; M Völk; C Manke; W R Nitz; M Strotzer; S Feuerbach; J Link
Journal:  Radiology       Date:  2000-10       Impact factor: 11.105

3.  Revealing in-stent stenoses of the iliac arteries: comparison of multidetector CT with MR angiography and digital radiographic angiography in a Phantom model.

Authors:  David Maintz; Bernd Tombach; Kai-Uwe Juergens; Stefanie Weigel; Walter Heindel; Roman Fischbach
Journal:  AJR Am J Roentgenol       Date:  2002-11       Impact factor: 3.959

4.  3.0 Tesla magnetic resonance angiography of endovascular aortic stent grafts: phantom measurements in comparison with 1.5 Tesla.

Authors:  Stefan C Krämer; Alexander Wall; David Maintz; Rainald Bachmann; Harald Kugel; Walter Heindel
Journal:  Invest Radiol       Date:  2004-07       Impact factor: 6.016

5.  3.0 T vs. 1.5 T MR angiography: in vitro comparison of intravascular stent artifacts.

Authors:  Alexander Wall; Harald Kugel; Rainald Bachman; Lars Matuszewski; Stefan Krämer; Walter Heindel; David Maintz
Journal:  J Magn Reson Imaging       Date:  2005-12       Impact factor: 4.813

6.  64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents.

Authors:  David Maintz; Harald Seifarth; Rainer Raupach; Thomas Flohr; Michael Rink; Torsten Sommer; Murat Ozgün; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2005-12-07       Impact factor: 5.315

7.  Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study.

Authors:  M Bendszus; M Koltzenburg; R Burger; M Warmuth-Metz; E Hofmann; L Solymosi
Journal:  Lancet       Date:  1999-11-06       Impact factor: 79.321

8.  Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent.

Authors:  Arno Buecker; Elmar Spuentrup; Alexander Ruebben; Rolf W Günther
Journal:  Circulation       Date:  2002-04-16       Impact factor: 29.690

9.  MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents.

Authors:  L W Bartels; H F Smits; C J Bakker; M A Viergever
Journal:  J Vasc Interv Radiol       Date:  2001-03       Impact factor: 3.464

10.  Complications of cerebral angiography: a prospective analysis of 2,924 consecutive procedures.

Authors:  A A Dawkins; A L Evans; J Wattam; C A J Romanowski; D J A Connolly; T J Hodgson; S C Coley
Journal:  Neuroradiology       Date:  2007-06-27       Impact factor: 2.804

View more
  7 in total

1.  [CT diagnostics of carotid artery stenosis].

Authors:  W Reith; C Roth
Journal:  Radiologe       Date:  2010-07       Impact factor: 0.635

2.  Comparing different MR angiography strategies of carotid stents in a vascular flow model: toward stent-specific recommendations in MR follow-up.

Authors:  Andreas M J Frölich; Sara M Pilgram-Pastor; Marios N Psychogios; Alexander Mohr; Michael Knauth
Journal:  Neuroradiology       Date:  2010-08-19       Impact factor: 2.804

3.  Angiographic CT: in vitro comparison of different carotid artery stents-does stent orientation matter?

Authors:  Michael Lettau; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2013-02-15       Impact factor: 2.804

4.  Off-resonance magnetic resonance angiography improves visualization of in-stent lumen in peripheral nitinol stents compared to conventional T1-weighted acquisitions: an in vitro comparison study.

Authors:  Gitsios Gitsioudis; Philipp Fortner; Matthias Stuber; Anna Missiou; Florian Andre; Oliver J Müller; Hugo A Katus; Grigorios Korosoglou
Journal:  Int J Cardiovasc Imaging       Date:  2016-08-17       Impact factor: 2.357

5.  In vitro stent lumen visualisation of various common and newly developed femoral artery stents using MR angiography at 1.5 and 3 tesla.

Authors:  R Syha; D Ketelsen; M Kaempf; S Mangold; S Sixt; T Zeller; F Springer; F Schick; C D Claussen; K Brechtel
Journal:  Eur Radiol       Date:  2012-08-17       Impact factor: 5.315

6.  Artificial luminal narrowing on contrast-enhanced magnetic resonance angiograms on an occasion of stent-assisted coiling of intracranial aneurysm: in vitro comparison using two different stents with variable imaging parameters.

Authors:  Jee Hyun Seok; Hyun Seok Choi; So-Lyung Jung; Kook-Jin Ahn; Myeong Jin Kim; Yong Sam Shin; Bum-soo Kim
Journal:  Korean J Radiol       Date:  2012-08-28       Impact factor: 3.500

7.  Pediatric cardiovascular interventional devices: effect on CMR images at 1.5 and 3 Tesla.

Authors:  Sarah N Khan; Stanislas Rapacchi; Daniel S Levi; J Paul Finn
Journal:  J Cardiovasc Magn Reson       Date:  2013-06-19       Impact factor: 5.364

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.